Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring Systems
Cardiovascular Engineering and Technology · 2026-10 · 17 (5) · 537-552
DOI: 10.1007/s13239-026-00850-0 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, online erschienen am 03.08.2026. Verfasst von Fabian Andres Castaño und 1 weiteren Autor:innen. Beteiligte Einrichtung: Bioinstrumentation and Clinical Engineering Research Group - GIBIC, Bioengineering Department, Engineering Faculty, Universidad de Antioquia UdeA, Ca…. Hauptthemen laut MEDLINE: Artifacts, Photoplethysmography, Oxygen Saturation, Signal Processing, Computer-Assisted. Die Arbeit nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Antioquia.
Abstract (Originalsprache)
Background and objective Chronic cardiovascular diseases have motivated the development of wearable systems capable of continuously monitoring physiological variables such as oxygen saturation (SpO $$_2$$ 2 ) and non-invasive blood pressure (NIBP). However, motion artifacts remain one of the main limitations affecting the reliability of these measurements in ambulatory environments. This article presents a novel framework for motion artifact reduction in photoplethysmography (PPG) and NIBP signals based on redundant sensing, inertial measurements, ECG-guided temporal synchrony, and multichannel signal processing.
Methods
The proposed framework combines independent component analysis and recursive least-squares adaptive filtering to separate physiological information from motion-induced interference. Performance was compared with conventional finite impulse response (FIR) filtering and wavelet shrinkage (WS) methods using signal-to-noise ratio (SNR), weighted distortion assessment (WDA), oxygen saturation estimation, and blood pressure estimation metrics.
Results
The proposed method achieved statistically significant improvements in signal quality and physiological parameter estimation. Average SNR improved from $$-5.93\pm 5.05$$ - 5.93 ± 5.05 dB to $$-3.43\pm 2.02$$ - 3.43 ± 2.02 dB in the red PPG channel ( $$p=0.02$$ p = 0.02 ), while the WDA index increased from 0.48 to 0.73. Oxygen saturation estimation improved from $$80.87\pm 9.09\%$$ 80.87 ± 9.09 % in motion-contaminated signals to $$94.16\pm 2.72\%$$ 94.16 ± 2.72 % after denoising ( $$p=0.04$$ p = 0.04 ), approaching resting measurements ( $$95.23\pm 2.31\%$$ 95.23 ± 2.31 % ). For NIBP signals, statistically significant improvements were also observed ( $$p=0.045$$ p = 0.045 ), particularly for systolic pressure estimation, although diastolic pressure remained affected by residual motion-related distortions.
Conclusions
The proposed framework improves the robustness of wearable PPG and NIBP monitoring under motion conditions through the combined use of sensor redundancy, ECG-guided temporal coupling, inertial measurements, and adaptive multichannel signal processing. The results demonstrate significant improvements in signal quality and physiological parameter estimation compared with conventional denoising approaches, supporting the potential use of the method in ambulatory cardiovascular monitoring applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Bioinstrumentation and Clinical Engineering Research Group - GIBIC, Bioengineering Department, Engineering Faculty, Universidad de Antioquia UdeA, Calle 70 No. 52-21, Medellín, 050010, Colombia. fabian.castano@udea.edu.…
- Physics Institute,Exact and Natural Sciences Faculty, Universidad de Antioquia UdeA, Calle 70 No. 52-21, Medellín, 050010, Colombia. fabian.castano@udea.edu.co.
- Bioinstrumentation and Clinical Engineering Research Group - GIBIC, Bioengineering Department, Engineering Faculty, Universidad de Antioquia UdeA, Calle 70 No. 52-21, Medellín, 050010, Colombia.
Förderung
- University of Antioquia
Zitieren
APA
Castaño, F. A., & Hernández, A. M. (2026). Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring Systems. Cardiovascular Engineering and Technology, 17(5), 537-552. https://doi.org/10.1007/s13239-026-00850-0
Vancouver
Castaño FA, Hernández AM. Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring Systems. Cardiovascular Engineering and Technology. 2026;17(5):537-552. doi:10.1007/s13239-026-00850-0
BibTeX
@article{Castao2026,
author = {Castaño, Fabian Andres and Hernández, Alher Mauricio},
title = {Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring Systems},
journal = {Cardiovascular Engineering and Technology},
year = {2026},
volume = {17},
number = {5},
pages = {537--552},
doi = {10.1007/s13239-026-00850-0}
}
EUVIMED