Dynamic extraction of residual solvents from pharmaceutical formulations using a needle-type extraction device

Ikuo Ueta, Daisuke Kondo, Suguru Masuda, Yuki Ariizumi, Ryosuke Kikuchi, Koki Nakagami, Yoshihiro Saito

Analytical Sciences · 2025

DOI: 10.1007/s44211-024-00674-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical Sciences, 2025. Verfasst von Ikuo Ueta und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Applied Chemistry, University of Yamanashi, 4-3-11 Takeda, Kofu, 400-8511, Japan. iueta@yamanashi.ac.jp. Hauptthemen laut MEDLINE: Solvents, Needles. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 16 Literaturangaben. Gefördert von: Japan Society for the Promotion of Science.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study developed a rapid screening method to determine residual solvents contained in a tablet formulation using a needle-type extraction device. For this, the tablet formulation was pulverized and the powdered sample was rapidly inserted into a pipette tip. After fixing the sample with quartz wool, the pipette tip was capped with a silicon septum. A needle-type extraction device packed with Carbopack X and a carbon molecular sieve was inserted into the pipette tip through the septum, and the gas sample was collected. During this gas sampling, pure N2 gas was introduced into the pipette tip. The extraction time for collecting 100 mL of the sample was approximately 10 min. After gas sampling, the extraction needle was connected to a gas-tight syringe and 0.5 mL of pure N2 gas (desorption gas) was collected. The extraction needle was then inserted into a heated gas chromatographic injection port. The extracted residual solvents were then thermally desorbed and determined using a gas chromatograph-flame ionization detector. The detector response is very low for carbon tetrachloride, although the proposed method showed sufficient sensitivity for five Class 1 compounds. Additionally, this study clearly indicated that the purge efficiency of residual solves with dynamic extraction is different between powdered pharmaceutical formulations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SolventsNeedlesChromatography, GasPharmaceutical PreparationsTablets

Stichwörter der Autor:innen

Gas chromatographyNeedle-type extraction deviceResidual solventsVOCs

Substanzen

SolventsPharmaceutical PreparationsTablets

Autor:innen und Institutionen

  • Department of Applied Chemistry, University of Yamanashi, 4-3-11 Takeda, Kofu, 400-8511, Japan. iueta@yamanashi.ac.jp.
  • Department of Applied Chemistry, University of Yamanashi, 4-3-11 Takeda, Kofu, 400-8511, Japan.
  • Department of Pharmacy, University of Yamanashi Hospital, 1110 Shimokato, Chuo, 409-3898, Japan.
  • Department of Applied Chemistry and Life Science, Toyohashi University of Technology, 1-1 Hibarigaoka, Tempaku, Toyohashi, 441-8580, Japan.

Förderung

  • Japan Society for the Promotion of Science

Zitieren

APA

Ueta, I., Kondo, D., Masuda, S., Ariizumi, Y., Kikuchi, R., Nakagami, K., & Saito, Y. (2025). Dynamic extraction of residual solvents from pharmaceutical formulations using a needle-type extraction device. Analytical Sciences. https://doi.org/10.1007/s44211-024-00674-9

Vancouver

Ueta I, Kondo D, Masuda S, Ariizumi Y, Kikuchi R, Nakagami K, et al. Dynamic extraction of residual solvents from pharmaceutical formulations using a needle-type extraction device. Analytical Sciences. 2025. doi:10.1007/s44211-024-00674-9

BibTeX
@article{Ueta2025,
  author = {Ueta, Ikuo and Kondo, Daisuke and Masuda, Suguru and Ariizumi, Yuki and Kikuchi, Ryosuke and Nakagami, Koki and Saito, Yoshihiro},
  title = {Dynamic extraction of residual solvents from pharmaceutical formulations using a needle-type extraction device},
  journal = {Analytical Sciences},
  year = {2025},
  doi = {10.1007/s44211-024-00674-9}
}
Kostenlos starten →