EUVIMEDEUVIMED entdecken
Crossref

Calibration-free and drift-robust AlGaN/GaN HEMT sensor arrays for intelligent pH detection

Jiang Zhu, Heqiu Zhang, Dawei Guo, Guanbing Ji, Jialin Wang, Shupeng Sun, Hongwei Liang

Analytica Chimica Acta · 2026-07 · 1408 · 345536

DOI: 10.1016/j.aca.2026.345536 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytica Chimica Acta, 2026-07. Verfasst von Jiang Zhu und 6 weiteren Autor:innen. Beteiligte Einrichtung: School of Integrated Circuits, Dalian University of Technology, Dalian, 116620, PR China; Dalian Key Laboratory of Wide Bandgap Semiconductor Devices…. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

AlGaN/GaN high-electron-mobility transistors (HEMTs) combine surface charge coupling with excellent chemical stability, making them highly attractive for liquid-phase biosensing. However, when extended to array-level implementations, sensor-to-sensor variation, chip-to-chip inconsistency, and response variation across repeated measurements can collectively undermine reproducibility. In this work, we report a hardware-algorithm co-optimized sensing platform that integrates a monolithically embedded on-chip reference electrode and multilayer metallization to ensure electrical isolation and matrix-addressed signal acquisition. To address array-level variability, a data-driven artificial neural network (ANN) was employed, improving robustness across sensors, chips, and repeated measurements without explicit recalibration. The model showed robust recognition performance across different chips and measurement conditions, with stable within-chip recognition and effective cross-chip transfer, achieving a classification accuracy of approximately 95% when trained on the combined datasets. X-ray photoelectron spectroscopy and Kelvin probe measurements reveal gradual surface-state evolution at the GaN cap layer during the 30-day storage period. By incorporating an attention-enhanced ANN, the degradation in classification accuracy observed on 30-day datasets is alleviated, yielding improved prediction accuracy without recalibration.

Schlagwörter

Stichwörter der Autor:innen

AlGaN/GaN HEMT sensor arrayArtificial neural networkspH sensing

Autor:innen und Institutionen

  • School of Integrated Circuits, Dalian University of Technology, Dalian, 116620, PR China; Dalian Key Laboratory of Wide Bandgap Semiconductor Devices Integration and System, Dalian, 116024, PR China.
  • School of Integrated Circuits, Dalian University of Technology, Dalian, 116620, PR China; Dalian Key Laboratory of Wide Bandgap Semiconductor Devices Integration and System, Dalian, 116024, PR China. Electronic address:…
  • School of Integrated Circuits, Dalian University of Technology, Dalian, 116620, PR China.

Förderung

  • National Natural Science Foundation of China

Zitieren

APA

Zhu, J., Zhang, H., Guo, D., Ji, G., Wang, J., Sun, S., & Liang, H. (2026). Calibration-free and drift-robust AlGaN/GaN HEMT sensor arrays for intelligent pH detection. Analytica Chimica Acta, 1408, 345536. https://doi.org/10.1016/j.aca.2026.345536

Vancouver

Zhu J, Zhang H, Guo D, Ji G, Wang J, Sun S, et al. Calibration-free and drift-robust AlGaN/GaN HEMT sensor arrays for intelligent pH detection. Analytica Chimica Acta. 2026;1408:345536. doi:10.1016/j.aca.2026.345536

BibTeX
@article{Zhu2026,
  author = {Zhu, Jiang and Zhang, Heqiu and Guo, Dawei and Ji, Guanbing and Wang, Jialin and Sun, Shupeng and Liang, Hongwei},
  title = {Calibration-free and drift-robust AlGaN/GaN HEMT sensor arrays for intelligent pH detection},
  journal = {Analytica Chimica Acta},
  year = {2026},
  volume = {1408},
  pages = {345536},
  doi = {10.1016/j.aca.2026.345536}
}
Kostenlos starten →