Validation of Automated 3D MRI Knee Cartilage Measurements
Academic Radiology · 2026-10
DOI: 10.1016/j.acra.2026.09.025 ↗Kurz gefasst
Fachartikel in der Zeitschrift Academic Radiology, 2026-10. Verfasst von Mirela Kostadinova und 13 weiteren Autor:innen. Beteiligte Einrichtung: Clinic for Radiology and Nuclear Medicine, University Hospital Frankfurt, Goethe University, Frankfurt am Main, Germany (M.K., I.H., S.M. S.B., E.H.,…. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
RATIONALE AND OBJECTIVES: To assess the agreement between automated cartilage thickness measurements derived from three-dimensional (3D) dual-echo steady-state (DESS) magnetic resonance imaging and manual radiologist measurements. MATERIALS AND METHODS: In 111 patients, cartilage thickness measurements across 21 knee subregions were compared between an automated segmentation software and manual radiologist measurements on 3D DESS and proton density (PD) fat-suppressed sequences. Agreement was assessed using Bland-Altman analysis, Pearson correlation, and intraclass correlation coefficient (ICC). A post-hoc calibration approach was applied to both comparisons. Intersequence variability between PD manual and DESS manual measurements served as a benchmark. RESULTS: The automated method showed a systematic overestimation relative to the PD manual (bias +0.7 mm) and the DESS manual (bias +0.9 mm). After calibration, bias was eliminated. Postcalibration ICC improved substantially (PD: from 0.29-0.61; DESS: from 0.18-0.67), reaching levels comparable to the intersequence benchmark (ICC = 0.73). Per-region analysis revealed the greatest variability in patellar and lateral tibial subregions. CONCLUSION: Following a simple global calibration, automated cartilage thickness measurements achieve overall acceptable agreement with manual reference measurements that is comparable to inherent intersequence variability. This supports the clinical acceptability of the automated approach as an efficient and reproducible alternative to manual segmentation.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Clinic for Radiology and Nuclear Medicine, University Hospital Frankfurt, Goethe University, Frankfurt am Main, Germany (M.K., I.H., S.M. S.B., E.H., P.R., A.G., L.S.A., L.D.G., C.B., K.E., T.J.V., and I.Y. ). Electroni…
- Clinic for Radiology and Nuclear Medicine, University Hospital Frankfurt, Goethe University, Frankfurt am Main, Germany (M.K., I.H., S.M. S.B., E.H., P.R., A.G., L.S.A., L.D.G., C.B., K.E., T.J.V., and I.Y. ).
- Diagnostic and Interventional Radiology Unit, BIOMORF Department, University Hospital "Policlinico G. Martino", Messina, Italy (T.D.).
Zitieren
APA
Kostadinova, M., Harrou, I., Mahmoudi, S., Bernatz, S., Hoehne, E., Reschke, P., Goekduman, A., Alizadeh, L. S., Gruenewald, L. D., D'Angelo, T., Booz, C., Eichler, K., Vogl, T. J., & Yel, I. (2026). Validation of Automated 3D MRI Knee Cartilage Measurements. Academic Radiology. https://doi.org/10.1016/j.acra.2026.09.025
Vancouver
Kostadinova M, Harrou I, Mahmoudi S, Bernatz S, Hoehne E, Reschke P, et al. Validation of Automated 3D MRI Knee Cartilage Measurements. Academic Radiology. 2026. doi:10.1016/j.acra.2026.09.025
BibTeX
@article{Kostadinova2026,
author = {Kostadinova, Mirela and Harrou, Iman and Mahmoudi, Scherwin and Bernatz, Simon and Hoehne, Elena and Reschke, Philipp and Goekduman, Aynur and Alizadeh, Leona S and Gruenewald, Leon D and D'Angelo, Tommaso and Booz, Christian and Eichler, Katrin and Vogl, Thomas J and Yel, Ibrahim},
title = {Validation of Automated 3D MRI Knee Cartilage Measurements},
journal = {Academic Radiology},
year = {2026},
doi = {10.1016/j.acra.2026.09.025}
}
EUVIMED