Analytical quality by design (AQbD) methodology for concurrent determination of perindopril erbumine and moxonidine hydrochloride using RP-HPLC with an eco-friendly evaluation
Discover Chemistry · 03.01.2025 · 2 (1)
DOI: 10.1007/s44371-024-00067-x ↗Kurz gefasst
Fachartikel in der Zeitschrift Discover Chemistry, online erschienen am 03.01.2025. Verfasst von Ashok H. Akabari und 8 weiteren Autor:innen. Die Arbeit wurde bisher 46-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
A sensitive, accurate, robust, and precise analytical quality by design (AQbD)-based RP-HPLC (reverse-phase high-performance liquid chromatography) method has been developed for the simultaneous estimation of perindopril erbumine and moxonidine hydrochloride, both used in hypertension treatment. Initially, a three-level factorial design screened various parameters impacting chromatographic responses, followed by Central Composite Design (CCD) for optimizing the critical parameters. This cost-effective isocratic method employed a Shimpack ODS C-18 column (250 mm × 4.6 mm, 5 μm) as the stationary phase. The method, developed and validated in line with ICH Q2 (R1) guidelines, utilized a mobile phase comprising methanol: acetonitrile: phosphate buffer (34:30:36 v/v/v) adjusted to pH 3.5 with 1% ortho phosphoric acid, at a flow rate of 1.0 mL/min, a column temperature of 40 ºC, and UV detection at 214 nm, achieving efficient separation of both drugs. Linearity ranges were 25–125 μg/mL for perindopril erbumine (r2 = 0.9996) and 1–5 μg/mL for moxonidine hydrochloride (r2 = 0.9993). This newly developed RP-HPLC method effectively enables routine quantitative and simultaneous qualitative analysis of these compounds in bulk and synthetic mixtures. Additionally, the method’s greenness and environmental safety were evaluated using eco-analytical metrics (eco analytical scale, NEMI, GAPI, and AGREE), confirming its ecological capability.
Zitieren
APA
Akabari, A. H., Patel, S. K., Shah, K. V., Patel, A., Modi, D., Sen, A. K., Ramani, V. D., Shah, D. P., & Patel, S. P. (2025). Analytical quality by design (AQbD) methodology for concurrent determination of perindopril erbumine and moxonidine hydrochloride using RP-HPLC with an eco-friendly evaluation. Discover Chemistry, 2(1). https://doi.org/10.1007/s44371-024-00067-x
Vancouver
Akabari AH, Patel SK, Shah KV, Patel A, Modi D, Sen AK, et al. Analytical quality by design (AQbD) methodology for concurrent determination of perindopril erbumine and moxonidine hydrochloride using RP-HPLC with an eco-friendly evaluation. Discover Chemistry. 2025;2(1). doi:10.1007/s44371-024-00067-x
BibTeX
@article{Akabari2025,
author = {Akabari, Ashok H. and Patel, Sagar K. and Shah, Ketan V. and Patel, Arun and Modi, Dixitkumar and Sen, Ashim Kumar and Ramani, Vinodkumar D. and Shah, Dhiren P. and Patel, Sagarkumar P.},
title = {Analytical quality by design (AQbD) methodology for concurrent determination of perindopril erbumine and moxonidine hydrochloride using RP-HPLC with an eco-friendly evaluation},
journal = {Discover Chemistry},
year = {2025},
volume = {2},
number = {1},
doi = {10.1007/s44371-024-00067-x}
}