EUVIMEDEUVIMED entdecken
Crossref

Laser-induced graphene electrode modified by platinum nanoparticle/zein/gelatin/glucose oxidase for non-invasive glucose sensor in multiple biofluids

Kittiya Sakdaphetsiri, Thana Thaweeskulchai, Wiwittawin Sukmas, Joseph Wang, Albert Schulte, Nadnudda Rodthongkum

Analytica Chimica Acta · 2025

DOI: 10.1016/j.aca.2025.343974 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytica Chimica Acta, 2025. Verfasst von Kittiya Sakdaphetsiri und 5 weiteren Autor:innen. Beteiligte Einrichtung: Metallurgy and Materials Science Research Institute, Chulalongkorn University, Bangkok, Thailand. Hauptthemen laut MEDLINE: Glucose Oxidase, Graphite, Biosensing Techniques, Glucose. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 53 Literaturangaben. Gefördert von: Ministry of Higher Education, Science, Research and Innovation, Thailand, Chulalongkorn University, Thailand Science Research and Innovation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: A non-invasive glucose biosensor for multiple biofluids based on platinum nanoparticle (Pt-NP)-modified laser-induced graphene (LIG) electrodes coated with a zein/gelatin/glucose oxidase (GOx) for amperometric detection of glucose is created. The biosensor fabrication is cost-effective and scalable, as it combines simple LIG electrode fabrication with direct Pt-NP electrodeposition and a sequence of drop-and-dry steps for zein and gelatin layer then GOx enzyme. The Pt-NP modification on the LIG electrode functions as an electrocatalyst to enhance the anodic H2O2 signal, which is directly proportional to glucose concentration. The zein layer acts as a diffusion barrier to mitigate potential interferences, while the gelatin film provides amine groups for the glutaraldehyde-mediated immobilization of the GOx enzyme. RESULTS: The key parameters of LIG were optimized, such as power laser, number of Pt-NP cycles, and zein concentration. In addition, LIG was characterized by Raman spectroscopy, SEM, and cyclic voltammetry (CV) to ensure graphitization and electron transfer performance. The as prepared LIG/Pt-NP/Zein/Gel-GOx glucose biosensor was in anodic H2O2 detection mode and tested for glucose measurements in multiple biofluids including sweat, saliva, and urine. At H2O2 detection potential of +0.4 V, a linear detection range from 0 up to 2 mM glucose was obtained with a limit of detection (LOD) of 0.01 mM, making it feasible for glucose determination in various clinically relevant biofluids. By comparing with the commercial SPE, this LIG-based sensor offered much higher detection sensitivity towards both H2O2 and glucose, making it a superior choice for electrochemical analysis. SIGNIFICANCE: This LIG/Pt-NPs/Zein/Gel-GOx offers a practical and high sensitivity approach to glucose measurement with a wide linearity for multiple biofluids. Given the straightforward and easily scalable process, this high-performance, LIG-based glucose biosensor presents a compelling alternative over commercial screen-printed electrode. Highlighted the novelty using zein as a protective layer to reduce interferences. Owing to the simplicity of fabrication with high potential for up-scaling, this high analytical performance biosensor might be an alternative tool for multiplex glucose biosensors in point-of-care applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Glucose OxidaseGraphiteBiosensing TechniquesGlucoseMetal NanoparticlesZeinLasersPlatinumElectrodesGelatinHumansElectrochemical TechniquesSalivaLimit of DetectionEnzymes, ImmobilizedHydrogen PeroxideSweat

Stichwörter der Autor:innen

BiofluidsBiosensorsGlucoseLaser-induced graphene (LIG)Non-invasiveZein

Substanzen

Glucose OxidasePlatinumGraphiteGlucoseGelatinZeinEnzymes, ImmobilizedHydrogen Peroxide

Autor:innen und Institutionen

  • Metallurgy and Materials Science Research Institute, Chulalongkorn University, Bangkok, Thailand.
  • School of Biomolecular Science and Engineering (BSE), Vidyasirimedhi Institute of Science and Technology (VISTEC), Rayong, Thailand.
  • Extreme Conditions Physics Research Laboratory and Center of Excellence in Physics of Energy Materials (CE:PEM), Department of Physics, Faculty of Science, Chulalongkorn University, Bangkok, Thailand.
  • Department of Chemical and Nano Engineering, University of California San Diego, La Jolla, CA, USA.
  • School of Biomolecular Science and Engineering (BSE), Vidyasirimedhi Institute of Science and Technology (VISTEC), Rayong, Thailand. Electronic address: Albert.s@vistec.ac.th.
  • Department of Chemistry, Faculty of Science, Chulalongkorn University, Bangkok, Thailand; Center of Excellence in Responsive Wearable Materials, Metallurgy and Materials Science Research Institute, Chulalongkorn Univers…

Förderung

  • Ministry of Higher Education, Science, Research and Innovation, Thailand
  • Chulalongkorn University
  • Thailand Science Research and Innovation

Zitieren

APA

Sakdaphetsiri, K., Thaweeskulchai, T., Sukmas, W., Wang, J., Schulte, A., & Rodthongkum, N. (2025). Laser-induced graphene electrode modified by platinum nanoparticle/zein/gelatin/glucose oxidase for non-invasive glucose sensor in multiple biofluids. Analytica Chimica Acta. https://doi.org/10.1016/j.aca.2025.343974

Vancouver

Sakdaphetsiri K, Thaweeskulchai T, Sukmas W, Wang J, Schulte A, Rodthongkum N. Laser-induced graphene electrode modified by platinum nanoparticle/zein/gelatin/glucose oxidase for non-invasive glucose sensor in multiple biofluids. Analytica Chimica Acta. 2025. doi:10.1016/j.aca.2025.343974

BibTeX
@article{Sakdaphetsiri2025,
  author = {Sakdaphetsiri, Kittiya and Thaweeskulchai, Thana and Sukmas, Wiwittawin and Wang, Joseph and Schulte, Albert and Rodthongkum, Nadnudda},
  title = {Laser-induced graphene electrode modified by platinum nanoparticle/zein/gelatin/glucose oxidase for non-invasive glucose sensor in multiple biofluids},
  journal = {Analytica Chimica Acta},
  year = {2025},
  doi = {10.1016/j.aca.2025.343974}
}
Kostenlos starten →