An Anterior Spike Decreases Bone-Implant Micromotion in Cementless Tibial Baseplates for Total Knee Arthroplasty: A Biomechanical Study
The Journal of Arthroplasty · 2024-05 · 39 (5) · 1323-1327
DOI: 10.1016/j.arth.2023.11.020 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Arthroplasty, 2024-05. Verfasst von Fernando J Quevedo Gonzalez und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomechanics, Hospital for Special Surgery, New York. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Knee Prosthesis. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 19 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: Cementless tibial baseplates in total knee arthroplasty include fixation features (eg, pegs, spikes, and keels) to ensure sufficient primary bone-implant stability. While the design of these features plays a fundamental role in biologic fixation, the effectiveness of anterior spikes in reducing bone-implant micromotion remains unclear. Therefore, we asked: Can an anterior spike reduce the bone-implant micromotion of cementless tibial implants? METHODS: We performed computational finite element analyses on 13 tibiae using the computed tomography scans of patients scheduled for primary total knee arthroplasty. The tibiae were virtually implanted with a cementless tibial baseplate with 2 designs of fixation of the baseplate: 2 pegs and 2 pegs with an anterior spike. We compared the bone-implant micromotion under the most demanding loads from stair ascent between both designs. RESULTS: Both fixation designs had peak micromotion at the anterior-lateral edge of the baseplate. The design with 2 pegs and an anterior spike had up to 15% lower peak micromotion and up to 14% more baseplate area with micromotions below the most conservative threshold for ingrowth, 20 μm, than the design with only 2 pegs. The greatest benefit of adding an anterior spike occurred for subjects who had the smallest area of tibial bone below the 20 μm threshold (ie, most at risk for failure to achieve bone ingrowth). CONCLUSIONS: An anteriorly placed spike for cementless tibial baseplates with 2 pegs can help decrease the bone-implant micromotion during stair ascent, especially for subjects with increased bone-implant micromotion and risk for bone ingrowth failure.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomechanics, Hospital for Special Surgery, New York.
- Adult Reconstruction and Joint Replacement Service, Hospital for Special Surgery, New York.
- Department of Geriatric Science and Orthopaedics, Università Cattolica del Sacro Cuore, Rome, Italy.
Zitieren
APA
Quevedo Gonzalez, F. J., Lipman, J. D., Sculco, P. K., Sculco, T. P., De Martino, I., & Wright, T. M. (2024). An Anterior Spike Decreases Bone-Implant Micromotion in Cementless Tibial Baseplates for Total Knee Arthroplasty: A Biomechanical Study. The Journal of Arthroplasty, 39(5), 1323-1327. https://doi.org/10.1016/j.arth.2023.11.020
Vancouver
Quevedo Gonzalez FJ, Lipman JD, Sculco PK, Sculco TP, De Martino I, Wright TM. An Anterior Spike Decreases Bone-Implant Micromotion in Cementless Tibial Baseplates for Total Knee Arthroplasty: A Biomechanical Study. The Journal of Arthroplasty. 2024;39(5):1323-1327. doi:10.1016/j.arth.2023.11.020
BibTeX
@article{QuevedoGonzalez2024,
author = {Quevedo Gonzalez, Fernando J and Lipman, Joseph D and Sculco, Peter K and Sculco, Thomas P and De Martino, Ivan and Wright, Timothy M},
title = {An Anterior Spike Decreases Bone-Implant Micromotion in Cementless Tibial Baseplates for Total Knee Arthroplasty: A Biomechanical Study},
journal = {The Journal of Arthroplasty},
year = {2024},
volume = {39},
number = {5},
pages = {1323--1327},
doi = {10.1016/j.arth.2023.11.020}
}