Validation of Automated 3D MRI Knee Cartilage Measurements

Mirela Kostadinova, Iman Harrou, Scherwin Mahmoudi, Simon Bernatz, Elena Hoehne, Philipp Reschke, Aynur Goekduman, Leona S Alizadeh, Leon D Gruenewald, Tommaso D'Angelo, Christian Booz, Katrin Eichler und 2 weitere

Academic Radiology · 2026-10

DOI: 10.1016/j.acra.2026.09.025 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Academic Radiology, 2026-10. Verfasst von Mirela Kostadinova und 13 weiteren Autor:innen. Beteiligte Einrichtung: Clinic for Radiology and Nuclear Medicine, University Hospital Frankfurt, Goethe University, Frankfurt am Main, Germany (M.K., I.H., S.M. S.B., E.H.,…. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

RATIONALE AND OBJECTIVES: To assess the agreement between automated cartilage thickness measurements derived from three-dimensional (3D) dual-echo steady-state (DESS) magnetic resonance imaging and manual radiologist measurements. MATERIALS AND METHODS: In 111 patients, cartilage thickness measurements across 21 knee subregions were compared between an automated segmentation software and manual radiologist measurements on 3D DESS and proton density (PD) fat-suppressed sequences. Agreement was assessed using Bland-Altman analysis, Pearson correlation, and intraclass correlation coefficient (ICC). A post-hoc calibration approach was applied to both comparisons. Intersequence variability between PD manual and DESS manual measurements served as a benchmark. RESULTS: The automated method showed a systematic overestimation relative to the PD manual (bias +0.7 mm) and the DESS manual (bias +0.9 mm). After calibration, bias was eliminated. Postcalibration ICC improved substantially (PD: from 0.29-0.61; DESS: from 0.18-0.67), reaching levels comparable to the intersequence benchmark (ICC = 0.73). Per-region analysis revealed the greatest variability in patellar and lateral tibial subregions. CONCLUSION: Following a simple global calibration, automated cartilage thickness measurements achieve overall acceptable agreement with manual reference measurements that is comparable to inherent intersequence variability. This supports the clinical acceptability of the automated approach as an efficient and reproducible alternative to manual segmentation.

Schlagwörter

Stichwörter der Autor:innen

3D DESSAutomated segmentationCartilage thicknessKnee MRIQuantitative imaging

Autor:innen und Institutionen

  • Clinic for Radiology and Nuclear Medicine, University Hospital Frankfurt, Goethe University, Frankfurt am Main, Germany (M.K., I.H., S.M. S.B., E.H., P.R., A.G., L.S.A., L.D.G., C.B., K.E., T.J.V., and I.Y. ). Electroni…
  • Clinic for Radiology and Nuclear Medicine, University Hospital Frankfurt, Goethe University, Frankfurt am Main, Germany (M.K., I.H., S.M. S.B., E.H., P.R., A.G., L.S.A., L.D.G., C.B., K.E., T.J.V., and I.Y. ).
  • Diagnostic and Interventional Radiology Unit, BIOMORF Department, University Hospital "Policlinico G. Martino", Messina, Italy (T.D.).

Zitieren

APA

Kostadinova, M., Harrou, I., Mahmoudi, S., Bernatz, S., Hoehne, E., Reschke, P., Goekduman, A., Alizadeh, L. S., Gruenewald, L. D., D'Angelo, T., Booz, C., Eichler, K., Vogl, T. J., & Yel, I. (2026). Validation of Automated 3D MRI Knee Cartilage Measurements. Academic Radiology. https://doi.org/10.1016/j.acra.2026.09.025

Vancouver

Kostadinova M, Harrou I, Mahmoudi S, Bernatz S, Hoehne E, Reschke P, et al. Validation of Automated 3D MRI Knee Cartilage Measurements. Academic Radiology. 2026. doi:10.1016/j.acra.2026.09.025

BibTeX
@article{Kostadinova2026,
  author = {Kostadinova, Mirela and Harrou, Iman and Mahmoudi, Scherwin and Bernatz, Simon and Hoehne, Elena and Reschke, Philipp and Goekduman, Aynur and Alizadeh, Leona S and Gruenewald, Leon D and D'Angelo, Tommaso and Booz, Christian and Eichler, Katrin and Vogl, Thomas J and Yel, Ibrahim},
  title = {Validation of Automated 3D MRI Knee Cartilage Measurements},
  journal = {Academic Radiology},
  year = {2026},
  doi = {10.1016/j.acra.2026.09.025}
}
Kostenlos starten →