Liver Iodine Quantification With Photon-Counting Detector CT: Accuracy in an Abdominal Phantom and Feasibility in Patients

Thomas Sartoretti, Victor Mergen, Lisa Jungblut, Hatem Alkadhi, André Euler

Academic Radiology · 2023-03 · 30 (3) · 461-469

DOI: 10.1016/j.acra.2022.04.021 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Academic Radiology, 2023-03. Verfasst von Thomas Sartoretti und 4 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Raemistrasse 100, CH-809, Zurich, Switzerland. Hauptthemen laut MEDLINE: Iodine. Die Arbeit wurde bisher 30-mal zitiert (Crossref) und nennt 31 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

RATIONALE AND OBJECTIVES: To compare the accuracy of iodine quantification in liver parenchyma and lesions between dual-source photon-counting detector CT (PCD-CT) and dual-source energy-integrating detector CT (EID-CT) in a phantom and to demonstrate the feasibility of iodine quantification with PCD-CT in liver parenchyma and lesions in patients. MATERIALS AND METHODS: An anthropomorphic abdominal phantom with a liver insert containing parenchyma and lesions was imaged on a clinical PCD-CT at 120kV and in the dual-energy mode on an EID-CT with kV-combinations of 80/Sn150kV, 90/Sn150kV, and 100/Sn150kV. Three patient sizes were imaged at three different radiation doses (CTDIvol: 5, 10, 15mGy). Thirty patients with liver cysts, hemangiomas or metastases imaged with PCD-CT were retrospectively included. Iodine maps were reconstructed and iodine concentrations were measured in liver parenchyma and lesions. For the phantom, iodine error was quantified as the absolute difference to the vendor's specifications as reference. RESULTS: Overall iodine error was 0.33 ± 0.29, 0.34 ± 0.32, 0.39 ± 0.37, 0.35 ± 0.39 mgI/mL for 80/Sn150kV, 90/Sn150kV, 100/Sn150kV of EID-CT, and PCD-CT, respectively, without significant differences between PCD-CT and EID-CT (p > 0.05). Radiation dose did not significantly influence error of PCD-CT (p > 0.05) nor EID-CT (p > 0.05). For both scanners, smaller patient sizes were associated with lower errors (p 0.05). In patients, iodine quantification was feasible in liver parenchyma, cysts, hemangiomas, and metastases. CONCLUSION: Iodine quantification with PCD-CT is accurate in simulated liver parenchyma and lesions irrespective of radiation dose, iodine concentration, and base attenuation and is feasible in common liver lesions in patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

IodineHumansFeasibility StudiesRetrospective StudiesPhotonsTomography, X-Ray ComputedPhantoms, ImagingLiver

Stichwörter der Autor:innen

Photon-counting detector CTiodine quantificationspectral imaging

Substanzen

Iodine

Autor:innen und Institutionen

  • Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Raemistrasse 100, CH-809, Zurich, Switzerland.
  • Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Raemistrasse 100, CH-809, Zurich, Switzerland. Electronic address: andre.euler@usz.ch.

Zitieren

APA

Sartoretti, T., Mergen, V., Jungblut, L., Alkadhi, H., & Euler, A. (2023). Liver Iodine Quantification With Photon-Counting Detector CT: Accuracy in an Abdominal Phantom and Feasibility in Patients. Academic Radiology, 30(3), 461-469. https://doi.org/10.1016/j.acra.2022.04.021

Vancouver

Sartoretti T, Mergen V, Jungblut L, Alkadhi H, Euler A. Liver Iodine Quantification With Photon-Counting Detector CT: Accuracy in an Abdominal Phantom and Feasibility in Patients. Academic Radiology. 2023;30(3):461-469. doi:10.1016/j.acra.2022.04.021

BibTeX
@article{Sartoretti2023,
  author = {Sartoretti, Thomas and Mergen, Victor and Jungblut, Lisa and Alkadhi, Hatem and Euler, André},
  title = {Liver Iodine Quantification With Photon-Counting Detector CT: Accuracy in an Abdominal Phantom and Feasibility in Patients},
  journal = {Academic Radiology},
  year = {2023},
  volume = {30},
  number = {3},
  pages = {461--469},
  doi = {10.1016/j.acra.2022.04.021}
}
Kostenlos starten →