Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, online erschienen am 13.07.2026. Verfasst von Lena Iam Spitz und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Simulation and Graphics, Otto-von-Guericke University, Magdeburg, Germany. spitz@isg.cs.ovgu.de. Hauptthemen laut MEDLINE: Intracranial Aneurysm, Cerebrovascular Circulation. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft, Bundesministerium für Bildung und Forschung, Otto-von-Guericke-Universität Magdeburg.
Abstract (Originalsprache)
Purpose
The Contour Neurovascular System (CNS) is a novel intrasaccular flow disruptor for the endovascular treatment of wide-necked aneurysms. While more longitudinal studies are needed to fully assess its efficacy, initial study results over the past few years have been promising. However, no studies have yet assessed the CNS’s robustness to deformation over time, and few have assessed its effect on the morphology of intracranial aneurysms.
Methods
Thirteen IAs were retrospectively analyzed using co-registered 3D meshes from serial imaging (2–5 scans per patient, up to 26 months of follow-up). Morphological parameters of the ostium and CNS were compared across defined time points. Intra-patient variance was also assessed through scans taken on the same day, and morphology was correlated with treatment outcome.
Results
Significant postprocedural changes were observed in the ostium’s ellipticity index (EI; $$\Delta$$ Δ = −0.07 ±0.06, p = 0.01, d = −1.18) and neck circularity index (NCI; $$\Delta$$ Δ = +0.04 ±0.04, p = 0.04, d = 0.90). No statistically significant long-term changes were detected for either the ostium or CNS. Most variations were within intra-patient variance, and many measured deformations were below imaging resolution (0.16−0.47 mm). No correlation was found between recurrence and morphology.
Conclusion
Only immediate postprocedural shape changes of the ostium were significant. No measurable long-term deformation of the CNS or ostium was observed. The CNS appears morphologically stable, though larger, high-resolution studies are needed to confirm these findings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Simulation and Graphics, Otto-von-Guericke University, Magdeburg, Germany. spitz@isg.cs.ovgu.de.
- Research campus STIMULATE, Magdeburg, Germany. spitz@isg.cs.ovgu.de.
- Research campus STIMULATE, Magdeburg, Germany.
- Laboratory of Fluid Dynamics and Technical Flows, Otto-von-Guericke University, Magdeburg, Germany.
- Department of Radiology and Neuroradiology, University Medical Center Schleswig-Holstein, Campus Kiel, Germany.
- University Clinic for Neuroradiology, Otto-von-Guericke University, Magdeburg, Germany.
Förderung
- Deutsche Forschungsgemeinschaft
- Bundesministerium für Bildung und Forschung
- Otto-von-Guericke-Universität Magdeburg
Zitieren
APA
Spitz, L. I., Korte, J., Gaidzik, F., Larsen, N., Behme, D., Preim, B., & Saalfeld, S. (2026). Morphological Stability of Intracranial Aneurysms and the Contour Neurovascular System: A Longitudinal Analysis. Cardiovascular Engineering and Technology, 17(5), 579-590. https://doi.org/10.1007/s13239-026-00847-9
Vancouver
Spitz LI, Korte J, Gaidzik F, Larsen N, Behme D, Preim B, et al. Morphological Stability of Intracranial Aneurysms and the Contour Neurovascular System: A Longitudinal Analysis. Cardiovascular Engineering and Technology. 2026;17(5):579-590. doi:10.1007/s13239-026-00847-9
BibTeX
@article{Spitz2026,
author = {Spitz, Lena Iam and Korte, Jana and Gaidzik, Franziska and Larsen, Naomi and Behme, Daniel and Preim, Bernhard and Saalfeld, Sylvia},
title = {Morphological Stability of Intracranial Aneurysms and the Contour Neurovascular System: A Longitudinal Analysis},
journal = {Cardiovascular Engineering and Technology},
year = {2026},
volume = {17},
number = {5},
pages = {579--590},
doi = {10.1007/s13239-026-00847-9}
}
EUVIMED