Measurement of Macular Pigment in Singaporeans: Macular Metrics Densitometer Versus Macular Pigment Screener

Kumari Neelam, Joanna H Cher, Chalani U Ubeynarayana, Ivan Y-F Leung, Ian J Murray

Ophthalmic and Physiological Optics · 03.11.2026

DOI: 10.1007/s44402-026-00192-3 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Ophthalmic and Physiological Optics, online erschienen am 03.11.2026. Verfasst von Kumari Neelam und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology and Visual Sciences, Khoo Teck Puat Hospital, Singapore, Singapore. Kumari.neelam@nhghealth.com.sg. Die Arbeit nennt 48 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To compare macular pigment optical density (MPOD) obtained using a macular pigment screener (MPS II) and macular metrics densitometer (MMD).

Methods

Following details were collected in 65 healthy Singaporeans: sociodemographic, lifestyle information, height and weight, MPOD using MMD and MPS II including time taken to obtain measurements, and blood sample for estimation of serum lutein (L) and zeaxanthin (Z). MPOD agreement between MMD and MPS II was assessed using Bland–Altman plots and 95% limits of agreement. Intraclass correlation coefficient (ICC) was calculated to assess absolute concordance between the two instruments.

Results

Mean (±SD) age of the population was 27.92 years (±8.60) with slight preponderance of females. Mean MPOD measured with MPS II (0.48 ± 0.17) was significantly lower than MMD (0.57 ± 0.21). MPOD values recorded using MMD and MPS II were correlated (r = 0.72, p < 0.0001). Using Bland–Altman plots, upper and lower 95% limits of agreement were 0.370 and -0.194, respectively. Concordance of MPOD measured by MMD and MPS II using ICC was 0.64 (95% CI: 0.343–0.798). MPOD recording time was significantly longer using MPS II than MMD ( p < 0.05). Recording time increased significantly in subjects with higher MPOD using MPS II ( p = 0.004), whereas MPOD recording time did not differ with MMD ( p = 0.99). A weak correlation was observed between serum L and MPOD measured with MPS II (r = 0.31, p = 0.01) and MMD (r = 0.24, p = 0.06). Coefficients of repeatability for MMD and MPS II were 0.21 and 0.06, respectively.

Conclusions

MPS II measures MPOD in a reliable and reproducible manner when compared with MMD. However, MPS II consistently recorded lower values than MMD and has a longer recording time, particularly in those with higher MPOD. Moreover, MPS II failed to reach necessary level of agreement with MMD to allow interchangeability on an individual basis.

Schlagwörter

Stichwörter der Autor:innen

LuteinMacular metrics densitometerMacular pigmentMacular pigment screenerZeaxanthin

Autor:innen und Institutionen

  • Department of Ophthalmology and Visual Sciences, Khoo Teck Puat Hospital, Singapore, Singapore. Kumari.neelam@nhghealth.com.sg.
  • Singapore Eye Research Institute, Singapore National Eye Centre, Singapore, Singapore. Kumari.neelam@nhghealth.com.sg.
  • Clinical Research Unit, Khoo Teck Puat Hospital, Singapore, Singapore.
  • Singapore Polytechnic, School of Chemical and Life Science, Singapore, Singapore.
  • Faculty of Biology, Medicine and Health, University of Manchester, Manchester, UK.

Zitieren

APA

Neelam, K., Cher, J. H., Ubeynarayana, C. U., Leung, I. Y., & Murray, I. J. (2026). Measurement of Macular Pigment in Singaporeans: Macular Metrics Densitometer Versus Macular Pigment Screener. Ophthalmic and Physiological Optics. https://doi.org/10.1007/s44402-026-00192-3

Vancouver

Neelam K, Cher JH, Ubeynarayana CU, Leung IY, Murray IJ. Measurement of Macular Pigment in Singaporeans: Macular Metrics Densitometer Versus Macular Pigment Screener. Ophthalmic and Physiological Optics. 2026. doi:10.1007/s44402-026-00192-3

BibTeX
@article{Neelam2026,
  author = {Neelam, Kumari and Cher, Joanna H and Ubeynarayana, Chalani U and Leung, Ivan Y-F and Murray, Ian J},
  title = {Measurement of Macular Pigment in Singaporeans: Macular Metrics Densitometer Versus Macular Pigment Screener},
  journal = {Ophthalmic and Physiological Optics},
  year = {2026},
  doi = {10.1007/s44402-026-00192-3}
}
Kostenlos starten →