Pulse Wave Velocity in Young South African Adults: Comparison Between the SphygmoCor CVMS and SphygmoCor XCEL Devices

Wessel L. du Toit, Thembelihle Myeza, Vernice R. Peterson, Kamohelo E. Sithole, Angela J. Woodiwiss, Andrea Kolkenbeck-Ruh

Artery Research · 08.05.2026 · 32 (1)

DOI: 10.1007/s44200-026-00115-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Artery Research, online erschienen am 08.05.2026. Verfasst von Wessel L. du Toit und 5 weiteren Autor:innen. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of the Witwatersrand, Johannesburg, National Research Foundation, South Africa.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Carotid-femoral pulse wave velocity (PWV) is the gold-standard non-invasive measure of arterial stiffness. Technological advances have led to the development of newer devices for measuring PWV, resulting in the discontinuation of earlier models. However, it is unclear whether PWV measurements from different devices can be combined within the same population. We therefore compared PWV obtained using the SphygmoCor CVMS (tonometer-ECG-gated) and XCEL (tonometer-thigh cuff) devices in a sample of young adults. South African adults ( n = 115, mean age: 35 ± 12 years, 50.4% male, 63.5% Black African) attending the ongoing Multi-Ethnic Early Vascular Ageing Study (MEEVA) had PWV measurements using SphygmoCor CVMS and XCEL devices. Wilcoxon signed-rank test, Spearman’s rank correlation, Bland-Altman plots, and regression analysis were employed to compare PWV across devices.

Results

No difference ( p = 0.07) in the PWV measurements between the CVMS (6.3 m/s, IQR: 5.5–7.4) and XCEL devices (6.2 m/s, IQR: 5.5–7.2), and a moderate correlation was observed (r s = 0.69, p < 0.001). There was an acceptable level of agreement (mean difference in PWV: 0.32 ± 1.67 m/s; limits of agreement: -2.95 to 3.59), albeit with a proportional bias, as CVMS recorded higher PWV than XCEL, with mean PWV > 8.5 m/s. Age ( p = 0.058) was the primary determinant of the difference in PWV between the devices.

Conclusion

There was no difference in PWV between CVMS and XCEL devices, and a good limit of agreement; a proportional bias was observed at higher PWV values. Hence, there is potential to pool data collected from either device, which would aid in creating larger South African-specific PWV databases.

Förderung

  • University of the Witwatersrand, Johannesburg
  • National Research Foundation, South Africa

Zitieren

APA

Toit, W. L. D., Myeza, T., Peterson, V. R., Sithole, K. E., Woodiwiss, A. J., & Kolkenbeck-Ruh, A. (2026). Pulse Wave Velocity in Young South African Adults: Comparison Between the SphygmoCor CVMS and SphygmoCor XCEL Devices. Artery Research, 32(1). https://doi.org/10.1007/s44200-026-00115-0

Vancouver

Toit WLD, Myeza T, Peterson VR, Sithole KE, Woodiwiss AJ, Kolkenbeck-Ruh A. Pulse Wave Velocity in Young South African Adults: Comparison Between the SphygmoCor CVMS and SphygmoCor XCEL Devices. Artery Research. 2026;32(1). doi:10.1007/s44200-026-00115-0

BibTeX
@article{Toit2026,
  author = {Toit, Wessel L. du and Myeza, Thembelihle and Peterson, Vernice R. and Sithole, Kamohelo E. and Woodiwiss, Angela J. and Kolkenbeck-Ruh, Andrea},
  title = {Pulse Wave Velocity in Young South African Adults: Comparison Between the SphygmoCor CVMS and SphygmoCor XCEL Devices},
  journal = {Artery Research},
  year = {2026},
  volume = {32},
  number = {1},
  doi = {10.1007/s44200-026-00115-0}
}
Kostenlos starten →