Patient-Specific Instrumentation Versus Standard of Care in Total Knee Arthroplasty in an Obese Population: Minimum 5-year Follow-Up

Sebastian Braun, James Allen, Lyndsay Somerville, James L Howard, Brent A Lanting, Edward M Vasarhelyi

The Journal of Arthroplasty · 2026

DOI: 10.1016/j.arth.2025.06.060 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift The Journal of Arthroplasty, 2026. Verfasst von Sebastian Braun und 5 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Fellow in Hip and Knee Adult Reconstruction, Division of Orthopaedic Surgery, Department of Surgery, Schulich School of Medicine and Dentist…. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Obesity, Osteoarthritis, Knee, Standard of Care. Die Arbeit nennt 39 Literaturangaben. Gefördert von: DePuy Synthes, Total, DePuy Synthes Mitek Sports Medicine.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Patient-specific instrumentation (PSI) in total knee arthroplasty (TKA) was developed to improve alignment accuracy by using custom cutting guides. Obese patients pose unique technical challenges, and enhanced alignment methods may be especially beneficial. This study compared PSI with standard of care (SOC) using conventional instrumentation in obese patients undergoing primary TKA. METHODS: Patients who had a body mass index greater than 30 and undergoing primary TKA were randomized to SOC or PSI. Patients were followed for five years postoperatively. Patient-reported outcome measures, surgical information, postoperative lower extremity alignment, complications, and revision rates were recorded. There were 158 patients included (80 in PSI and 78 in SOC). RESULTS: The primary radiographic outcome, medial proximal tibial angle, showed a statistically significant, but not clinically meaningful, difference between the groups (PSI = 88.2 ± 2.2, SOC = 88.7 ± 1.8, P = 0.02). No significant differences were observed in the anatomic femoral-tibial angle (P = 0.075) or lateral distal femoral angle (P = 0.53). Patient-reported outcomes, including Western Ontario and McMaster Universities Arthritis Index (PSI: 82.5 ± 18.7, SOC: 83.2 ± 15.9, P = 0.83), 12-Item Short-Form Survey physical and mental composite scores (P = 0.35 and P = 0.40), and Knee Society Score total scores (PSI: 177.3 ± 24.3, SOC: 168.2 ± 31.2, P = 0.08), were similar between the groups. A higher rate of deviation from the preoperative surgical plan was observed in the PSI group (30.2%) compared to SOC (5.1%). Complication and revision rates were comparable between the groups. CONCLUSIONS: Use of PSI in primary TKA did not result in clinically relevant improvements in radiographic alignment compared to SOC. Despite increased preoperative planning, intraoperative deviations from the PSI surgical plan were common. There were no differences in patient-reported outcome measures, complications, or revision rates observed at five years.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, KneeObesityOsteoarthritis, KneeStandard of CareHumansFemaleMaleAgedMiddle AgedFollow-Up StudiesPatient Reported Outcome MeasuresKnee JointTreatment OutcomeBody Mass IndexKnee Prosthesis

Stichwörter der Autor:innen

custom instrumentationobese patientspatient specific-instrumentationradiographic leg alignmenttotal knee arthroplasty

Autor:innen und Institutionen

  • Clinical Fellow in Hip and Knee Adult Reconstruction, Division of Orthopaedic Surgery, Department of Surgery, Schulich School of Medicine and Dentistry, Western University and London Health Sciences Centre, London, Onta…

Förderung

  • DePuy Synthes
  • Total
  • DePuy Synthes Mitek Sports Medicine

Zitieren

APA

Braun, S., Allen, J., Somerville, L., Howard, J. L., Lanting, B. A., & Vasarhelyi, E. M. (2026). Patient-Specific Instrumentation Versus Standard of Care in Total Knee Arthroplasty in an Obese Population: Minimum 5-year Follow-Up. The Journal of Arthroplasty. https://doi.org/10.1016/j.arth.2025.06.060

Vancouver

Braun S, Allen J, Somerville L, Howard JL, Lanting BA, Vasarhelyi EM. Patient-Specific Instrumentation Versus Standard of Care in Total Knee Arthroplasty in an Obese Population: Minimum 5-year Follow-Up. The Journal of Arthroplasty. 2026. doi:10.1016/j.arth.2025.06.060

BibTeX
@article{Braun2026,
  author = {Braun, Sebastian and Allen, James and Somerville, Lyndsay and Howard, James L and Lanting, Brent A and Vasarhelyi, Edward M},
  title = {Patient-Specific Instrumentation Versus Standard of Care in Total Knee Arthroplasty in an Obese Population: Minimum 5-year Follow-Up},
  journal = {The Journal of Arthroplasty},
  year = {2026},
  doi = {10.1016/j.arth.2025.06.060}
}
Kostenlos starten →