Fachartikel in der Zeitschrift Academic Radiology, 2024-02. Verfasst von Mi-Ri Kwon und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, 29 Saemunan-ro, Jongno-gu, Seoul, 03181, Republic of K…. Hauptthemen laut MEDLINE: Neoplasms, Breast Neoplasms. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 42 Literaturangaben.
Abstract (Originalsprache)
RATIONALE AND OBJECTIVES: This study aimed to evaluate the diagnostic performance of radiologists following the utilization of artificial intelligence (AI)-based computer-aided detection software (CAD) in detecting suspicious lesions in automated breast ultrasounds (ABUS). MATERIALS AND METHODS: ABUS-detected 262 breast lesions (histopathological verification; January 2020 to December 2022) were included. Two radiologists reviewed the images and assigned a Breast Imaging Reporting and Data System (BI-RADS) category. ABUS images were classified as positive or negative using AI-CAD. The BI-RADS category was readjusted in four ways: the radiologists modified the BI-RADS category using the AI results (AI-aided 1), upgraded or downgraded based on AI results (AI-aided 2), only upgraded for positive results (AI-aided 3), or only downgraded for negative results (AI-aided 4). The AI-aided diagnostic performances were compared to radiologists. The AI-CAD-positive and AI-CAD-negative cancer characteristics were compared. RESULTS: For 262 lesions (145 malignant and 117 benign) in 231 women (mean age, 52.2 years), the area under the receiver operator characteristic curve (AUC) of radiologists was 0.870 (95% confidence interval [CI], 0.832-0.908). The AUC significantly improved to 0.919 (95% CI, 0.890-0.947; P = 0.001) using AI-aided 1, whereas it improved without significance to 0.884 (95% CI, 0.844-0.923), 0.890 (95% CI, 0.852-0.929), and 0.890 (95% CI, 0.853-0.928) using AI-aided 2, 3, and 4, respectively. AI-CAD-negative cancers were smaller, less frequently exhibited retraction phenomenon, and had lower BI-RADS category. Among nonmass lesions, AI-CAD-negative cancers showed no posterior shadowing. CONCLUSION: AI-CAD implementation significantly improved the radiologists' diagnostic performance and may serve as a valuable diagnostic tool.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, 29 Saemunan-ro, Jongno-gu, Seoul, 03181, Republic of Korea (M.K., I.Y., H.-A.L.).
- Department of Radiology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, 29 Saemunan-ro, Jongno-gu, Seoul, 03181, Republic of Korea (M.K., I.Y., H.-A.L.). Electronic address: iy.youn@samsung.com.
- Division of Biostatistics, Department of R&D Management, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea (M.Y.L.).
Zitieren
APA
Kwon, M. R., Youn, I., Lee, M. Y., & Lee, H. A. (2024). Diagnostic Performance of Artificial Intelligence–Based Computer-Aided Detection Software for Automated Breast Ultrasound. Academic Radiology, 31(2), 480-491. https://doi.org/10.1016/j.acra.2023.09.013
Vancouver
Kwon MR, Youn I, Lee MY, Lee HA. Diagnostic Performance of Artificial Intelligence–Based Computer-Aided Detection Software for Automated Breast Ultrasound. Academic Radiology. 2024;31(2):480-491. doi:10.1016/j.acra.2023.09.013
BibTeX
@article{Kwon2024,
author = {Kwon, Mi-Ri and Youn, Inyoung and Lee, Mi Yeon and Lee, Hyun-Ah},
title = {Diagnostic Performance of Artificial Intelligence–Based Computer-Aided Detection Software for Automated Breast Ultrasound},
journal = {Academic Radiology},
year = {2024},
volume = {31},
number = {2},
pages = {480--491},
doi = {10.1016/j.acra.2023.09.013}
}
EUVIMED