EUVIMEDEUVIMED entdecken
Randomisierte kontrollierte StudieOpen Access · CC BY-NC 4.0

Impact of Ascorbic Acid and Exercise on Accuracy of a Glucose Monitoring System with GDH-FAD Technology in Adult Patients with Type 1 Diabetes Mellitus: A Double-Blinded, Randomized Trial

Ronald Brazg, Tina Mitchell, Jiangfeng Fei, Jiyun Zheng, Fei Gao, Ao Gao, Siting Zhu, Frank Flacke, Leon Shi, Poul Strange

Diabetes Therapy · 2026-11 · 17 (11) · 1745-1755

DOI: 10.1007/s13300-026-01918-9 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift Diabetes Therapy, online erschienen am 09.10.2026. Verfasst von Ronald Brazg und 9 weiteren Autor:innen. Beteiligte Einrichtung: Rainier Clinical Research Center, Renton, WA, USA. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Sinocare Inc.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

The performance of a new continuous glucose monitoring system (CGM) with flavin adenine dinucleotide-dependent glucose dehydrogenase (GDH-FAD) technology in adults with type 1 diabetes mellitus (T1DM) was assessed, with a focus on the impact of ascorbic acid and exercise on the accuracy of glucose measurements relative to Yellow Springs Instrument (YSI) 2300 control.

Methods

Adults with T1DM aged 18–65 years using a stable glucose management treatment regimen for at least 3 months with multiple daily injections or continuous subcutaneous insulin infusion participated in this study at a single center. Participants attended a screening visit and three in-clinic visits for frequent blood sampling with blood glucose measurements using YSI measurements as the reference measurement as endorsed by the Food and Drug Administration (FDA). Participants had two meals at each clinic visit leading to pronounced glucose excursions. On the second and third in-clinic visits, participants performed moderate exercise in between the two meals and were randomized to receive 1000 mg of ascorbic acid orally at either of those visits. Accuracy evaluation included the proportion of CGM values within 15% of YSI glucose values > 100 mg/dL or within 15 mg/dL of YSI values ≤ 100 mg/dL (%15/15), along with %20/20, %30/30, and %40/40 agreement rates (AR) across CGM ranges. The mean absolute difference (MAD) and mean absolute relative difference (MARD) with and without ascorbic acid and exercise, as well as their differences, are presented across glucose ranges.

Results

Data from 16 participants were analyzed. In the overall CGM range (35–450 mg/dL), the %20/20 AR was 92.0%. The overall changes in MAD between the exercise/nonexercise and the ascorbic acid/nonascorbic acid conditions were −1.3 mg/dL and −1.1 mg/dL. The overall changes in MARD between those two sets of conditions were −1.5% and −1.6%, respectively, none of which were considered to be clinically relevant. There were no serious adverse events.

Conclusions

This study provides evidence that neither exercise nor ascorbic acid results in any clinically significant interference on glucose measurements using the iCan CGM system. Trial Registration ClinicalTrials.gov identifier, NCT05348928.

Schlagwörter

Stichwörter der Autor:innen

AccuracyAscorbic acidContinuous glucose monitoringExerciseInterferenceType 1 diabetes mellitusiCan i3 CGM

Autor:innen und Institutionen

  • Rainier Clinical Research Center, Renton, WA, USA.
  • Sinocare Meditech, Inc., Lauderdale, FL, USA.
  • Sinocare, 265 GuYuan Road, Hi-Tech Zone, Changsha, Hunan, 410205, People's Republic of China.
  • Flacke Consulting GmbH, Haydnstr. 12, 55130, Mainz, Germany.
  • Integrated Medical Development, 186 Princeton Hightstown Rd, Princeton Junction, NJ, 08550, USA.
  • Integrated Medical Development, 186 Princeton Hightstown Rd, Princeton Junction, NJ, 08550, USA. pstrange@imdcro.com.

Förderung

  • Sinocare Inc

Zitieren

APA

Brazg, R., Mitchell, T., Fei, J., Zheng, J., Gao, F., Gao, A., Zhu, S., Flacke, F., Shi, L., & Strange, P. (2026). Impact of Ascorbic Acid and Exercise on Accuracy of a Glucose Monitoring System with GDH-FAD Technology in Adult Patients with Type 1 Diabetes Mellitus: A Double-Blinded, Randomized Trial. Diabetes Therapy, 17(11), 1745-1755. https://doi.org/10.1007/s13300-026-01918-9

Vancouver

Brazg R, Mitchell T, Fei J, Zheng J, Gao F, Gao A, et al. Impact of Ascorbic Acid and Exercise on Accuracy of a Glucose Monitoring System with GDH-FAD Technology in Adult Patients with Type 1 Diabetes Mellitus: A Double-Blinded, Randomized Trial. Diabetes Therapy. 2026;17(11):1745-1755. doi:10.1007/s13300-026-01918-9

BibTeX
@article{Brazg2026,
  author = {Brazg, Ronald and Mitchell, Tina and Fei, Jiangfeng and Zheng, Jiyun and Gao, Fei and Gao, Ao and Zhu, Siting and Flacke, Frank and Shi, Leon and Strange, Poul},
  title = {Impact of Ascorbic Acid and Exercise on Accuracy of a Glucose Monitoring System with GDH-FAD Technology in Adult Patients with Type 1 Diabetes Mellitus: A Double-Blinded, Randomized Trial},
  journal = {Diabetes Therapy},
  year = {2026},
  volume = {17},
  number = {11},
  pages = {1745--1755},
  doi = {10.1007/s13300-026-01918-9}
}
Kostenlos starten →