Flow-Based Coronary Artery Bypass Graft Patency Metrics: Uncertainty Quantification Simulations to Guide Development
Cardiovascular Engineering and Technology · 2025
DOI: 10.1007/s13239-024-00765-8 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, 2025. Verfasst von Sita Drost und 1 weiteren Autor:innen. Beteiligte Einrichtung: Transonic Europe B.V., Business Park Stein 205, Elsloo, 6181 MB, The Netherlands. sita.drost@transonic.com. Hauptthemen laut MEDLINE: Coronary Artery Bypass, Vascular Patency, Models, Cardiovascular, Computer Simulation. Die Arbeit nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Over time, transit time flow measurement (TTFM) has proven itself as a simple and effective tool for intra-operative evaluation of coronary artery bypass grafts (CABGs). However, metrics used to screen for possible technical error show considerable spread, preventing the definition of sharp cut-off values to distinguish between patent, questionable, and failed grafts. The simulation study presented in this paper aims to quantify this uncertainty for commonly used patency metrics, and to identify the most important physiological parameters influencing it.
Methods
Uncertainty quantification was performed on a realistic multiscale numerical model of the coronary circulation, guided by Morris screening sensitivity analysis of a simpler, lumped-parameter model. Simulation results were qualitatively verified against results of a recent clinical study.
Results
Correspondence with clinical study data is reasonable, especially considering that the model was not fitted in any way. Stenosis severity was confirmed to be an influential parameter. However, also cardiac period and graft diameter were observed to be important, particularly for mean flow rate and pulsatility index.
Conclusion
Metrics quantifying the flow waveform’s diastolic dominance show the highest sensitivity to graft stenosis, and seem to be least affected by autoregulation. Among these, the novel diastolic resistance index shows the strongest sensitivity to stenosis severity. Significance The approach used in this study is expected to benefit the development of improved patency metrics, by allowing medical engineers to include sensitivity and uncertainty in assessing, in-silico, the potential of novel metrics, thus enabling them to provide better guidance in the design of clinical studies.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Transonic Europe B.V., Business Park Stein 205, Elsloo, 6181 MB, The Netherlands. sita.drost@transonic.com.
- Transonic Systems Inc., 34 Dutch Mill Road, Ithaca, New York, 14850, USA.
ORCID-Profile: Sita Drost
Zitieren
APA
Drost, S., & Drost, C. J. (2025). Flow-Based Coronary Artery Bypass Graft Patency Metrics: Uncertainty Quantification Simulations to Guide Development. Cardiovascular Engineering and Technology. https://doi.org/10.1007/s13239-024-00765-8
Vancouver
Drost S, Drost CJ. Flow-Based Coronary Artery Bypass Graft Patency Metrics: Uncertainty Quantification Simulations to Guide Development. Cardiovascular Engineering and Technology. 2025. doi:10.1007/s13239-024-00765-8
BibTeX
@article{Drost2025,
author = {Drost, Sita and Drost, Cornelis J},
title = {Flow-Based Coronary Artery Bypass Graft Patency Metrics: Uncertainty Quantification Simulations to Guide Development},
journal = {Cardiovascular Engineering and Technology},
year = {2025},
doi = {10.1007/s13239-024-00765-8}
}
EUVIMED