Detection of H. pylori outer membrane protein (HopQ) biomarker using electrochemical impedimetric immunosensor with polypyrrole nanotubes and carbon nanotubes nanocomposite on screen-printed carbon electrode
Biosensors and Bioelectronics · 2024-04 · 249 · 115937
DOI: 10.1016/j.bios.2023.115937 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2024-04. Verfasst von Hussamaldeen Jaradat und 7 weiteren Autor:innen. Beteiligte Einrichtung: Professorship of Measurement and Sensor Technology, Faculty of Electrical Engineering and Information Technology, Chemnitz University of Technology, …. Hauptthemen laut MEDLINE: Nanotubes, Carbon, Helicobacter pylori, Biosensing Techniques, Nanocomposites. Die Arbeit wurde bisher 36-mal zitiert (Crossref) und nennt 59 Literaturangaben. Gefördert von: Coordenação de Aperfeiçoamento de Pessoal de Nível Superior, Deutsche Forschungsgemeinschaft, CNPq.
Abstract (Originalsprache)
Helicobacter pylori (H. pylori) is classified as a class I carcinogen that colonizes the human gastrointestinal (GI) tract. The detection at low concentrations is crucial in combatting H. pylori. HopQ protein is located on H. pylori's outer membrane and is expressed at an early stage of contamination, which signifies it as an ideal biomarker. In this study, we presented the development of an electrochemical impedimetric immunosensor for the ultra-sensitive detection of HopQ at low concentrations. The sensor employed polypyrrole nanotubes (PPy-NTs) and carboxylated multi-walled carbon nanotubes (MWCNT-COOH) nanocomposite. PPy-NTs were chosen for their excellent conductivity, biocompatibility, and redox capabilities, simplifying sample preparation by eliminating the need to add redox probes upon measurement. MWCNT-COOH provided covalent binding sites for HopQ antibodies (HopQ-Ab) on the biosensor surface. Characterization of the biosensor was performed using scanning electron microscopy (SEM), Fourier transform infrared spectroscopy (FTIR), contact angle measurements, and electrochemical impedance spectroscopy (EIS), complemented by numerical semiempirical quantum calculations. The results demonstrated a dynamic linear range of 5 pg/mL to 1.063 ng/mL and an excellent selectivity, with the possibility of excluding interference using EIS data, specifically charge transfer resistance and double-layer capacitance as multivariants for the calibration curve. Using two EIS components, the limit of detection is calculated to be 2.06 pg/mL. The biosensor was tested with a spiked drinking water sample and showed a signal recovery of 105.5% when detecting 300 pg/mL of HopQ. This novel H. pylori biosensor offers reliable, simple, portable, and rapid screening of the bacteria.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Professorship of Measurement and Sensor Technology, Faculty of Electrical Engineering and Information Technology, Chemnitz University of Technology, 09126, Chemnitz, Germany. Electronic address: hujar@hrz.tu-chemnitz.de.
- Grupo de Pesquisa em Macromoléculas e Interfaces, Departamento de Química, Universidade Federal do Paraná (UFPR), Curitiba, 81531-980, PR, Brazil. Electronic address: bruna.hry@ufpr.br.
- University of Belgrade-Faculty of Physical Chemistry, Studentski trg 12-16, 11158, Belgrade, Serbia. Electronic address: igor@ffh.bg.ac.rs.
- Grupo de Pesquisa em Macromoléculas e Interfaces, Departamento de Química, Universidade Federal do Paraná (UFPR), Curitiba, 81531-980, PR, Brazil. Electronic address: tatianalvalerio@ufpr.br.
- Professorship of Measurement and Sensor Technology, Faculty of Electrical Engineering and Information Technology, Chemnitz University of Technology, 09126, Chemnitz, Germany. Electronic address: ammar.al-hamry@etit.tu-c…
- Grupo de Estudos em Espectroscopia de Impedância Eletroquímica (GEIS), Universidade Tecnológica Federal Do Paraná, Rua Dr. Washington Subtil Chueire, 330 - Jd. Carvalho, CEP 84017-220, Ponta Grossa, PR, Brazil. Electron…
Förderung
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
- Deutsche Forschungsgemeinschaft
- CNPq
- FAPESP
- Deutscher Akademischer Austauschdienst
- Instituto Nacional de Ciência e Tecnologia para Excitotoxicidade e Neuroproteção
- Republic of Serbia Ministry of Education Science and Technological Development
Zitieren
APA
Jaradat, H., Hryniewicz, B. M., Pašti, I. A., Valério, T. L., Al-Hamry, A., Marchesi, L. F., Vidotti, M., & Kanoun, O. (2024). Detection of H. pylori outer membrane protein (HopQ) biomarker using electrochemical impedimetric immunosensor with polypyrrole nanotubes and carbon nanotubes nanocomposite on screen-printed carbon electrode. Biosensors and Bioelectronics, 249, 115937. https://doi.org/10.1016/j.bios.2023.115937
Vancouver
Jaradat H, Hryniewicz BM, Pašti IA, Valério TL, Al-Hamry A, Marchesi LF, et al. Detection of H. pylori outer membrane protein (HopQ) biomarker using electrochemical impedimetric immunosensor with polypyrrole nanotubes and carbon nanotubes nanocomposite on screen-printed carbon electrode. Biosensors and Bioelectronics. 2024;249:115937. doi:10.1016/j.bios.2023.115937
BibTeX
@article{Jaradat2024,
author = {Jaradat, Hussamaldeen and Hryniewicz, Bruna M and Pašti, Igor A and Valério, Tatiana L and Al-Hamry, Ammar and Marchesi, Luís F and Vidotti, Marcio and Kanoun, Olfa},
title = {Detection of H. pylori outer membrane protein (HopQ) biomarker using electrochemical impedimetric immunosensor with polypyrrole nanotubes and carbon nanotubes nanocomposite on screen-printed carbon electrode},
journal = {Biosensors and Bioelectronics},
year = {2024},
volume = {249},
pages = {115937},
doi = {10.1016/j.bios.2023.115937}
}
EUVIMED