High-Fidelity Modeling and in Vitro Validation of a Novel NiTi Stent Retriever
Cardiovascular Engineering and Technology · 09.11.2026
DOI: 10.1007/s13239-026-00869-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, online erschienen am 09.11.2026. Verfasst von Shuo Wu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering and Automation, Robotics Institute, Beihang University, Road Xueyuan 37, Beijing, 100191, China. Die Arbeit nennt 51 Literaturangaben. Gefördert von: National Natural Science Foundation of China, National Plan for NRRP, China Scholarship Council.
Abstract (Originalsprache)
PURPOSE: Nickel-titanium (NiTi) stent retrievers leverage superelasticity to navigate tortuous vascular anatomies, yet their intricate geometries pose significant challenges for accurate finite element (FE) modeling. This study evaluates the influence of mesh type and design representation on the predictive fidelity of FE simulations for a novel NiTi stent retriever. METHODS: NiTi tube properties for a novel stent retriever prototype were characterized. FE models were developed using tetrahedral and hexahedral solid elements, as well as beam elements with alternative junction representations. Lateral compression, axial tension, and bending simulations were validated against experiments. The selected beam formulation was then incorporated into a full-process model of crimping, delivery, deployment, and retrieval, with in vitro validation in straight, L-shaped, and U-shaped glass vessel models. RESULTS: The hexahedral solid model showed the closest agreement with device-level experiments and most clearly resolved local stress and strain concentrations. Among the beam formulations, the T-type junction model preserved the primary load-transfer path and provided an efficient representation for full-process simulations. After interface calibration, the full-process model reproduced deployment configuration, retrieval-force evolution, and procedure-dependent deformation. In the U-shaped model, the predicted mean retrieval force differed from the experiment by 2.1%, and peak axial strain remained below 8%. CONCLUSION: The discretization strategy affected both device-level mechanical prediction and procedure-level retrieval response. The mechanically validated and interface-calibrated beam model provides an efficient framework for simulating NiTi stent retriever procedures in rigid in vitro vessel models. The identified interface coefficient may serve as a reference for similar NiTi-glass retrieval simulations.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechanical Engineering and Automation, Robotics Institute, Beihang University, Road Xueyuan 37, Beijing, 100191, China.
- Department of Chemistry, Materials and Chemical Engineering, Politecnico di Milano, Piazza L. da Vinci 32, 20133, Milan, Italy.
- Department of Mechanical Engineering and Automation, Robotics Institute, Beihang University, Road Xueyuan 37, Beijing, 100191, China. shengnan_lyu@buaa.edu.cn.
- Fondazione IRCCS Cà Granda Ospedale Maggiore Policlinico, Milan, Italy.
Förderung
- National Natural Science Foundation of China
- National Plan for NRRP
- China Scholarship Council
Zitieren
APA
Wu, S., Fregona, V., Quartieri, V., Lyu, S., Migliavacca, F., Luraghi, G., Zhang, T., & Berti, F. (2026). High-Fidelity Modeling and in Vitro Validation of a Novel NiTi Stent Retriever. Cardiovascular Engineering and Technology. https://doi.org/10.1007/s13239-026-00869-3
Vancouver
Wu S, Fregona V, Quartieri V, Lyu S, Migliavacca F, Luraghi G, et al. High-Fidelity Modeling and in Vitro Validation of a Novel NiTi Stent Retriever. Cardiovascular Engineering and Technology. 2026. doi:10.1007/s13239-026-00869-3
BibTeX
@article{Wu2026,
author = {Wu, Shuo and Fregona, Virginia and Quartieri, Valentina and Lyu, Shengnan and Migliavacca, Francesco and Luraghi, Giulia and Zhang, Tianxiao and Berti, Francesca},
title = {High-Fidelity Modeling and in Vitro Validation of a Novel NiTi Stent Retriever},
journal = {Cardiovascular Engineering and Technology},
year = {2026},
doi = {10.1007/s13239-026-00869-3}
}
EUVIMED