Electrochemical biosensor for early Alzheimer's detection and patient risk stratification using plasma exosomes
Biosensors and Bioelectronics · 2026
DOI: 10.1016/j.bios.2025.118061 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2026. Verfasst von Rosanna Rossi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Biosensing and Bioanalysis Group, Institute of Biotechnology and Biomedicine, Universitat Autònoma de Barcelona, Spain; Grup de Sensors i Biosensors,…. Hauptthemen laut MEDLINE: Alzheimer Disease, Exosomes, Biosensing Techniques, Amyloid Precursor Protein Secretases. Die Arbeit nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Instituto de Salud Carlos III, Ministerio de Ciencia e Innovación, Centro de Investigación Biomédica en Red sobre Enfermedades Neurodegenerativas.
Themen: Demenz
Abstract (Originalsprache)
Alzheimer's Disease (AD) is the leading cause of dementia, accounting for 60-70 % of cases worldwide. Early diagnosis remains challenging due to the limitations of current diagnostic tools, which are costly, invasive, and suffer from low patient compliance. Blood-based biomarkers, particularly plasma brain-derived exosomes (BDEs), have emerged as a promising alternative since they carry AD-related molecules and can be isolated non-invasively. In this study, an immunoassay was developed to isolate BDEs using magnetic particles functionalized with an anti-neuroligin-3 (NLGN3) antibody, while the AD-related marker β-secretase (BACE-1) was detected on the captured exosomes. This is the first report combining NLGN3 for for the isolation of BDEs with BACE-1 as a detection target, establishing a novel biomarker panel for AD diagnostics. The assay was evaluated across three readout platforms-optical, chemiluminescent, and electrochemical-with detection limits in the range of 104-105 exosomes μL-1. Among them, the portable electrochemical platform achieved the improved LOD (1.51 × 104 exosomes μL-1, R2 = 0.9829). Plasma samples from patients with AD, mild cognitive impairment (MCI), and healthy controls were analyzed, revealing differences in exosomal BACE-1 levels (p < 0.1, t-test). These findings demonstrate, for the first time, an in vitro diagnostic approach based on a portable electrochemical biosensor for early AD detection in plasma through a novel exosomal biomarker panel. Compared to conventional diagnostics, this biosensor offers a non-invasive and cost-effective solution for AD screening, with the potential to support earlier intervention and patient risk stratification.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Biosensing and Bioanalysis Group, Institute of Biotechnology and Biomedicine, Universitat Autònoma de Barcelona, Spain; Grup de Sensors i Biosensors, Departament de Química, Universitat Autònoma de Barcelona, Bellaterra…
- Research Center and Memory Clinic, Ace Alzheimer Center Barcelona, Barcelona, Spain; Biomedical Research Networking Centre in Neurodegenerative Diseases (CIBERNED), Madrid, Spain.
- BioEclosion SL, Avda. Can Domènech, Edifici Eureka, Universitat Autònoma de Barcelona, Bellaterra, Spain.
- Research Center and Memory Clinic, Ace Alzheimer Center Barcelona, Barcelona, Spain; Biomedical Research Networking Centre in Neurodegenerative Diseases (CIBERNED), Madrid, Spain; Glenn Biggs Institute for Alzheimer's &…
- Biosensing and Bioanalysis Group, Institute of Biotechnology and Biomedicine, Universitat Autònoma de Barcelona, Spain; Immunology Unit, Department of Cell Biology, Physiology, and Immunology, Institute of Biotechnology…
Förderung
- Instituto de Salud Carlos III
- Ministerio de Ciencia e Innovación
- Centro de Investigación Biomédica en Red sobre Enfermedades Neurodegenerativas
- Spain Ministry of Science and Innovation
Zitieren
APA
Rossi, R., Cano, A., Pallarès-Rusiñol, A., Ruiz, A., Martí, M., & Pividori, M. I. (2026). Electrochemical biosensor for early Alzheimer's detection and patient risk stratification using plasma exosomes. Biosensors and Bioelectronics. https://doi.org/10.1016/j.bios.2025.118061
Vancouver
Rossi R, Cano A, Pallarès-Rusiñol A, Ruiz A, Martí M, Pividori MI. Electrochemical biosensor for early Alzheimer's detection and patient risk stratification using plasma exosomes. Biosensors and Bioelectronics. 2026. doi:10.1016/j.bios.2025.118061
BibTeX
@article{Rossi2026,
author = {Rossi, Rosanna and Cano, Amanda and Pallarès-Rusiñol, Arnau and Ruiz, Agustín and Martí, Merce and Pividori, Maria Isabel},
title = {Electrochemical biosensor for early Alzheimer's detection and patient risk stratification using plasma exosomes},
journal = {Biosensors and Bioelectronics},
year = {2026},
doi = {10.1016/j.bios.2025.118061}
}
EUVIMED