Operating Room Air May Harbor Pathogens: The Role of an Ultraviolet Air Filtration Unit
The Journal of Arthroplasty · 2024
DOI: 10.1016/j.arth.2024.05.072 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Arthroplasty, 2024. Verfasst von Diana Fernández-Rodríguez und 4 weiteren Autor:innen. Beteiligte Einrichtung: Rothman Institute, Department of Orthopedic Surgery, Thomas Jefferson University, Philadelphia, Pennsylvania; Plan de Estudios Combinados en Medicina…. Hauptthemen laut MEDLINE: Operating Rooms, Air Microbiology, Ultraviolet Rays, Surgical Wound Infection. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 37 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: One important factor for the prevention of surgical site infections is ultraclean air in the operating room (OR). Still, the direct sterilization potential of most technologies, especially in a dynamic clinical setting, is not well understood. We aimed to determine and compare the microbial presence from the inlet and outlet flow of a filtration unit with crystalline ultraviolet-C (C-UVC) light. METHODS: A prospective study was conducted at a single institution, where primary total joint arthroplasty and spine surgeries were performed. The OR was fitted with a positive ventilation system. In addition, a filtration unit with a C-UVC sterilizing light was placed in the OR. The inlet and outlet air flows were swabbed simultaneously and compared. Swabs were processed for culture and next-generation sequencing. RESULTS: The mean length of the surgical procedures sampled was 68 ± 13 minutes. Overall, 19 out of 200 (9.5%) swabs isolated microorganisms. Inlet air swabs were positive at a higher rate (16 versus 3%; P < .01) compared to the outlet air swabs. A wide variety of Gram-positive, Gram-negative, and anaerobic bacteria were isolated, but fungi were only recovered from inlet air swabs. The detection of microorganisms was also higher when more door openings were performed (32.5 ± 7.1 versus 27.9 ± 5.6; P < .01). CONCLUSIONS: Air swabs mainly isolated microorganisms from the inlet flow to the filtration unit with a C-UVC light. The sterilizing unit counteracted factors affecting the air quality in the OR, namely door openings, surgical personnel, and tissue combustion.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Rothman Institute, Department of Orthopedic Surgery, Thomas Jefferson University, Philadelphia, Pennsylvania; Plan de Estudios Combinados en Medicina (PECEM) MD/PhD, Facultad de Medicina, Universidad Nacional Autónoma d…
- Department of Orthopaedic Surgery, Mayo Clinic Arizona, Phoenix, Arizona.
- Rothman Institute, Department of Orthopedic Surgery, Thomas Jefferson University, Philadelphia, Pennsylvania.
- International Joint Center, Acibadem University Hospital, Istanbul, Turkey.
Zitieren
APA
Fernández-Rodríguez, D., Tarabichi, S., Golankiewicz, K., Zappley, N., & Parvizi, J. (2024). Operating Room Air May Harbor Pathogens: The Role of an Ultraviolet Air Filtration Unit. The Journal of Arthroplasty. https://doi.org/10.1016/j.arth.2024.05.072
Vancouver
Fernández-Rodríguez D, Tarabichi S, Golankiewicz K, Zappley N, Parvizi J. Operating Room Air May Harbor Pathogens: The Role of an Ultraviolet Air Filtration Unit. The Journal of Arthroplasty. 2024. doi:10.1016/j.arth.2024.05.072
BibTeX
@article{FernndezRodrguez2024,
author = {Fernández-Rodríguez, Diana and Tarabichi, Saad and Golankiewicz, Krystal and Zappley, Nicolina and Parvizi, Javad},
title = {Operating Room Air May Harbor Pathogens: The Role of an Ultraviolet Air Filtration Unit},
journal = {The Journal of Arthroplasty},
year = {2024},
doi = {10.1016/j.arth.2024.05.072}
}
EUVIMED