Fachartikel in der Zeitschrift Analytica Chimica Acta, 2024. Verfasst von Kasper Hobin und 4 weiteren Autor:innen. Beteiligte Einrichtung: Atomic & Mass Spectrometry - A&MS Research Unit, Department of Chemistry, Ghent University, 9000, Ghent, Belgium. Hauptthemen laut MEDLINE: Potassium, Mass Spectrometry. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 38 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Potassium isotopic analysis is increasingly performed in both geological and biological contexts as a result of the introduction of MC-ICP-MS instrumentation either equipped with a collision/reaction cell or having the capability of working at "extra-high" mass resolution in order to deal with spectral interference caused by argon hydride (ArH+) ions. Potassium plays an important role in the central nervous system, and its isotopic analysis could provide an enhanced insight into the corresponding processes, but K isotopic analysis of cerebrospinal fluid is challenging due to the small volume, a few microliter only, typically available. This work aimed at developing a method for determining the K isotopic signature of serum and cerebrospinal fluid at a final K concentration of 25 ng mL-1 using Faraday cup amplifiers equipped with a 1013 Ω resistor. RESULTS: Potassium isotope ratios obtained for reference materials measured at a final K concentration of 25 ng mL-1 were in excellent agreement with the corresponding reference values and the internal and external precision for the δ41K value was 0.11 ‰ (2SE, N = 50) and 0.10 ‰ (2SD, N = 6), respectively. The robustness against the presence of matrix elements and the concentration mismatch between sample and standard observed at higher K concentrations is preserved at low K concentration. Finally, K isotopic analysis of serum and cerebrospinal fluid (3-12 μL of sample) of healthy mice of both sexes was performed, revealing a trend towards an isotopically lighter signature for serum and cerebrospinal fluid from female individuals, however being significant for serum only. SIGNIFICANCE: This work provides a robust method for high-precision K isotopic analysis at a concentration of 25 ng mL-1. By monitoring both K isotopes, 39K and 41K, with Faraday cups connected to amplifiers with 1013 Ω resistors, accurate K isotope ratio results are obtained with a two-fold improvement in internal and external precision compared to those obtained with the set-up with traditional 1011 Ω resistors. The difference in the K isotope ratio in CSF and serum between the sexes, is possibly indicating an influence of the sex or hormones on the fractionation effects accompanying cellular uptake/release.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PotassiumMass SpectrometryAnimalsFemaleMaleMiceIsotopesHumans

Stichwörter der Autor:innen

10(13) Ω resistorsBlood serumCerebrospinal fluidIsotopic analysisMC-ICP-MSPotassium

Substanzen

PotassiumIsotopes

Autor:innen und Institutionen

  • Atomic & Mass Spectrometry - A&MS Research Unit, Department of Chemistry, Ghent University, 9000, Ghent, Belgium.
  • Atomic & Mass Spectrometry - A&MS Research Unit, Department of Chemistry, Ghent University, 9000, Ghent, Belgium; Centro de Investigación Mariña, Universidade de Vigo, Departamento de Química Analítica y Alimentaria, Gr…
  • Barriers in Inflammation Lab, VIB Center for Inflammation Research, 9052, Ghent, Belgium; Department of Biomedical Molecular Biology, Ghent University, 9052, Ghent, Belgium.
  • Atomic & Mass Spectrometry - A&MS Research Unit, Department of Chemistry, Ghent University, 9000, Ghent, Belgium. Electronic address: frank.vanhaecke@ugent.be.

Zitieren

APA

Hobin, K., Costas Rodríguez, M., Van Wonterghem, E., Vandenbroucke, R. E., & Vanhaecke, F. (2024). High-precision K isotopic analysis of cerebrospinal fluid and blood serum microsamples via multicollector inductively coupled plasma-mass spectrometry equipped with 1013 Ω faraday cup amplifier resistors. Analytica Chimica Acta. https://doi.org/10.1016/j.aca.2024.342812

Vancouver

Hobin K, Costas Rodríguez M, Van Wonterghem E, Vandenbroucke RE, Vanhaecke F. High-precision K isotopic analysis of cerebrospinal fluid and blood serum microsamples via multicollector inductively coupled plasma-mass spectrometry equipped with 1013 Ω faraday cup amplifier resistors. Analytica Chimica Acta. 2024. doi:10.1016/j.aca.2024.342812

BibTeX
@article{Hobin2024,
  author = {Hobin, Kasper and Costas Rodríguez, Marta and Van Wonterghem, Elien and Vandenbroucke, Roosmarijn E and Vanhaecke, Frank},
  title = {High-precision K isotopic analysis of cerebrospinal fluid and blood serum microsamples via multicollector inductively coupled plasma-mass spectrometry equipped with 1013 Ω faraday cup amplifier resistors},
  journal = {Analytica Chimica Acta},
  year = {2024},
  doi = {10.1016/j.aca.2024.342812}
}
Kostenlos starten →