EUVIMEDEUVIMED entdecken
Open Access

Radiographic Accuracy of Malseated Dual-Mobility Liners Varies Based on Implant Design

Richard D Rames, Cory G Couch, Rafael J Sierra, Daniel J Berry, Robert T Trousdale, Matthew P Abdel

The Journal of Arthroplasty · 2023

DOI: 10.1016/j.arth.2023.05.020 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Arthroplasty, 2023. Verfasst von Richard D Rames und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic Surgery, Mayo Clinic, Rochester, Minnesota. Hauptthemen laut MEDLINE: Hip Prosthesis, Arthroplasty, Replacement, Hip. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIAMS NIH HHS, NHLBI NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Dual-mobility constructs have gained popularity to mitigate dislocations after high-risk primary and revision total hip arthroplasties. Contemporary data have indicated that malseating of modular dual-mobility liners occurs in up to 6% of cases. The purpose of this cadaveric-based radiographic study was to determine the ability to accurately determine if modular dual-mobility liners were seated. METHODS: There were 10 hips (5 cadaveric pelvic specimens) used to implant modular dual-mobility liners of 2 designs. One had a liner that seated flush and the other had an extended rim. There were 20 constructs that were well-seated and 20 constructs were intentionally malseated. A comprehensive series of radiographs was reviewed by 2 blinded surgeons. Statistical analyses included Chi-squared testing, logistic regressions, and kappa statistics. RESULTS: Radiographic assessment of liner malseating was not accurate with an elevated rim design with misdiagnosis in 40% (16 of 40). The flush design had diagnostic errors in 5% (2 of 40; P = .0002). Logistic regressions demonstrated a significantly higher risk of misdiagnosing a malseated liner in the elevated rim group (odds ratio 13). There were 12 of 16 misdiagnoses in the elevated rim group failing to recognize a malseated liner. Each surgeon had almost perfect agreement for intraobserver reliability for flush designs (k 0.90) and fair agreement in the elevated rim design (k 0.35). CONCLUSION: A comprehensive series of plain radiographs can reliably detect a malseated modular dual-mobility liner with a flush rim design in 95% of cases. However, elevated rim designs are more difficult to accurately identify malseating on plain radiographs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hip ProsthesisArthroplasty, Replacement, HipHumansReproducibility of ResultsProsthesis DesignProsthesis FailureReoperationCadaver

Stichwörter der Autor:innen

cross-table lateral radiographsdislocationmalseatingmodular dual-mobility constructstotal hip arthroplasty

Autor:innen und Institutionen

  • Department of Orthopedic Surgery, Mayo Clinic, Rochester, Minnesota.

Förderung

  • NIAMS NIH HHS
  • NHLBI NIH HHS
  • National Institutes of Health
  • National Institute of Arthritis and Musculoskeletal and Skin Diseases

Zitieren

APA

Rames, R. D., Couch, C. G., Sierra, R. J., Berry, D. J., Trousdale, R. T., & Abdel, M. P. (2023). Radiographic Accuracy of Malseated Dual-Mobility Liners Varies Based on Implant Design. The Journal of Arthroplasty. https://doi.org/10.1016/j.arth.2023.05.020

Vancouver

Rames RD, Couch CG, Sierra RJ, Berry DJ, Trousdale RT, Abdel MP. Radiographic Accuracy of Malseated Dual-Mobility Liners Varies Based on Implant Design. The Journal of Arthroplasty. 2023. doi:10.1016/j.arth.2023.05.020

BibTeX
@article{Rames2023,
  author = {Rames, Richard D and Couch, Cory G and Sierra, Rafael J and Berry, Daniel J and Trousdale, Robert T and Abdel, Matthew P},
  title = {Radiographic Accuracy of Malseated Dual-Mobility Liners Varies Based on Implant Design},
  journal = {The Journal of Arthroplasty},
  year = {2023},
  doi = {10.1016/j.arth.2023.05.020}
}
Kostenlos starten →