High-resolution quadrupole improves spectral purity and reduces interference from non-target ions in isobaric multiplexed quantitative proteomics
Analytica Chimica Acta · 2024
DOI: 10.1016/j.aca.2024.343135 ↗Kurz gefasst
Fachartikel in der Zeitschrift Analytica Chimica Acta, 2024. Verfasst von Shen Zhang und 7 weiteren Autor:innen. Beteiligte Einrichtung: SCIEX, Vaughan, Ontario, L4K 4V8, Canada; Lunenfeld-Tanenbaum Research Institute, Mount Sinai Hospital, Sinai Health, Toronto, Ontario, M5G 1X5, Cana…. Hauptthemen laut MEDLINE: Proteomics. Die Arbeit nennt 20 Literaturangaben. Gefördert von: Canadian Institutes of Health Research, Mitacs, Canada Research Chairs Program.
Abstract (Originalsprache)
BACKGROUND: Mass spectrometry (MS)-based proteomics is a powerful tool for identifying and quantifying proteins. However, chimeric spectra caused by the fragmentation of multiple precursors within the same isolation window impair the accuracy of peptide identification and isobaric mass tag-based quantification. While there have been advances in computational deconvolution of chimeric spectra and methods to further separate the peptides by ion mobility or through MSn, the use of narrower isolation windows to decrease the fraction of chimeric species remains to be fully explored. RESULTS: We present results obtained on a SCIEX TripleTOF instrument where the quadrupole was optimized and tuned for precursor isolation at 0.1 Da (FWHH). Using a three-proteome model (trypsin digest of protein lysates from yeast, human and E. coli) and 8-plex iTRAQ labeling to document the interference effect, we investigated the impact of co-fragmentation on spectral purity, identification accuracy and quantification accuracy. The narrow quadrupole isolation window significantly improved the spectral purity and reduced the interference of non-target precursors on quantification accuracy. The high-resolution isolation strategy also reduced the number of false identifications caused by chimeric spectra. While these improvements came at the cost of sensitivity loss, combining high-resolution isolation with other advanced techniques, including ion mobility, may result in improved accuracy in identification and quantification. SIGNIFICANCE: Compared to standard-resolution quadrupole isolation (0.7 Da), high-resolution quadrupole isolation (0.1 Da) significantly improved the spectral purity and quantification accuracy while reducing the number of potential false identifications caused by chimeric spectra, thus showing excellent potential for further development to analyze clinical proteomics samples, for which high accuracy is essential.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- SCIEX, Vaughan, Ontario, L4K 4V8, Canada; Lunenfeld-Tanenbaum Research Institute, Mount Sinai Hospital, Sinai Health, Toronto, Ontario, M5G 1X5, Canada; NHC Key Laboratory of Human Stem Cell and Reproductive Engineering…
- SCIEX, Vaughan, Ontario, L4K 4V8, Canada.
- Lunenfeld-Tanenbaum Research Institute, Mount Sinai Hospital, Sinai Health, Toronto, Ontario, M5G 1X5, Canada.
- Lunenfeld-Tanenbaum Research Institute, Mount Sinai Hospital, Sinai Health, Toronto, Ontario, M5G 1X5, Canada; Department of Molecular Genetics, University of Toronto, Toronto, Ontario, M5G 1X8, Canada. Electronic addre…
- SCIEX, Vaughan, Ontario, L4K 4V8, Canada. Electronic address: Stephen.Tate@sciex.com.
Förderung
- Canadian Institutes of Health Research
- Mitacs
- Canada Research Chairs Program
- American Society for Mass Spectrometry
- Genome Canada
Zitieren
APA
Zhang, S., Le Blanc, J. C. Y., Larsen, B., Colwill, K., Burton, L., Guna, M., Gingras, A. C., & Tate, S. (2024). High-resolution quadrupole improves spectral purity and reduces interference from non-target ions in isobaric multiplexed quantitative proteomics. Analytica Chimica Acta. https://doi.org/10.1016/j.aca.2024.343135
Vancouver
Zhang S, Le Blanc JCY, Larsen B, Colwill K, Burton L, Guna M, et al. High-resolution quadrupole improves spectral purity and reduces interference from non-target ions in isobaric multiplexed quantitative proteomics. Analytica Chimica Acta. 2024. doi:10.1016/j.aca.2024.343135
BibTeX
@article{Zhang2024,
author = {Zhang, Shen and Le Blanc, J C Yves and Larsen, Brett and Colwill, Karen and Burton, Lyle and Guna, Mircea and Gingras, Anne-Claude and Tate, Stephen},
title = {High-resolution quadrupole improves spectral purity and reduces interference from non-target ions in isobaric multiplexed quantitative proteomics},
journal = {Analytica Chimica Acta},
year = {2024},
doi = {10.1016/j.aca.2024.343135}
}
EUVIMED