Continuous glucose monitor overestimates glycemia, with the magnitude of bias varying by postprandial test and individual – a randomized crossover trial

Katie M Hutchins, James A Betts, Dylan Thompson, Aaron Hengist, Javier T Gonzalez

The American Journal of Clinical Nutrition · 2025-05 · 121 (5) · 1025-1034

DOI: 10.1016/j.ajcnut.2025.02.024 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift The American Journal of Clinical Nutrition, 2025-05. Verfasst von Katie M Hutchins und 4 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Nutrition, Exercise, and Metabolism, University of Bath, Bath, United Kingdom; Department for Health, University of Bath, Bath, United Kin…. Hauptthemen laut MEDLINE: Blood Glucose, Postprandial Period, Blood Glucose Self-Monitoring. Die Arbeit wurde bisher 27-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Continuous glucose monitors (CGM) are used to characterize postprandial glycemia, yet no study has directly tested how different test foods/beverages alter CGM accuracy. OBJECTIVES: Assess glycemic responses to test foods/drinks using CGM compared with capillary sampling (criterion). METHODS: Fifteen healthy females (n = 9) and males (n = 6) completed 7 laboratory visits in a randomized crossover design with ≥48 h washout between visits. During each visit, participants consumed an oral carbohydrate challenge comprising either 50 g glucose or equivalent 50 g carbohydrate as whole fruits, 50 g carbohydrate as blended fruit, 50 g carbohydrate as commercially available fruit smoothie, 50 g carbohydrate as commercially available fruit smoothie ingested over 30 ± 4 min, 50 g carbohydrate as commercially available fruit smoothie with 5 g inulin, 30 g carbohydrate as commercially available fruit smoothie. The glycemia was recorded from both CGM and capillary samples every 15 min for 120 min and expressed as incremental areas under the curve. The glycemic index (GI) was calculated relative to 50 g glucose where appropriate. Exploratory analyses examined 1) interindividual heterogeneity of CGM bias compared with criterion and 2) whether CGM bias could be improved with adjustment for baseline differences. RESULTS: CGM-estimated fasting and postprandial glucose concentrations were (mean ± standard deviation) 0.9 ± 0.6 and 0.9 ± 0.5 mmol/L higher than capillary estimates, respectively(both, P 7.8 mmol/L by ∼4-fold, and adjustment for baseline differences reduced this overestimate to ∼2-fold (both P < 0.01). CONCLUSIONS: CGM overestimated glycemic responses in numerous contexts. At times, this can mischaracterize the GI. In addition, there is interindividual heterogeneity in the accuracy of CGM in estimating fasting glucose concentrations. Correction for this difference reduces, but does not eliminate, postprandial overestimate of glycemia by CGM. Caution should be applied when inferring absolute or relative glycemic responses to foods using CGM, and capillary sampling should be prioritized for accurate quantification of glycemic response. This trial was registered at clinicaltrials.gov as NCT06333184.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood GlucosePostprandial PeriodBlood Glucose Self-MonitoringHumansFemaleMaleCross-Over StudiesAdultGlycemic IndexYoung AdultDietary CarbohydratesFruitMiddle Aged

Stichwörter der Autor:innen

capillarycarbohydratefruit smoothieglycemic indexglycemic responseinterstitialmetabolism

Substanzen

Blood GlucoseDietary Carbohydrates

Autor:innen und Institutionen

  • Centre for Nutrition, Exercise, and Metabolism, University of Bath, Bath, United Kingdom; Department for Health, University of Bath, Bath, United Kingdom.
  • National Institute of Diabetes and Digestive and Kidney Diseases, NIH, Bethesda, MD, United States.
  • Centre for Nutrition, Exercise, and Metabolism, University of Bath, Bath, United Kingdom; Department for Health, University of Bath, Bath, United Kingdom. Electronic address: j.t.gonzalez@bath.ac.uk.

Zitieren

APA

Hutchins, K. M., Betts, J. A., Thompson, D., Hengist, A., & Gonzalez, J. T. (2025). Continuous glucose monitor overestimates glycemia, with the magnitude of bias varying by postprandial test and individual – a randomized crossover trial. The American Journal of Clinical Nutrition, 121(5), 1025-1034. https://doi.org/10.1016/j.ajcnut.2025.02.024

Vancouver

Hutchins KM, Betts JA, Thompson D, Hengist A, Gonzalez JT. Continuous glucose monitor overestimates glycemia, with the magnitude of bias varying by postprandial test and individual – a randomized crossover trial. The American Journal of Clinical Nutrition. 2025;121(5):1025-1034. doi:10.1016/j.ajcnut.2025.02.024

BibTeX
@article{Hutchins2025,
  author = {Hutchins, Katie M and Betts, James A and Thompson, Dylan and Hengist, Aaron and Gonzalez, Javier T},
  title = {Continuous glucose monitor overestimates glycemia, with the magnitude of bias varying by postprandial test and individual – a randomized crossover trial},
  journal = {The American Journal of Clinical Nutrition},
  year = {2025},
  volume = {121},
  number = {5},
  pages = {1025--1034},
  doi = {10.1016/j.ajcnut.2025.02.024}
}
Kostenlos starten →