CCTA without ECG Leads Using Whole-Heart Motion Correction Algorithm Based on Wide-Detector CT

Cheng Zhao, Zhongying Wang, Shuo Dong, Jinfu Bu, Chong Zheng, Wang Chao, Jie Lu

Academic Radiology · 2026-10 · 33 (10) · 4397-4408

DOI: 10.1016/j.acra.2026.06.054 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Academic Radiology, 2026-10. Verfasst von Cheng Zhao und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Nuclear Medicine, Xuanwu Hospital, Capital Medical University, 100053 Beijing, China (C.Z., C.Z., W.C., J.L.). Hauptthemen laut MEDLINE: Algorithms, Computed Tomography Angiography, Coronary Angiography, Radiographic Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Gefördert von: Beijing Natural Science Foundation, Hebei Provincial Department of Science and Technology, Natural Science Foundation of Beijing Municipality.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: This study evaluates the impact of a whole heart motion correction algorithm on image quality and interpretability in coronary computed tomography angiography (CCTA) without ECG leads. MATERIALS AND METHODS: Seventy patients with acute chest pain underwent CCTA without ECG leads examination. Images were reconstructed with standard (STD), coronary vessel tracking motion correction (Snapshot Freeze 1, SSF1), and whole‑heart motion correction (Snapshot Freeze 2, SSF2) algorithms. Additionally, all three groups underwent deep learning image reconstruction (DLIR) algorithm. Objective evaluation, including noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured. Subjective image quality was scored by two radiologists using a 5 - point Likert scale. Interpretability of coronary vessels was assessed per segment, vessel, and patient. RESULTS: SSF2 significantly reduced noise and improved SNR/CNR compared to STD and SSF1 (all P < .05). Subjective scores were highest for SSF2 (median 5.0 vs 4.0 for SSF1 and 3.0 for STD, P < .001), with inter-observer agreement (κ = 0.87). SSF2 achieved higher interpretability for proximal/mid-segments (99.6%) and improved distal/small-branch interpretability (97.2% vs 89.4% for SSF1, P < .017). Patient-level interpretability reached 100% with SSF2, vs 92.9% (SSF1) and 51.4% (STD). CONCLUSION: SSF2 enhances image quality and interpretability in CCTA without ECG leads, and exhibits favorable in the visualization of distal vessels and small branches.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsComputed Tomography AngiographyCoronary AngiographyRadiographic Image Interpretation, Computer-AssistedHeartHumansFemaleElectrocardiographyMaleSignal-To-Noise RatioMiddle AgedAgedArtifactsReproducibility of Results

Stichwörter der Autor:innen

CCTA without ECG leadsImage quality assessmentMotion correctionWide-detector CT

Autor:innen und Institutionen

  • Department of Radiology and Nuclear Medicine, Xuanwu Hospital, Capital Medical University, 100053 Beijing, China (C.Z., C.Z., W.C., J.L.).
  • Department of Radiology, Xiongan Xuanwu Hospital, 070001 Xiongan New Area, China (Z.W., J.B.).
  • Department of Medical Engineering, Xuanwu Hospital, Capital Medical University, 100053 Beijing, China (S.D.); Department of Medical Engineering, Xiongan Xuanwu Hospital, 070001 Xiongan New Area, China (S.D.).
  • Department of Radiology and Nuclear Medicine, Xuanwu Hospital, Capital Medical University, 100053 Beijing, China (C.Z., C.Z., W.C., J.L.). Electronic address: imaginglu@hotmail.com.

Förderung

  • Beijing Natural Science Foundation
  • Hebei Provincial Department of Science and Technology
  • Natural Science Foundation of Beijing Municipality

Zitieren

APA

Zhao, C., Wang, Z., Dong, S., Bu, J., Zheng, C., Chao, W., & Lu, J. (2026). CCTA without ECG Leads Using Whole-Heart Motion Correction Algorithm Based on Wide-Detector CT. Academic Radiology, 33(10), 4397-4408. https://doi.org/10.1016/j.acra.2026.06.054

Vancouver

Zhao C, Wang Z, Dong S, Bu J, Zheng C, Chao W, et al. CCTA without ECG Leads Using Whole-Heart Motion Correction Algorithm Based on Wide-Detector CT. Academic Radiology. 2026;33(10):4397-4408. doi:10.1016/j.acra.2026.06.054

BibTeX
@article{Zhao2026,
  author = {Zhao, Cheng and Wang, Zhongying and Dong, Shuo and Bu, Jinfu and Zheng, Chong and Chao, Wang and Lu, Jie},
  title = {CCTA without ECG Leads Using Whole-Heart Motion Correction Algorithm Based on Wide-Detector CT},
  journal = {Academic Radiology},
  year = {2026},
  volume = {33},
  number = {10},
  pages = {4397--4408},
  doi = {10.1016/j.acra.2026.06.054}
}
Kostenlos starten →