Optimizing Non-invasive Fractional Flow Reserve Estimation with Machine Learning-Enhanced 1D Hemodynamic Modeling
Cardiovascular Engineering and Technology · 2026-08 · 17 (4) · 425-438
DOI: 10.1007/s13239-026-00836-y ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Cardiovascular Engineering and Technology, online erschienen am 29.04.2026. Verfasst von Cyrus Tanade und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Duke University, 534 Research Dr., Durham, NC, 27705, USA. Hauptthemen laut MEDLINE: Fractional Flow Reserve, Myocardial, Coronary Angiography, Models, Cardiovascular, Coronary Artery Disease. Die Arbeit nennt 72 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIA NIH HHS, National Science Foundation, National Institute on Aging.
Abstract (Originalsprache)
Patient-specific computational models exhibit strong concordance with invasively measured fractional flow reserve (FFR)—the clinical gold standard for diagnosing coronary ischemia. However, current modeling techniques frequently rely on computationally intensive assumptions such as pulsatile flow dynamics and often fail to optimally leverage patient-specific clinical data that is routinely available, limiting their practical clinical adoption. In this study, we propose a hybrid coronary angiography-based approach that reduces computational complexity through simplified steady-state flow assumptions, while simultaneously better leveraging available clinical information. Specifically, we integrate physics–based modeling with a machine learning (ML) feedback loop designed to refine and improve FFR predictions. We evaluated this hybrid framework using a retrospective two-center cohort comprising 132 patients with 132 coronary lesions. Our results demonstrate that steady-state models effectively capture essential hemodynamic patterns, closely matching pulsatile model predictions. The ML refinement step enhances diagnostic accuracy, yielding a sensitivity of 83.3%, specificity of 100.0%, positive predictive value of 100.0%, negative predictive value of 88.2%, and overall precision of 92.6%. By effectively combining efficient computational modeling with targeted ML-driven refinements, our approach represents a robust, clinically viable solution for accurate patient-specific FFR estimation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Duke University, 534 Research Dr., Durham, NC, 27705, USA.
- Department of Biomedical Engineering, Duke University, 534 Research Dr., Durham, NC, 27705, USA. amanda.randles@duke.edu.
Förderung
- NIA NIH HHS
- National Science Foundation
- National Institute on Aging
Zitieren
APA
Tanade, C., Mavi, J. K., Ferreira, G., Schwaller, S., & Randles, A. (2026). Optimizing Non-invasive Fractional Flow Reserve Estimation with Machine Learning-Enhanced 1D Hemodynamic Modeling. Cardiovascular Engineering and Technology, 17(4), 425-438. https://doi.org/10.1007/s13239-026-00836-y
Vancouver
Tanade C, Mavi JK, Ferreira G, Schwaller S, Randles A. Optimizing Non-invasive Fractional Flow Reserve Estimation with Machine Learning-Enhanced 1D Hemodynamic Modeling. Cardiovascular Engineering and Technology. 2026;17(4):425-438. doi:10.1007/s13239-026-00836-y
BibTeX
@article{Tanade2026,
author = {Tanade, Cyrus and Mavi, Japneet Kaur and Ferreira, Guinevere and Schwaller, Sam and Randles, Amanda},
title = {Optimizing Non-invasive Fractional Flow Reserve Estimation with Machine Learning-Enhanced 1D Hemodynamic Modeling},
journal = {Cardiovascular Engineering and Technology},
year = {2026},
volume = {17},
number = {4},
pages = {425--438},
doi = {10.1007/s13239-026-00836-y}
}
EUVIMED