Y-90 PET/MR imaging optimization with a Bayesian penalized likelihood reconstruction algorithm
Physical and Engineering Sciences in Medicine · 2024
DOI: 10.1007/s13246-024-01452-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physical and Engineering Sciences in Medicine, 2024. Verfasst von José Calatayud-Jordán und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nuclear Medicine, La Fe University and Polytechnical Hospital, Av. Fernando Abril Martorell 106, 46026, Valencia, Spain. calatayud_josj…. Hauptthemen laut MEDLINE: Bayes Theorem, Algorithms, Positron-Emission Tomography, Phantoms, Imaging. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 61 Literaturangaben.
Abstract (Originalsprache)
Positron Emission Tomography (PET) imaging after 90 Y liver radioembolization is used for both lesion identification and dosimetry. Bayesian penalized likelihood (BPL) reconstruction algorithms are an alternative to ordered subset expectation maximization (OSEM) with improved image quality and lesion detectability. The investigation of optimal parameters for 90 Y image reconstruction of Q.Clear, a commercial BPL algorithm developed by General Electric (GE), in PET/MR is a field of interest and the subject of this study. The NEMA phantom was filled at an 8:1 sphere-to-background ratio. Acquisitions were performed on a PET/MR scanner for clinically relevant activities between 0.7 and 3.3 MBq/ml. Reconstructions with Q.Clear were performed varying the β penalty parameter between 20 and 6000, the acquisition time between 5 and 20 min and pixel size between 1.56 and 4.69 mm. OSEM reconstructions of 28 subsets with 2 and 4 iterations with and without Time-of-Flight (TOF) were compared to Q.Clear with β = 4000. Recovery coefficients (RC), their coefficient of variation (COV), background variability (BV), contrast-to-noise ratio (CNR) and residual activity in the cold insert were evaluated. Increasing β parameter lowered RC, COV and BV, while CNR was maximized at β = 4000; further increase resulted in oversmoothing. For quantification purposes, β = 1000-2000 could be more appropriate. Longer acquisition times resulted in larger CNR due to reduced image noise. Q.Clear reconstructions led to higher CNR than OSEM. A β of 4000 was obtained for optimal image quality, although lower values could be considered for quantification purposes. An optimal acquisition time of 15 min was proposed considering its clinical use.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Nuclear Medicine, La Fe University and Polytechnical Hospital, Av. Fernando Abril Martorell 106, 46026, Valencia, Spain. calatayud_josjor@gva.es.
- Radiophysics and Radiological Protection Service, Clinical University Hospital of Valencia, Av. Blasco Ibáñez 17, 46010, Valencia, Spain.
- Department of Nuclear Medicine, La Fe University and Polytechnical Hospital, Av. Fernando Abril Martorell 106, 46026, Valencia, Spain.
- Biomedical Imaging Research Group (GIBI230) at Health Research Institute Hospital La Fe (IIS La Fe), La Fe University and Polytechnical Hospital, Av. Fernando Abril Martorell 106, 46026, Valencia, Spain.
- Department of Radiology, La Fe University and Polytechnical Hospital, Av. Fernando Abril Martorell 106, 46026, Valencia, Spain.
Zitieren
APA
Calatayud-Jordán, J., Carrasco-Vela, N., Chimeno-Hernández, J., Carles-Fariña, M., Olivas-Arroyo, C., Bello-Arqués, P., Pérez-Enguix, D., Martí-Bonmatí, L., & Torres-Espallardo, I. (2024). Y-90 PET/MR imaging optimization with a Bayesian penalized likelihood reconstruction algorithm. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-024-01452-7
Vancouver
Calatayud-Jordán J, Carrasco-Vela N, Chimeno-Hernández J, Carles-Fariña M, Olivas-Arroyo C, Bello-Arqués P, et al. Y-90 PET/MR imaging optimization with a Bayesian penalized likelihood reconstruction algorithm. Physical and Engineering Sciences in Medicine. 2024. doi:10.1007/s13246-024-01452-7
BibTeX
@article{CalatayudJordn2024,
author = {Calatayud-Jordán, José and Carrasco-Vela, Nuria and Chimeno-Hernández, José and Carles-Fariña, Montserrat and Olivas-Arroyo, Consuelo and Bello-Arqués, Pilar and Pérez-Enguix, Daniel and Martí-Bonmatí, Luis and Torres-Espallardo, Irene},
title = {Y-90 PET/MR imaging optimization with a Bayesian penalized likelihood reconstruction algorithm},
journal = {Physical and Engineering Sciences in Medicine},
year = {2024},
doi = {10.1007/s13246-024-01452-7}
}
EUVIMED