Reactive oxygen species, electrode potential and pH affect CoCrMo alloy corrosion and semiconducting behavior in simulated inflammatory environments

Hwaran Lee, Michael A Kurtz, Jeremy L Gilbert

Acta Biomaterialia · 2024

DOI: 10.1016/j.actbio.2024.08.010 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta Biomaterialia, 2024. Verfasst von Hwaran Lee und 2 weiteren Autor:innen. Beteiligte Einrichtung: Clemson - Medical University of South Carolina Bioengineering Program, Department of Bioengineering, Clemson University, Bioengineering Building, 101…. Hauptthemen laut MEDLINE: Reactive Oxygen Species, Electrodes, Vitallium, Inflammation. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 55 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Crevice corrosion in modular taper junctions of hip or knee replacements using cobalt-chrome-molybdenum (CoCrMo) alloys remains a clinical concern. Non-mechanically-driven corrosion has been less explored compared to mechanically assisted crevice corrosion. This study hypothesized that solution chemistry within crevices, inflammation, and cathodic electrode potential shifts during fretting result in low pH and generate reactive oxygen species (ROS), affecting oxide film behavior. This study investigated how resistance and capacitance of the CoCrMo oxide film (i.e., corrosion resistance) are modified in simulated in vivo crevice environments of modular taper junctions. Six solutions were evaluated (two pH levels: 1 and 7.4 and four hydrogen peroxide (H2O2) concentrations: 0, 0.001, 0.01 and 0.1 M). Rp versus voltage and Mott-Schottky plots were created from symmetry-based electrochemical impedance spectroscopy (sbEIS). At pH 1, the semiconductor transition to p-type occurs at more anodic potentials and higher flat band potentials were found. H2O2 decreased the flat band potential and slope in the Mott-Schottky plot. Higher H2O2 in pH 7.4 solution significantly modified the oxide film, leading to increased donor density (p = 0.0004) and a 150-fold reduction in Rp in the cathodic potential range at -1 V (p = 0.0005). The most unfavorable condition (0.1 M H2O2 pH 1) resulted in a 250-fold lower resistance compared to phosphate buffered saline (PBS) pH 7.4 at -1 V (p = 0.0013). This study highlights the corrosion susceptibility of CoCrMo under adverse chemical and potential conditions, identifying increased defects in the oxide film due to ROS, hydrogen ions and electrode potential. STATEMENT OF SIGNIFICANCE: Corrosion of cobalt chrome molybdenum alloy caused by direct chemical attack in the crevice region of hip replacements, such as modular taper junctions, remains a clinical concern. The junction environment contains adverse chemical compositions, including high acidity and reactive oxygen species (ROS) due to inflammatory responses against the corrosion products. We simulate inflammatory environments with different pH levels and hydrogen peroxide, representative of ROS. We employ electrochemical impedance spectroscopy and apply stepwise voltage over the range induced by tribocorrosion processes. We relate the effect of adverse chemical components on corrosion and semiconducting behavior of the oxide film using Mott-Schottky analysis. This study shows how pH and ROS concentration compromises the oxide film potentially leading to non-mechanically induced corrosion.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Reactive Oxygen SpeciesElectrodesVitalliumInflammationSemiconductorsCorrosionHydrogen-Ion ConcentrationHydrogen PeroxideAlloysDielectric Spectroscopy

Stichwörter der Autor:innen

CoCrMoElectrochemical impedance spectroscopy (EIS)Hydrogen peroxide (H(2)O(2))Mott–Schottky analysisPassive oxide filmReactive oxygen species (ROS)

Substanzen

Reactive Oxygen SpeciesVitalliumHydrogen PeroxideAlloys

Autor:innen und Institutionen

  • Clemson - Medical University of South Carolina Bioengineering Program, Department of Bioengineering, Clemson University, Bioengineering Building, 101D, MSC 501, 68 Presidents St, BE 325, Medical University of South Caro…

Zitieren

APA

Lee, H., Kurtz, M. A., & Gilbert, J. L. (2024). Reactive oxygen species, electrode potential and pH affect CoCrMo alloy corrosion and semiconducting behavior in simulated inflammatory environments. Acta Biomaterialia. https://doi.org/10.1016/j.actbio.2024.08.010

Vancouver

Lee H, Kurtz MA, Gilbert JL. Reactive oxygen species, electrode potential and pH affect CoCrMo alloy corrosion and semiconducting behavior in simulated inflammatory environments. Acta Biomaterialia. 2024. doi:10.1016/j.actbio.2024.08.010

BibTeX
@article{Lee2024,
  author = {Lee, Hwaran and Kurtz, Michael A and Gilbert, Jeremy L},
  title = {Reactive oxygen species, electrode potential and pH affect CoCrMo alloy corrosion and semiconducting behavior in simulated inflammatory environments},
  journal = {Acta Biomaterialia},
  year = {2024},
  doi = {10.1016/j.actbio.2024.08.010}
}
Kostenlos starten →