Assessment of remote digital auscultation for identifying significant aortic stenosis: A pilot study
Archives of Cardiovascular Diseases · 2025
DOI: 10.1016/j.acvd.2025.06.072 ↗Kurz gefasst
Fachartikel in der Zeitschrift Archives of Cardiovascular Diseases, 2025. Verfasst von Yuval Avidan und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Lady Davis Carmel Medical Center, 7 Michal St., Haifa, Israel. Electronic address: Yuvalavidan10@gmail.com. Hauptthemen laut MEDLINE: Aortic Valve Stenosis, Heart Auscultation, Stethoscopes, Aortic Valve. Die Arbeit nennt 28 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: Aortic stenosis (AS) is the most common valvular disease in the Western world. Technological advances, such as the TytoCare™ device, have introduced remote telemedicine-based cardiac auscultation. Its utility in diagnosing and grading AS remains unexplored. This study aims to evaluate the reliability of the TytoCare™ for assessing AS. METHODS: Sixty patients were classified by standard echocardiogram into three groups: significant AS (at least moderate AS; n=20), non-significant AS (aortic sclerosis or mild AS; n=20) and controls (n=20). Heart sounds were self-recorded using the TytoCare™ stethoscope. Fifteen blinded physicians - five each from primary care, internal medicine and cardiology - reviewed the recordings using standardized equipment and assigned diagnoses based on a categorical scale: 0=insufficient data, 1=normal, 2=mild AS and 3=at least moderate AS. RESULTS: The study groups had comparable baseline characteristics, with a mean±standard deviation age of 72.6±9.1years. The correct response rate for significant AS was 85% (95% confidence interval [95% CI]: 80-88%), with both cardiologists and non-cardiologists achieving≥80%. In contrast, performance declined in non-significant AS (57%, 95% CI: 51-62%), but remained high in the control group (87%, 95% CI: 82-90%). Cardiologists outperformed other physicians, overall (P=0.029) and in significant AS cases (P=0.022), while no significant differences were found among non-cardiologists. CONCLUSIONS: TytoCare™ demonstrated high reliability in the remote assessment of significant AS by both cardiologists and non-cardiologists. Its intuitive and user-friendly design enables self-application, even among older populations, emphasizing its potential as an effective screening tool for AS across diverse healthcare settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Lady Davis Carmel Medical Center, 7 Michal St., Haifa, Israel. Electronic address: Yuvalavidan10@gmail.com.
- Department of Cardiology, Lady Davis Carmel Medical Center, 7 Michal St., Haifa, Israel.
- Department of Cardiology, Lady Davis Carmel Medical Center, 7 Michal St., Haifa, Israel; The Ruth and Bruce Rappaport Faculty of Medicine, Technion, Israel Institute of Technology, Haifa, Israel.
Zitieren
APA
Avidan, Y., Tabachnikov, V., & Aker, A. (2025). Assessment of remote digital auscultation for identifying significant aortic stenosis: A pilot study. Archives of Cardiovascular Diseases. https://doi.org/10.1016/j.acvd.2025.06.072
Vancouver
Avidan Y, Tabachnikov V, Aker A. Assessment of remote digital auscultation for identifying significant aortic stenosis: A pilot study. Archives of Cardiovascular Diseases. 2025. doi:10.1016/j.acvd.2025.06.072
BibTeX
@article{Avidan2025,
author = {Avidan, Yuval and Tabachnikov, Vsevolod and Aker, Amir},
title = {Assessment of remote digital auscultation for identifying significant aortic stenosis: A pilot study},
journal = {Archives of Cardiovascular Diseases},
year = {2025},
doi = {10.1016/j.acvd.2025.06.072}
}
EUVIMED