Fachartikel in der Zeitschrift Spine Deformity, 2024. Verfasst von Niloufar Shekouhi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Engineering Center for Orthopedic Research Excellence (E-CORE), Departments of Bioengineering and Orthopaedic Surgery, University of Toledo, Toledo, …. Hauptthemen laut MEDLINE: Osteotomy, Lumbar Vertebrae. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NSF Industry/University Cooperative Research Center.
Abstract (Originalsprache)
Purpose
To develop and validate a finite element (FE) model of a sacral pedicle subtraction osteotomy (S1-PSO) and to compare biomechanical properties of various multi-rod configurations to stabilize S1-PSOs.
Methods
A previously validated FE spinopelvic model was used to develop a 30° PSO at the sacrum. Five multi-rod techniques spanning the S1-PSO were made using 4 iliac screws and a variety of primary rods (PR) and accessory rods (AR; lateral: Lat-AR or medial: Med-AR). All constructs, except one, utilized a horizontal rod (HR) connecting the iliac bolts to which PRs and Med-ARs were connected. Lat-ARs were connected to proximal iliac bolts. The simulation was performed in two steps with the acetabula fixed. For each model, PSO ROM and maximum stress on the PRs, ARs, and HRs were recorded and compared. The maximum stress on the L5–S1 disc and the PSO forces were captured and compared.
Results
Highest PSO ROMs were observed for 4-Rods (HR + 2 Med-AR). Constructs consisting of 5-Rods (HR + 2 Lat-ARs + 1 Med-AR) and 6-Rods (HR + 2 Lat-AR + 2 Med-AR) had the lowest PSO ROM. The least stress on the primary rods was seen with 6-Rods, followed by 5-Rods and 4-Rods (HR + 2 Lat-ARs). Lowest PSO forces and lowest L5–S1 disc stresses were observed for 4-Rod (Lat-AR), 5-Rod, and 6-Rod constructs, while 4-Rods (HR + Med-AR) had the highest.
Conclusion
In this first FE analysis of an S1-PSO, the 4-Rod construct (HR + Med-AR) created the least rigid environment and highest PSO forces anteriorly. While 5- and 6-Rods created the stiffest constructs and lowest stresses on the primary rods, it also jeopardized load transfer to the anterior column, which may not be favorable for healing anteriorly. A balance between the construct’s rigidity and anterior load sharing is essential.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Engineering Center for Orthopedic Research Excellence (E-CORE), Departments of Bioengineering and Orthopaedic Surgery, University of Toledo, Toledo, OH, USA.
- Department of Orthopaedic Surgery, University of California-San Francisco (UCSF), 500 Parnassus Ave, MUW 3rd Floor, San Francisco, CA, 94143, USA. alekos.theologis@ucsf.edu.
ORCID-Profile: Alekos A Theologis
Förderung
- NSF Industry/University Cooperative Research Center
Zitieren
APA
Shekouhi, N., Tripathi, S., Goel, V. K., & Theologis, A. A. (2024). Biomechanical evaluation of multi-rod constructs to stabilize an S1 pedicle subtraction osteotomy (PSO): a finite element analysis. Spine Deformity. https://doi.org/10.1007/s43390-023-00784-w
Vancouver
Shekouhi N, Tripathi S, Goel VK, Theologis AA. Biomechanical evaluation of multi-rod constructs to stabilize an S1 pedicle subtraction osteotomy (PSO): a finite element analysis. Spine Deformity. 2024. doi:10.1007/s43390-023-00784-w
BibTeX
@article{Shekouhi2024,
author = {Shekouhi, Niloufar and Tripathi, Sudharshan and Goel, Vijay K and Theologis, Alekos A},
title = {Biomechanical evaluation of multi-rod constructs to stabilize an S1 pedicle subtraction osteotomy (PSO): a finite element analysis},
journal = {Spine Deformity},
year = {2024},
doi = {10.1007/s43390-023-00784-w}
}
EUVIMED