Pulse wave velocity measurement assessing the superficial carotid–femoral distance via smartphone-based light detection and ranging

Igor N Posokhov

Journal of Ultrasound · 29.10.2026

DOI: 10.1007/s40477-026-01252-7 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Ultrasound, online erschienen am 29.10.2026. Verfasst von Igor N Posokhov. Beteiligte Einrichtung: West Kazakhstan Marat Ospanov Medical University, Maresyev Str., 68, 030000, Aktobe, Kazakhstan. igor@posohov.ru. Die Arbeit nennt 11 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Carotid–femoral pulse wave velocity (cfPWV) is the gold standard for non-invasive assessment of aortic stiffness. Tape measurement of the carotid–femoral distance typically requires an assistant to hold the tape while the operator maintains the Doller transducer position. This study assessed the concordance and intra-observer reliability of distance measurements obtained via smartphone-based light detection and ranging (LiDAR) technology compared with traditional tape measurement.

Methods

A cohort of 46 adult participants (mean age, 50.0±12.7 years) underwent Doppler ultrasound evaluation to determine carotid and femoral pulse transit times. The operator measured the superficial carotid–femoral distance using tape and smartphone LiDAR in alternating sequences. Duplicate measurements were taken following a washout period to evaluate intra-observer reliability.

Results

LiDAR distance measurements strongly agreed with tape measurements (ICC (2,1) = 0.92, 95% CI 0.81–0.97; mean bias 10 m/s) showed near-perfect concordance between the methods (kappa > 0.98).

Conclusions

Smartphone-based LiDAR telemetry provides distance measurements and cfPWV estimates that are highly concordant with traditional tape measurements, offering a reliable single-operator alternative for superficial distance acquisition. While these findings confirm technical equivalence and intra-observer reliability, further studies incorporating reference standards and quantitative time-motion analyses are required to confirm the clinical workflow advantages of this method. Graphical abstract

Schlagwörter

Stichwörter der Autor:innen

Aortic stiffnessCarotid–femoral pulse wave velocityLiDARReliabilitySmartphoneVascular ultrasound

Autor:innen und Institutionen

  • West Kazakhstan Marat Ospanov Medical University, Maresyev Str., 68, 030000, Aktobe, Kazakhstan. igor@posohov.ru.

Zitieren

APA

Posokhov, I. N. (2026). Pulse wave velocity measurement assessing the superficial carotid–femoral distance via smartphone-based light detection and ranging. Journal of Ultrasound. https://doi.org/10.1007/s40477-026-01252-7

Vancouver

Posokhov IN. Pulse wave velocity measurement assessing the superficial carotid–femoral distance via smartphone-based light detection and ranging. Journal of Ultrasound. 2026. doi:10.1007/s40477-026-01252-7

BibTeX
@article{Posokhov2026,
  author = {Posokhov, Igor N},
  title = {Pulse wave velocity measurement assessing the superficial carotid–femoral distance via smartphone-based light detection and ranging},
  journal = {Journal of Ultrasound},
  year = {2026},
  doi = {10.1007/s40477-026-01252-7}
}
Kostenlos starten →