Artifact-resilient design and early reading extraction in a wearable microcavity NDIR transcutaneous CO2 prototype: validation of a PDMS interface layer and the R2-max rate-window method

Xin Xiong, Hanzheng Xu iD, Yunying Cai, Xiao Mo, Heng Su, JianFeng He

Biomedical Engineering Letters · 2026-11 · 16 (6) · 1489-1502

DOI: 10.1007/s13534-026-00611-x ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Engineering Letters, online erschienen am 24.09.2026. Verfasst von Xin Xiong und 5 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Information Engineering and Automation, Kunming University of Science and Technology, Kunming, 650500 China. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitieren

Abstract (Originalsprache)

Wearable skin-attached NDIR CO2 sensing is vulnerable to humidity, condensation, and contact perturbations, which can cause unstable signals and delayed valid readings. This study developed a wearable microcavity NDIR tcCO2 prototype and evaluated whether a PDMS interface layer combined with an R2-max rate-window method could improve signal stability and early reading extraction. The prototype integrated a miniature NDIR CO2 sensor, a 3D-printed wrist microcavity, and a wristband fixation structure. Human wrist recordings were obtained under direct detection without a membrane (NoMem) and with a PDMS membrane. Signals were preprocessed using median filtering and local polynomial smoothing. Sliding-window linear fitting was then applied, and valid rate windows were selected using R2, drop ratio, and slope constraints. Supplementary gas-gap, PDMS-thickness, repeatability, and post-exercise experiments were also performed. Valid rate windows were identified in 4/4 PDMS records and 4/8 NoMem records. The PDMS condition showed a median rise rate of 0.954 raw/s (IQR: 0.105 raw/s) and a median window center time of 280.5 s, whereas NoMem showed 1.159 raw/s (IQR: 0.581 raw/s) and 360.0 s. PDMS yielded more concentrated rate distributions and fewer unstable or invalid outcomes. Supplementary analyses supported the 0.3 cm gas gap, 100 μm PDMS film, repeated valid-window extraction, and post-exercise early-window formation. The PDMS interface and R2-max method improved prototype-level raw-signal stability and interpretable early-window extraction. These findings support engineering feasibility for wearable microcavity NDIR CO2 monitoring, but do not yet establish clinical quantitative tcCO2 accuracy.

Schlagwörter

Stichwörter der Autor:innen

Artifact suppressionNon-dispersive infrared (NDIR)Polydimethylsiloxane (PDMS)Rate windowTranscutaneous carbon dioxideWearable sensing

Autor:innen und Institutionen

  • Faculty of Information Engineering and Automation, Kunming University of Science and Technology, Kunming, 650500 China.
  • Endocrinology Department, First People's Hospital of Yunnan Province, The Affiliated Hospital of Kunming University of Science and Technology, Type 1 Diabetes Alliance of Yunnan Province, Kunming, 650032 China.
  • Department of Internal Medicine, Fuwai Yunnan Hospital, Chinese Academy of Medical Sciences, Affiliated Cardiovascular Hospital of Kunming Medical University, Kunming, 650102 China.

ORCID-Profile: Hanzheng Xu

Zitieren

APA

Xiong, X., Xu, H., Cai, Y., Mo, X., Su, H., & He, J. (2026). Artifact-resilient design and early reading extraction in a wearable microcavity NDIR transcutaneous CO2 prototype: validation of a PDMS interface layer and the R2-max rate-window method. Biomedical Engineering Letters, 16(6), 1489-1502. https://doi.org/10.1007/s13534-026-00611-x

Vancouver

Xiong X, Xu H, Cai Y, Mo X, Su H, He J. Artifact-resilient design and early reading extraction in a wearable microcavity NDIR transcutaneous CO2 prototype: validation of a PDMS interface layer and the R2-max rate-window method. Biomedical Engineering Letters. 2026;16(6):1489-1502. doi:10.1007/s13534-026-00611-x

BibTeX
@article{Xiong2026,
  author = {Xiong, Xin and Xu, Hanzheng and Cai, Yunying and Mo, Xiao and Su, Heng and He, JianFeng},
  title = {Artifact-resilient design and early reading extraction in a wearable microcavity NDIR transcutaneous CO2 prototype: validation of a PDMS interface layer and the R2-max rate-window method},
  journal = {Biomedical Engineering Letters},
  year = {2026},
  volume = {16},
  number = {6},
  pages = {1489--1502},
  doi = {10.1007/s13534-026-00611-x}
}
Kostenlos starten →