Results of Cementless Total Hip Arthroplasty Using Third-Generation Ceramic-On-Ceramic Bearings: A Minimum 15-Year Follow-Up
The Journal of Arthroplasty · 2025
DOI: 10.1016/j.arth.2025.08.034 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Arthroplasty, 2025. Verfasst von Ji Hoon Bahk und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic Surgery, Seoul St. Mary's Hospital, School of Medicine, The Catholic University of Korea, Seoul, South Korea. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Hip Prosthesis, Ceramics, Prosthesis Design. Die Arbeit nennt 38 Literaturangaben. Gefördert von: The Catholic University of Korea.
Abstract (Originalsprache)
BACKGROUND: The third-generation ceramic-on-ceramic total hip arthroplasty (THA) offers excellent wear resistance and biocompatibility but is associated with unique complications such as ceramic fracture and squeaking. While fourth-generation ceramics have become the standard, long-term outcomes of third-generation bearings remain clinically relevant, especially in patients still living with these implants. METHODS: We retrospectively reviewed 533 hips in 417 patients who underwent primary alumina-on-alumina THA using ceramic bearings with a minimum 15-year follow-up (mean 17.4). After excluding deaths and losses to follow-up, 317 hips in 234 patients were analyzed. Kaplan-Meier analysis assessed implant survivorship, and multivariate logistic regressions identified risk factors for ceramic fractures, reoperations, and audible joint noises. RESULTS: The 20-year implant survivorship was 85.2% when considering reoperation for any cause and 94.4% when limited to ceramic fracture. Most fractures (11 of 12) occurred after 10.9 years postoperatively. Significant risk factors for ceramic fracture included men (odds ratio (OR) = 6.94, P = 0.022), higher body mass index (exponential β = 1.22, P = 0.035), short-neck heads (OR = 9.17, P = 0.020), sandwich liners (OR = 17.41, P = 0.007), and prior audible noise (OR = 6.03, P = 0.013). Squeaking and clunking were particularly associated with increased fracture and reoperation risk. Despite these complications, average patient-reported outcome measures remained within acceptable ranges. CONCLUSIONS: A gradual decline in survivorship was observed beyond 15 years in third-generation ceramic-on-ceramic THAs, with a notable incidence of late-onset ceramic fractures. Risk factors included men, higher body mass index, short-neck heads, sandwich liners, and articular noises. The use of 28-mm heads, even with medium necks, should be avoided when possible in favor of 32-mm or larger heads. For patients who have previously had a 28-mm implant, annual follow-up is recommended to monitor for late complications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopedic Surgery, Seoul St. Mary's Hospital, School of Medicine, The Catholic University of Korea, Seoul, South Korea.
Förderung
- The Catholic University of Korea
Zitieren
APA
Bahk, J. H., Jo, W. L., Lee, K. H., Song, J. H., Kim, S. C., & Lim, Y. W. (2025). Results of Cementless Total Hip Arthroplasty Using Third-Generation Ceramic-On-Ceramic Bearings: A Minimum 15-Year Follow-Up. The Journal of Arthroplasty. https://doi.org/10.1016/j.arth.2025.08.034
Vancouver
Bahk JH, Jo WL, Lee KH, Song JH, Kim SC, Lim YW. Results of Cementless Total Hip Arthroplasty Using Third-Generation Ceramic-On-Ceramic Bearings: A Minimum 15-Year Follow-Up. The Journal of Arthroplasty. 2025. doi:10.1016/j.arth.2025.08.034
BibTeX
@article{Bahk2025,
author = {Bahk, Ji Hoon and Jo, Woo-Lam and Lee, Kee-Haeng and Song, Joo-Hyoun and Kim, Seung-Chan and Lim, Young Wook},
title = {Results of Cementless Total Hip Arthroplasty Using Third-Generation Ceramic-On-Ceramic Bearings: A Minimum 15-Year Follow-Up},
journal = {The Journal of Arthroplasty},
year = {2025},
doi = {10.1016/j.arth.2025.08.034}
}
EUVIMED