EUVIMEDEUVIMED entdecken
Crossref

Cyclodextrin-integrated molecularly imprinted polymer-based QCM sensor for sensitive, label-free detection of Gal3

Phoomintara Longsompurana, Peter Wolschann, Kittitat Subannajui, Peter A Lieberzeit, Rungtiva P Poo-Arporn

Biosensors and Bioelectronics · 2025-11 · 287 · 117712

DOI: 10.1016/j.bios.2025.117712 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2025-11. Verfasst von Phoomintara Longsompurana und 4 weiteren Autor:innen. Beteiligte Einrichtung: Biological Engineering Program, Faculty of Engineering, King Mongkut's University of Technology Thonburi, Bangkok, 10140, Thailand. Hauptthemen laut MEDLINE: Quartz Crystal Microbalance Techniques, Biosensing Techniques, Molecularly Imprinted Polymers, Galectin 3. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 46 Literaturangaben. Gefördert von: Thailand Science Research and Innovation, National Research Council of Thailand, King Mongkut's University of Technology Thonburi.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The global prevalence of heart failure (HF) poses a significant public health challenge. Elevated serum levels of Galectin-3 (Gal3) are a reliable biomarker for the early detection and monitoring of HF progression. Given the high mortality and morbidity associated with HF, early diagnosis is critical, necessitating a rapid, sensitive, and straightforward method for Gal3 detection. This study presents an innovative, label-free quartz crystal microbalance (QCM) sensor designed for Gal3 detection. The sensor uses hexakis-(6-mercapto-6-deoxy)-β-cyclodextrin (mCD) to facilitate the oriented immobilization of the Gal3 template through self-assembly and employs a molecularly imprinted polymer (MIP) for Gal3-specific detection. The mCD-integrated MIP was synthesized by electropolymerizing poly-aniline-3-carboxylic acid (ANI3C), chosen from 20 candidate polymers via molecular docking. Under optimal conditions, the MIP-based QCM sensor demonstrated an imprinting factor (IF) of 4.71 and an equilibrium dissociation constant (KD) of 2.03 nM. This indicates a stronger binding affinity compared to the non-molecularly imprinted polymer (NIP), which exhibited a KD of 26.75 nM. The sensor exhibits a linear detection range of 1-30 ng/mL, with a limit of detection (LOD) of 1.08 ng/mL (S/N = 3) and a limit of quantification (LOQ) of 3.61 ng/mL (S/N = 10). It demonstrated excellent accuracy and precision with no interference, and successfully detected Gal3 in lipid-free human serum, highlighting its potential for in heart failure diagnostics and monitoring.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Quartz Crystal Microbalance TechniquesBiosensing TechniquesMolecularly Imprinted PolymersGalectin 3CyclodextrinsHeart FailureHumansLimit of DetectionMolecular ImprintingMolecular Docking SimulationPolymersBlood ProteinsGalectins

Stichwörter der Autor:innen

Galectin-3Heart failureHexakis-(6-mercapto-6-deoxy)-β-cyclodextrinMolecular dockingMolecularly imprinted polymerQuartz crystal microbalance sensor

Substanzen

Molecularly Imprinted PolymersGalectin 3CyclodextrinsLGALS3 protein, humanPolymersBlood ProteinsGalectins

Autor:innen und Institutionen

  • Biological Engineering Program, Faculty of Engineering, King Mongkut's University of Technology Thonburi, Bangkok, 10140, Thailand.
  • Institute of Theoretical Chemistry, University of Vienna, Vienna, 1090, Austria.
  • School of Materials Science and Innovation, Faculty of Science, Mahidol University, Bangkok, 10400, Thailand.
  • Department of Physical Chemistry, Faculty for Chemistry, University of Vienna, Vienna, 1090, Austria.
  • Biological Engineering Program, Faculty of Engineering, King Mongkut's University of Technology Thonburi, Bangkok, 10140, Thailand. Electronic address: rungtiva.pal@kmutt.ac.th.

Förderung

  • Thailand Science Research and Innovation
  • National Research Council of Thailand
  • King Mongkut's University of Technology Thonburi

Zitieren

APA

Longsompurana, P., Wolschann, P., Subannajui, K., Lieberzeit, P. A., & Poo-Arporn, R. P. (2025). Cyclodextrin-integrated molecularly imprinted polymer-based QCM sensor for sensitive, label-free detection of Gal3. Biosensors and Bioelectronics, 287, 117712. https://doi.org/10.1016/j.bios.2025.117712

Vancouver

Longsompurana P, Wolschann P, Subannajui K, Lieberzeit PA, Poo-Arporn RP. Cyclodextrin-integrated molecularly imprinted polymer-based QCM sensor for sensitive, label-free detection of Gal3. Biosensors and Bioelectronics. 2025;287:117712. doi:10.1016/j.bios.2025.117712

BibTeX
@article{Longsompurana2025,
  author = {Longsompurana, Phoomintara and Wolschann, Peter and Subannajui, Kittitat and Lieberzeit, Peter A and Poo-Arporn, Rungtiva P},
  title = {Cyclodextrin-integrated molecularly imprinted polymer-based QCM sensor for sensitive, label-free detection of Gal3},
  journal = {Biosensors and Bioelectronics},
  year = {2025},
  volume = {287},
  pages = {117712},
  doi = {10.1016/j.bios.2025.117712}
}
Kostenlos starten →