Protein-metal interactions due to fretting corrosion at the taper junction of hip implants: An in vitro investigation using Raman spectroscopy
Acta Biomaterialia · 2024
DOI: 10.1016/j.actbio.2024.10.006 ↗Kurz gefasst
Fachartikel in der Zeitschrift Acta Biomaterialia, 2024. Verfasst von Adrian Wittrock und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, TU Dortmund University, Otto-Hahn-Str. 4a, Dortmund, 44227, Germany. Hauptthemen laut MEDLINE: Spectrum Analysis, Raman, Hip Prosthesis, Alloys, Titanium. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 94 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Modular hip implants are a clinically successful and widely used treatment for patients with arthritis. Despite ongoing retrieval studies the understanding of the fundamental physico-chemical mechanisms of friction and wear within the head-taper interface is still limited. Here, we Raman-spectroscopically analyze structural features of the biotribological material which is formed within the taper joint between Ti6Al4V and low-carbon cobalt alloy or high-nitrogen steel surfaces in in vitro gross-slip fretting corrosion tests with bovine calf serum. As a function of the fretting duration, we investigate short and long aliphatic chains and their adsorption behavior on the cobalt- and steel-type surfaces. Using the intensity and frequency shifts of the amide I and III Raman bands, we furthermore identify progressive protein folding and unfolding including the secondary structures of α-helix, β-sheet, and random-coil configuration as well as the formation of proteinaceous clusters depending on the hydrophilicity of the metallic surfaces. We additionally find a mixture of chromates and iron oxides with tryptophan and tyrosine at the worn cobalt alloy and high-nitrogen steel surfaces, respectively. Also, for long fretting duration, sp2 hybridized amorphous carbon is formed due to fretting-induced cleavage of proteins. STATEMENT OF SIGNIFICANCE: Despite efforts enhancing the biomedical tribology of hip implants, the impact of the organic environment on friction and wear at the femoral head-stem taper interface is limitedly understood. Using Raman spectroscopy we resolve structural changes within the biotribological material agglomerated at biomedical-grade metal alloys due to metal-organic interactions during in vitro fretting corrosion tests. Adsorption of short and long aliphatic chains, progressive protein (un)folding and proteinaceous cluster formation depend to a distinguishable extent on the fretting duration and type of alloy. Chromates and iron oxides are mixed with tryptophan and tyrosine, and amorphous carbon is formed resulting from a fretting-induced cleavage of serum proteins. Such information spectroscopically gleaned from biotribological material are vital to improve the design and performance of taper junctions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Physics, TU Dortmund University, Otto-Hahn-Str. 4a, Dortmund, 44227, Germany.
- Department of Orthopedic Surgery, Rush University Medical Center, 1611 W. Harrison St., Chicago, 60612, IL, USA.
- Department of Orthopedic Surgery, Rush University Medical Center, 1611 W. Harrison St., Chicago, 60612, IL, USA; Department Microstructure Physics and Alloy Design, Max Planck Institute for Sustainable Materials, Max-Pl…
- Department of Physics, TU Dortmund University, Otto-Hahn-Str. 4a, Dortmund, 44227, Germany. Electronic address: joerg.debus@tu-dortmund.de.
Zitieren
APA
Wittrock, A., Heermant, S., Beckmann, C., Wimmer, M. A., Fischer, A., Aßmann, M., & Debus, J. (2024). Protein-metal interactions due to fretting corrosion at the taper junction of hip implants: An in vitro investigation using Raman spectroscopy. Acta Biomaterialia. https://doi.org/10.1016/j.actbio.2024.10.006
Vancouver
Wittrock A, Heermant S, Beckmann C, Wimmer MA, Fischer A, Aßmann M, et al. Protein-metal interactions due to fretting corrosion at the taper junction of hip implants: An in vitro investigation using Raman spectroscopy. Acta Biomaterialia. 2024. doi:10.1016/j.actbio.2024.10.006
BibTeX
@article{Wittrock2024,
author = {Wittrock, Adrian and Heermant, Saskia and Beckmann, Christian and Wimmer, Markus A and Fischer, Alfons and Aßmann, Marc and Debus, Jörg},
title = {Protein-metal interactions due to fretting corrosion at the taper junction of hip implants: An in vitro investigation using Raman spectroscopy},
journal = {Acta Biomaterialia},
year = {2024},
doi = {10.1016/j.actbio.2024.10.006}
}
EUVIMED