EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY 4.0

Development and validation of the effective CNR analysis method for evaluating the contrast resolution of CT images

Kengo Igarashi, Kuniharu Imai, Shigeru Matsushima, Chiyo Yamauchi-Kawaura, Keisuke Fujii

Physical and Engineering Sciences in Medicine · 2024-06 · 47 (2) · 717-727

DOI: 10.1007/s13246-024-01400-5 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Physical and Engineering Sciences in Medicine, online erschienen am 07.03.2024. Verfasst von Kengo Igarashi und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Integrated Health Sciences, Nagoya University Graduate School of Medicine, 1-1-20, Daiko-Minami, Higashi-ku, Nagoya, Aichi, 461-8673, J…. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Signal-To-Noise Ratio, Algorithms, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Nagoya University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Contrast resolution is an important index for evaluating the signal detectability of computed tomographic (CT) images. Recently, various noise reduction algorithms, such as iterative reconstruction (IR) and deep learning reconstruction (DLR), have been proposed to reduce the image noise in CT images. However, these algorithms cause changes in the image noise texture and blurred image signals in CT images. Furthermore, the contrast-to-noise ratio (CNR) cannot be accurately evaluated in CT images reconstructed using noise reduction methods. Therefore, in this study, we devised a new method, namely, “effective CNR analysis,” for evaluating the contrast resolution of CT images. We verified whether the proposed algorithm could evaluate the effective contrast resolution based on the signal detectability of CT images. The findings showed that the effective CNR values obtained using the proposed method correlated well with the subjective visual impressions of CT images. To investigate whether signal detectability was appropriately evaluated using effective CNR analysis, the conventional CNR analysis method was compared with the proposed method. The CNRs of the IR and DLR images calculated using conventional CNR analysis were 13.2 and 10.7, respectively. By contrast, those calculated using the effective CNR analysis were estimated to be 0.7 and 1.1, respectively. Considering that the signal visibility of DLR images was superior to that of IR images, our proposed effective CNR analysis was shown to be appropriate for evaluating the contrast resolution of CT images.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedSignal-To-Noise RatioAlgorithmsImage Processing, Computer-AssistedHumansPhantoms, Imaging

Stichwörter der Autor:innen

Contrast resolutionContrast-to-noise ratioImage noiseStochastic differential equation

Autor:innen und Institutionen

  • Department of Integrated Health Sciences, Nagoya University Graduate School of Medicine, 1-1-20, Daiko-Minami, Higashi-ku, Nagoya, Aichi, 461-8673, Japan. igarashi.kengo.a5@s.mail.nagoya-u.ac.jp.
  • Department of Integrated Health Sciences, Nagoya University Graduate School of Medicine, 1-1-20, Daiko-Minami, Higashi-ku, Nagoya, Aichi, 461-8673, Japan.

Förderung

  • Nagoya University

Zitieren

APA

Igarashi, K., Imai, K., Matsushima, S., Yamauchi-Kawaura, C., & Fujii, K. (2024). Development and validation of the effective CNR analysis method for evaluating the contrast resolution of CT images. Physical and Engineering Sciences in Medicine, 47(2), 717-727. https://doi.org/10.1007/s13246-024-01400-5

Vancouver

Igarashi K, Imai K, Matsushima S, Yamauchi-Kawaura C, Fujii K. Development and validation of the effective CNR analysis method for evaluating the contrast resolution of CT images. Physical and Engineering Sciences in Medicine. 2024;47(2):717-727. doi:10.1007/s13246-024-01400-5

BibTeX
@article{Igarashi2024,
  author = {Igarashi, Kengo and Imai, Kuniharu and Matsushima, Shigeru and Yamauchi-Kawaura, Chiyo and Fujii, Keisuke},
  title = {Development and validation of the effective CNR analysis method for evaluating the contrast resolution of CT images},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2024},
  volume = {47},
  number = {2},
  pages = {717--727},
  doi = {10.1007/s13246-024-01400-5}
}