EUVIMEDEUVIMED entdecken
Crossref

Harnessing vinylferrocene microgel-based photonic interferometers for clinical X-ray sensing

Jinnan Liu, Yajie Wang, Rui Hu, Liang Hu

Analytica Chimica Acta · 2026

DOI: 10.1016/j.aca.2026.345507 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytica Chimica Acta, 2026. Verfasst von Jinnan Liu und 3 weiteren Autor:innen. Beteiligte Einrichtung: State Key Laboratory of Radiation Medicine and Protection, School of Radiation Medicine and Protection, Collaborative Innovation Centre of Radiologic…. Hauptthemen laut MEDLINE: Ferrous Compounds, Interferometry, Microgels, Metallocenes. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 23 Literaturangaben. Gefördert von: Jiangsu Province Natural Science Foundation, Priority Academic Program Development of Jiangsu Higher Education Institutions, Soochow University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: X-rays play an essential role in modern tumor radiotherapy. Precise X-ray dosimetry is crucial for tumor radiotherapy to maximize therapeutic efficacy while minimizing normal tissue damage. Compared to traditional X-ray physical dosimeters, hydrogels are strong candidates for detecting X-ray doses due to their excellent tissue equivalence, ease of chemical modification, tunable sizes and low costs. However, most of these sensors are time-consuming, dose rate dependent, and irreversible. Additionally, they require sophisticated device readout systems and skilled personnel. To overcome these shortcomings, this study focuses on developing a hydrogel photonic material that is capable of sensing X-rays. RESULTS: In this paper, we first synthesize a series of poly(N-isopropylacrylamide-co-vinylferrocene) microgels (PVFc MG) with a variety of microstructures by modulating chemical compositions and synthetic strategies. Subsequently, Au-PVFc MG-Au photonic interferometers are generated by sandwiching PVFc MG monolayer between two Au metal layers. After depositing the X-ray-irradiated Fe2+ solution, the optical signals of the Au-PVFc MG-Au interferometer redshift due to the swelling of the microgel. As a result, the optimized interferometer exhibits a fast and linear response in the dose range of 0.1-20 Gy, a sensitivity of ∼4 nm/Gy, 95% of accuracy and dose-rate independence (0.847-5.085 Gy/min). It can return to its initial state in the presence of ascorbic acid, showing excellent reversible sensing capability. This work provides a promising platform for precise clinical X-ray dose monitoring in radiotherapy. SIGNIFICANCE AND NOVELTY: This study explores the use of vinylferrocene microgel-based photonic interferometers for detecting clinical X-ray doses. The sensing capabilities of the Au-PVFc MG-Au interferometers can be enhanced by modifying the chemical compositions and structures of the microgels. The underlying response mechanism is examined in detail. When applied in clinical X-ray dose sensing, these interferometers demonstrate exceptional performance, surpassing previous reports in terms of response time, linear responsivity, accuracy, reversibility, and dose rate independence.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Ferrous CompoundsInterferometryMicrogelsMetallocenesX-RaysPhotonsGoldAcrylic Resins

Stichwörter der Autor:innen

InterferometerMicrogelSensingVinylferroceneX-ray

Substanzen

Ferrous CompoundsMicrogelsMetallocenesGoldAcrylic Resinspoly-N-isopropylacrylamide

Autor:innen und Institutionen

  • State Key Laboratory of Radiation Medicine and Protection, School of Radiation Medicine and Protection, Collaborative Innovation Centre of Radiological Medicine of Jiangsu Higher Education Institutions, Soochow Universi…
  • Department of Radiation Oncology, The Affiliated Suzhou Hospital of Nanjing Medical University, Suzhou Municipal Hospital, Suzhou, 215002, China. Electronic address: hur.2011@tsinghua.org.cn.

Förderung

  • Jiangsu Province Natural Science Foundation
  • Priority Academic Program Development of Jiangsu Higher Education Institutions
  • Soochow University

Zitieren

APA

Liu, J., Wang, Y., Hu, R., & Hu, L. (2026). Harnessing vinylferrocene microgel-based photonic interferometers for clinical X-ray sensing. Analytica Chimica Acta. https://doi.org/10.1016/j.aca.2026.345507

Vancouver

Liu J, Wang Y, Hu R, Hu L. Harnessing vinylferrocene microgel-based photonic interferometers for clinical X-ray sensing. Analytica Chimica Acta. 2026. doi:10.1016/j.aca.2026.345507

BibTeX
@article{Liu2026,
  author = {Liu, Jinnan and Wang, Yajie and Hu, Rui and Hu, Liang},
  title = {Harnessing vinylferrocene microgel-based photonic interferometers for clinical X-ray sensing},
  journal = {Analytica Chimica Acta},
  year = {2026},
  doi = {10.1016/j.aca.2026.345507}
}