Systematic review of the diagnostic accuracy of noninvasive intracranial pressure monitoring using optic nerve sheath diameter and the Brain4Care system
Brain and Spine · 2026 · 6 · 106036
DOI: 10.1016/j.bas.2026.106036 ↗Kurz gefasst
Übersichtsarbeit in der Zeitschrift Brain and Spine, 2026. Verfasst von Marcus Roland Victor Gustafsson und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physiology and Pharmacology, Karolinska Institutet, Stockholm, Sweden. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Swedish Research Council.
Abstract (Originalsprache)
Background
Assessing the intracranial pressure (ICP) with accuracy is essential in management of acute brain injury. Invasive monitoring remains the gold standard but carries procedural risks and requires neurosurgical competence. Noninvasive methods such as optic nerve sheath diameter (ONSD) ultrasonography and the Brain4Care (B4C) system have emerged as potential alternatives for estimating ICP or intracranial compliance.
Methods
A systematic search of PubMed and Cochrane Library was conducted for studies published up to October 2025 evaluating ONSD or B4C against invasive ICP monitoring in adult patients. Data on diagnostic performance, including sensitivity, specificity and correlation with invasive ICP, were extracted and summarized.
Results
A total of 47 studies was included, comprising 36 ONSD and 11 B4C investigations. ONSD cut-offs ranged from 4.1 to 7.2 mm, most commonly between 5.0 and 5.9 mm. Reported sensitivities were 70-100%, specificities 70-100% and correlation coefficients r = 0.64-0.90 with invasive ICP. B4C studies evaluated the P2/P1 ratio and time-to-peak (TTP) parameters, with sensitivities of 70-100% and specificities 41-92%. While some studies demonstrated strong correlations with invasive ICP (r up to 0.98), others reported modest accuracy and low specificity, emphasizing dependence on waveform quality and patient population.
Conclusion
ONSD currently remains as the most validated noninvasive method for estimating raised ICP, demonstrating consistently high diagnostic accuracy across studies. The B4C system shows promise for continuous, noninvasive monitoring of intracranial compliance but requires further validation in larger, multicenter settings.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Physiology and Pharmacology, Karolinska Institutet, Stockholm, Sweden.
- Unit of Neurosurgery, Department of Neurosciences, Azienda Ospedaliera Universitaria Senese Le Scotte, Siena, Italy.
- Universidade Do Estado Do Rio de Janeiro (UERJ), Rio de Janeiro, RJ, Brazil.
- Department of Medical Sciences, Section of Neurosurgery, Uppsala University, Uppsala, Sweden.
Förderung
- Swedish Research Council
Zitieren
APA
Gustafsson, M. R. V., Lippa, L., Barros de Sá, G., Velle, F., & Rostami, E. (2026). Systematic review of the diagnostic accuracy of noninvasive intracranial pressure monitoring using optic nerve sheath diameter and the Brain4Care system. Brain and Spine, 6, 106036. https://doi.org/10.1016/j.bas.2026.106036
Vancouver
Gustafsson MRV, Lippa L, Barros de Sá G, Velle F, Rostami E. Systematic review of the diagnostic accuracy of noninvasive intracranial pressure monitoring using optic nerve sheath diameter and the Brain4Care system. Brain and Spine. 2026;6:106036. doi:10.1016/j.bas.2026.106036
BibTeX
@article{Gustafsson2026,
author = {Gustafsson, Marcus Roland Victor and Lippa, Laura and Barros de Sá, Grace and Velle, Fartein and Rostami, Elham},
title = {Systematic review of the diagnostic accuracy of noninvasive intracranial pressure monitoring using optic nerve sheath diameter and the Brain4Care system},
journal = {Brain and Spine},
year = {2026},
volume = {6},
pages = {106036},
doi = {10.1016/j.bas.2026.106036}
}
EUVIMED