Analysis of tether breakages during revision after failed thoracolumbar double-row vertebral body tethering
Spine Deformity · 2026-09 · 14 (5) · 1807-1815
DOI: 10.1007/s43390-026-01345-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift Spine Deformity, online erschienen am 03.04.2026. Verfasst von Jil Frank und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics, Trauma and Reconstructive Surgery, Uniklinik RWTH Aachen, Pauwelsstraße 30, 52074, Aachen, Germany. jil.frank@rwth-aachen…. Hauptthemen laut MEDLINE: Reoperation, Lumbar Vertebrae, Thoracic Vertebrae, Spinal Fusion. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: RWTH Aachen University.
Abstract (Originalsprache)
Purpose
Vertebral Body Tethering (VBT) yields improving postoperative results. However, tether breakage remains a common complication, especially in thoracolumbar (TL) curves, and can negatively impact the outcome. The pathomechanism behind TL tether breakages is not well understood, yet this knowledge is essential for improving surgical planning. Therefore, the study aimed to investigate the timing and location of tether breakage in double-row (TL) revision cases. Material This study performed a descriptive analysis of consecutively explanted double-row implants in TL VBT patients who underwent revision due to tether breakage and for whom photographic documentation was available. A > 5° increase in the interscrew angle on consecutive radiographs suggested tether breakage.
Results
Data from 50 instrumented levels in ten patients were available. The UIV ranged from T9 to T11 and the LIV from L3 to L4. Revision surgery was performed, on average, 35 months after the index surgery. The study showed that the > 5° rule identified a higher percentage of segments with both tethers broken (68%) than segments with only one broken tether (54%). There was no difference in the occurrence of anterior (61%) and posterior (62%) tether breakage. The anterior tether broke in the distal segment in all ten cases, while the posterior tether broke in the distal segment in nine out of ten cases.
Conclusion
This study showed that the distal segment of the lumbar spine is prone for tether breakage. Fewer breakages were observed in the thoracic segments. Additionally, no clear breakage pattern in relation to implant standing time was observed.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedics, Trauma and Reconstructive Surgery, Uniklinik RWTH Aachen, Pauwelsstraße 30, 52074, Aachen, Germany. jil.frank@rwth-aachen.de.
- Casa Di Cura Humanitas San Pio X, Milan, Italy.
- Department of Spine Surgery, Eifelklinik St Brigida, Simmerath, Germany.
Förderung
- RWTH Aachen University
Zitieren
APA
Frank, J., Baroncini, A., Alkharsawi, M., & Trobisch, P. (2026). Analysis of tether breakages during revision after failed thoracolumbar double-row vertebral body tethering. Spine Deformity, 14(5), 1807-1815. https://doi.org/10.1007/s43390-026-01345-7
Vancouver
Frank J, Baroncini A, Alkharsawi M, Trobisch P. Analysis of tether breakages during revision after failed thoracolumbar double-row vertebral body tethering. Spine Deformity. 2026;14(5):1807-1815. doi:10.1007/s43390-026-01345-7
BibTeX
@article{Frank2026,
author = {Frank, Jil and Baroncini, Alice and Alkharsawi, Mahmoud and Trobisch, Per},
title = {Analysis of tether breakages during revision after failed thoracolumbar double-row vertebral body tethering},
journal = {Spine Deformity},
year = {2026},
volume = {14},
number = {5},
pages = {1807--1815},
doi = {10.1007/s43390-026-01345-7}
}