Custom Triflange Acetabular Components For Catastrophic Bone Loss: Minimum 5-Year Results

Jacob S Alexander, Adolph V Lombardi, Keith R Berend, Michael J Morris, Joanne B Adams, David A Crawford

The Journal of Arthroplasty · 2023-07 · 38 (7) · 1330-1334

DOI: 10.1016/j.arth.2023.01.023 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Arthroplasty, 2023-07. Verfasst von Jacob S Alexander und 5 weiteren Autor:innen. Beteiligte Einrichtung: JIS Orthopedics, New Albany, Ohio. Hauptthemen laut MEDLINE: Hip Prosthesis, Arthroplasty, Replacement, Hip. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 20 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

INTRODUCTION: Custom acetabular components are an increasingly utilized reconstructive option in total hip arthroplasty for catastrophic bone loss and acetabular failure. The purpose of this study was to determine the survivorship of such components for reconstruction due to catastrophic bone loss at a minimum 5-year follow-up. METHODS: From August 2003 to July 2016, 64 patients (66 hips) underwent acetabular reconstruction with custom triflange components. All hips were classified as Paprosky 3B or 3C. Harris hip scores were analyzed. Overall survivorship was determined by survival analysis. RESULTS: Seventeen patients (18 hips) died prior to returning for 5-year follow-up. One presumed living patient was lost to contact, yielding a cohort of 46 patients (47 hips) who had minimum 5-year follow-up. The mean age was 65 years (range, 46 to 85), mean body mass index was 29.4 (range, 18 to 45), and 72% were women. Mean follow-up was 8 years (range, 5 to 16). There were 3 revisions of the triflange device (6.4%) due to infection. Survivorship to end point of triflange removal for any reason was 94.1% (95% confidence interval: ±3.4%) at a mean of 16 years. In the overall series (n = 66), there were 9 (13.6%) additional reoperations as follows: 5 incision and debridements, one open reduction internal fixation, two stem revisions for periprosthetic femoral fracture, and one head revision. Harris hip scores improved significantly from a mean of 41 points preoperatively to 64 points postoperatively. CONCLUSIONS: Custom acetabular triflange components represent a highly effective tool in a surgeon's armamentarium. These devices are extremely helpful in managing catastrophic bone loss and have a good mean 16-year survival.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hip ProsthesisArthroplasty, Replacement, HipHumansFemaleAgedMaleProsthesis FailureAcetabulumReoperationFollow-Up StudiesRetrospective Studies

Stichwörter der Autor:innen

THAoutcomesreconstructionrevisionsurvivaltriflange

Autor:innen und Institutionen

  • JIS Orthopedics, New Albany, Ohio.

Zitieren

APA

Alexander, J. S., Lombardi, A. V., Berend, K. R., Morris, M. J., Adams, J. B., & Crawford, D. A. (2023). Custom Triflange Acetabular Components For Catastrophic Bone Loss: Minimum 5-Year Results. The Journal of Arthroplasty, 38(7), 1330-1334. https://doi.org/10.1016/j.arth.2023.01.023

Vancouver

Alexander JS, Lombardi AV, Berend KR, Morris MJ, Adams JB, Crawford DA. Custom Triflange Acetabular Components For Catastrophic Bone Loss: Minimum 5-Year Results. The Journal of Arthroplasty. 2023;38(7):1330-1334. doi:10.1016/j.arth.2023.01.023

BibTeX
@article{Alexander2023,
  author = {Alexander, Jacob S and Lombardi, Adolph V and Berend, Keith R and Morris, Michael J and Adams, Joanne B and Crawford, David A},
  title = {Custom Triflange Acetabular Components For Catastrophic Bone Loss: Minimum 5-Year Results},
  journal = {The Journal of Arthroplasty},
  year = {2023},
  volume = {38},
  number = {7},
  pages = {1330--1334},
  doi = {10.1016/j.arth.2023.01.023}
}