Variational mode decomposition based ECG denoising using non-local means and wavelet domain filtering
Australasian Physical & Engineering Sciences in Medicine · 2018-12 · 41 (4) · 891-904
DOI: 10.1007/s13246-018-0685-0 ↗Kurz gefasst
Fachartikel in der Zeitschrift Australasian Physical & Engineering Sciences in Medicine, online erschienen am 06.09.2018. Verfasst von Pratik Singh und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electronics and Communication Engineering, National Institute of Technology, Patna, Patna, 800005, India. pratik140871@nitp.ac.in. Hauptthemen laut MEDLINE: Signal Processing, Computer-Assisted. Die Arbeit wurde bisher 75-mal zitiert (Crossref) und nennt 42 Literaturangaben.
Abstract (Originalsprache)
This paper presents a novel electrocardiogram (ECG) denoising approach based on variational mode decomposition (VMD). This work also incorporates the efficacy of the non-local means (NLM) estimation and the discrete wavelet transform (DWT) filtering technique. Current ECG denoising methods fail to remove noise from the entire frequency range of the ECG signal. To achieve the effective ECG denoising goal, the noisy ECG signal is decomposed into narrow-band variational mode functions (VMFs) using VMD method. The idea is to filter out noise from these narrow-band VMFs. To achieve that, the center frequency information associated with each VMFs is used to exclusively divide them into lower- and higher-frequency signal groups. The higher frequency VMFs were filtered out using DWT-thresholding technique. The lower frequency VMFs are denoised through NLM estimation technique. The non-recursive nature of VMD enables the parallel processing of NLM estimation and DWT filtering. The traditional DWT-based approaches need large decomposition levels to filter low frequency noises and at the same time NLM technique suffers from the rare-patch effect in high-frequency region. On the contrary, in the proposed framework both NLM and DWT approaches complement each other to overcome their individual ill-effects. The signal reconstruction is performed using the denoised high frequency and low frequency VMFs. The simulation performed on the MIT-BIH Arrhythmia database shows that the proposed method outperforms the existing state-of-the-art ECG denoising techniques.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electronics and Communication Engineering, National Institute of Technology, Patna, Patna, 800005, India. pratik140871@nitp.ac.in.
- Department of Electronics and Communication Engineering, National Institute of Technology, Patna, Patna, 800005, India.
Zitieren
APA
Singh, P., & Pradhan, G. (2018). Variational mode decomposition based ECG denoising using non-local means and wavelet domain filtering. Australasian Physical & Engineering Sciences in Medicine, 41(4), 891-904. https://doi.org/10.1007/s13246-018-0685-0
Vancouver
Singh P, Pradhan G. Variational mode decomposition based ECG denoising using non-local means and wavelet domain filtering. Australasian Physical & Engineering Sciences in Medicine. 2018;41(4):891-904. doi:10.1007/s13246-018-0685-0
BibTeX
@article{Singh2018,
author = {Singh, Pratik and Pradhan, Gayadhar},
title = {Variational mode decomposition based ECG denoising using non-local means and wavelet domain filtering},
journal = {Australasian Physical & Engineering Sciences in Medicine},
year = {2018},
volume = {41},
number = {4},
pages = {891--904},
doi = {10.1007/s13246-018-0685-0}
}
EUVIMED