Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring Systems

Fabian Andres Castaño, Alher Mauricio Hernández

Cardiovascular Engineering and Technology · 2026-10 · 17 (5) · 537-552

DOI: 10.1007/s13239-026-00850-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, online erschienen am 03.08.2026. Verfasst von Fabian Andres Castaño und 1 weiteren Autor:innen. Beteiligte Einrichtung: Bioinstrumentation and Clinical Engineering Research Group - GIBIC, Bioengineering Department, Engineering Faculty, Universidad de Antioquia UdeA, Ca…. Hauptthemen laut MEDLINE: Artifacts, Photoplethysmography, Oxygen Saturation, Signal Processing, Computer-Assisted. Die Arbeit nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Antioquia.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and objective Chronic cardiovascular diseases have motivated the development of wearable systems capable of continuously monitoring physiological variables such as oxygen saturation (SpO $$_2$$ 2 ) and non-invasive blood pressure (NIBP). However, motion artifacts remain one of the main limitations affecting the reliability of these measurements in ambulatory environments. This article presents a novel framework for motion artifact reduction in photoplethysmography (PPG) and NIBP signals based on redundant sensing, inertial measurements, ECG-guided temporal synchrony, and multichannel signal processing.

Methods

The proposed framework combines independent component analysis and recursive least-squares adaptive filtering to separate physiological information from motion-induced interference. Performance was compared with conventional finite impulse response (FIR) filtering and wavelet shrinkage (WS) methods using signal-to-noise ratio (SNR), weighted distortion assessment (WDA), oxygen saturation estimation, and blood pressure estimation metrics.

Results

The proposed method achieved statistically significant improvements in signal quality and physiological parameter estimation. Average SNR improved from $$-5.93\pm 5.05$$ - 5.93 ± 5.05 dB to $$-3.43\pm 2.02$$ - 3.43 ± 2.02 dB in the red PPG channel ( $$p=0.02$$ p = 0.02 ), while the WDA index increased from 0.48 to 0.73. Oxygen saturation estimation improved from $$80.87\pm 9.09\%$$ 80.87 ± 9.09 % in motion-contaminated signals to $$94.16\pm 2.72\%$$ 94.16 ± 2.72 % after denoising ( $$p=0.04$$ p = 0.04 ), approaching resting measurements ( $$95.23\pm 2.31\%$$ 95.23 ± 2.31 % ). For NIBP signals, statistically significant improvements were also observed ( $$p=0.045$$ p = 0.045 ), particularly for systolic pressure estimation, although diastolic pressure remained affected by residual motion-related distortions.

Conclusions

The proposed framework improves the robustness of wearable PPG and NIBP monitoring under motion conditions through the combined use of sensor redundancy, ECG-guided temporal coupling, inertial measurements, and adaptive multichannel signal processing. The results demonstrate significant improvements in signal quality and physiological parameter estimation compared with conventional denoising approaches, supporting the potential use of the method in ambulatory cardiovascular monitoring applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ArtifactsPhotoplethysmographyOxygen SaturationSignal Processing, Computer-AssistedWearable Electronic DevicesBlood PressureBlood Pressure DeterminationBlood Pressure Monitoring, AmbulatoryHumansReproducibility of ResultsPredictive Value of TestsSignal-To-Noise RatioElectrocardiographyFemaleMotionMaleTime FactorsEquipment DesignBiomarkersDigital HealthAdaptive Algorithms

Stichwörter der Autor:innen

Biomedical signal processingHome health careMotion artifactsNon-invasive blood pressurePhotoplethysmographySignal denoising

Substanzen

Biomarkers

Autor:innen und Institutionen

  • Bioinstrumentation and Clinical Engineering Research Group - GIBIC, Bioengineering Department, Engineering Faculty, Universidad de Antioquia UdeA, Calle 70 No. 52-21, Medellín, 050010, Colombia. fabian.castano@udea.edu.…
  • Physics Institute,Exact and Natural Sciences Faculty, Universidad de Antioquia UdeA, Calle 70 No. 52-21, Medellín, 050010, Colombia. fabian.castano@udea.edu.co.
  • Bioinstrumentation and Clinical Engineering Research Group - GIBIC, Bioengineering Department, Engineering Faculty, Universidad de Antioquia UdeA, Calle 70 No. 52-21, Medellín, 050010, Colombia.

Förderung

  • University of Antioquia

Zitieren

APA

Castaño, F. A., & Hernández, A. M. (2026). Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring Systems. Cardiovascular Engineering and Technology, 17(5), 537-552. https://doi.org/10.1007/s13239-026-00850-0

Vancouver

Castaño FA, Hernández AM. Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring Systems. Cardiovascular Engineering and Technology. 2026;17(5):537-552. doi:10.1007/s13239-026-00850-0

BibTeX
@article{Castao2026,
  author = {Castaño, Fabian Andres and Hernández, Alher Mauricio},
  title = {Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring Systems},
  journal = {Cardiovascular Engineering and Technology},
  year = {2026},
  volume = {17},
  number = {5},
  pages = {537--552},
  doi = {10.1007/s13239-026-00850-0}
}
Kostenlos starten →