The effect of crowding on the accuracy of 3-dimensional printing
American Journal of Orthodontics and Dentofacial Orthopedics · 2023
DOI: 10.1016/j.ajodo.2023.08.005 ↗Kurz gefasst
Fachartikel in der Zeitschrift American Journal of Orthodontics and Dentofacial Orthopedics, 2023. Verfasst von Tonya Spangler und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthodontics, School of Dentistry, Virginia Commonwealth University, Richmond, Va. Hauptthemen laut MEDLINE: Computer-Aided Design, Models, Dental. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben.
Abstract (Originalsprache)
INTRODUCTION: The purpose of this study was to evaluate the accuracy of 3-dimensional (3D) printed aligners compared to conventional vacuum-formed thermoplastic aligners with varying levels of dental crowding. METHODS: Digital intraoral scans of 10 cases were assigned to their respective groups (n = 10, each, 30 total) as follows: no crowding (control), moderate crowding, and severe crowding. Digital images of these models were created in standard tessellation language (STL) file format using 3Shape software and randomly 3D printed. The STL files of each case were also sent to a dental laboratory to fabricate vacuum-formed samples, the current technology used for manufacturing aligners. The intaglio surfaces of fabricated aligners in both groups were scanned using cone beam computed tomography to create STL files, which were then compared to the original STL files of the cases using Geomagic Control X software. Absolute deviations from the original file and root mean square values were recorded. A Kruskal-Wallis test was conducted to analyze the difference in average deviation, and a t-test was repeated for the RMS measure. The significance level was set at 0.05. RESULTS: The crowding did not affect the trueness of aligners manufactured using 3D printing or conventional vacuum-forming techniques (P = 0.79). 3D-printed aligners showed less deviation than the vacuum-formed samples (0.1125 mm vs 0.1312 mm; P
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Orthodontics, School of Dentistry, Virginia Commonwealth University, Richmond, Va.
- Department of General Practice, School of Dentistry, Virginia Commonwealth University, Richmond, Va.
- Department of Dental Public Health and Policy, School of Dentistry, Virginia Commonwealth University, Richmond, Va.
- Office of Oral Health Innovation, Department of Oral Rehabilitation, James B. Edwards College of Dental Medicine, Medical University of South Carolina, Charleston, SC.
- Department of Orthodontics, School of Dentistry, Virginia Commonwealth University, Richmond, Va. Electronic address: etufekci@vcu.edu.
Zitieren
APA
Spangler, T., Ammoun, R., Carrico, C. K., Bencharit, S., & Tüfekçi, E. (2023). The effect of crowding on the accuracy of 3-dimensional printing. American Journal of Orthodontics and Dentofacial Orthopedics. https://doi.org/10.1016/j.ajodo.2023.08.005
Vancouver
Spangler T, Ammoun R, Carrico CK, Bencharit S, Tüfekçi E. The effect of crowding on the accuracy of 3-dimensional printing. American Journal of Orthodontics and Dentofacial Orthopedics. 2023. doi:10.1016/j.ajodo.2023.08.005
BibTeX
@article{Spangler2023,
author = {Spangler, Tonya and Ammoun, Rami and Carrico, Caroline K and Bencharit, Sompop and Tüfekçi, Eser},
title = {The effect of crowding on the accuracy of 3-dimensional printing},
journal = {American Journal of Orthodontics and Dentofacial Orthopedics},
year = {2023},
doi = {10.1016/j.ajodo.2023.08.005}
}
EUVIMED