Design and Characterisation of a Novel Z-Shaped Inductor-Based Wireless Implantable Sensor for Surveillance of Abdominal Aortic Aneurysm Post-Endovascular Repair
Cardiovascular Engineering and Technology · 2025-02 · 16 (1) · 1-19
DOI: 10.1007/s13239-024-00753-y ↗Kurz gefasst
Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, online erschienen am 07.10.2024. Verfasst von Nuno P Silva und 4 weiteren Autor:innen. Beteiligte Einrichtung: Translational Medical Device Lab, University of Galway, Galway, H91 TK33, Ireland. n.silva1@universityofgalway.ie. Hauptthemen laut MEDLINE: Aortic Aneurysm, Abdominal, Endovascular Aneurysm Repair, Equipment Design, Wireless Technology. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 51 Literaturangaben. Gefördert von: Science Foundation Ireland.
Abstract (Originalsprache)
PURPOSE: An abdominal aortic aneurysm (AAA) is a dilation of the aorta over its normal diameter (> 3 cm). The minimally invasive treatment adopted uses a stent graft to be deployed into the aneurysm by a catheter to flow blood through it. However, this approach demands frequent monitoring using imaging modalities that involve radiation and contrast agents. Moreover, the multiple follow-ups are expensive, time-consuming, and resource-demanding for healthcare systems. This study proposes a novel wireless implantable medical sensor (WIMS) to measure the aneurysm growth after the endovascular repair. METHODS: The proposed sensor is composed of a Z-shaped inductor, similar to a stent ring. The proposed design of the sensor is explored by investigating the inductance, resistance, and quality factor of different possible geometries related to a Z-shaped configuration, such as the height and number of struts. The study is conducted through a combination of numerical simulations and experimental tests, with the assessment being carried out at a frequency of 13.56 MHz. RESULTS: The results show that a higher number of struts result in higher values of inductance and resistance. On the other hand, the increase in the number of struts decreases the quality factor of the Z-shaped inductor due to the presence of high resistance from the inductor. Moreover, it is observed that the influence of the number of struts present in the Z-shaped inductor tends to decrease for larger radii. CONCLUSIONS: The numerical and experimental evaluation concludes the ability of the proposed sensor to measure the size of the aneurysm.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Translational Medical Device Lab, University of Galway, Galway, H91 TK33, Ireland. n.silva1@universityofgalway.ie.
- Electrical and Electronic Engineering, University of Galway, Galway, H91 TK33, Ireland. n.silva1@universityofgalway.ie.
- Translational Medical Device Lab, University of Galway, Galway, H91 TK33, Ireland.
- Electrical and Electronic Engineering, University of Galway, Galway, H91 TK33, Ireland.
- School of Medicine, University of Galway, Galway, H91 TK33, Ireland.
ORCID-Profile: Nuno P Silva
Förderung
- Science Foundation Ireland
Zitieren
APA
Silva, N. P., Amin, B., Dunne, E., O'Halloran, M., & Elahi, A. (2025). Design and Characterisation of a Novel Z-Shaped Inductor-Based Wireless Implantable Sensor for Surveillance of Abdominal Aortic Aneurysm Post-Endovascular Repair. Cardiovascular Engineering and Technology, 16(1), 1-19. https://doi.org/10.1007/s13239-024-00753-y
Vancouver
Silva NP, Amin B, Dunne E, O'Halloran M, Elahi A. Design and Characterisation of a Novel Z-Shaped Inductor-Based Wireless Implantable Sensor for Surveillance of Abdominal Aortic Aneurysm Post-Endovascular Repair. Cardiovascular Engineering and Technology. 2025;16(1):1-19. doi:10.1007/s13239-024-00753-y
BibTeX
@article{Silva2025,
author = {Silva, Nuno P and Amin, Bilal and Dunne, Eoghan and O'Halloran, Martin and Elahi, Adnan},
title = {Design and Characterisation of a Novel Z-Shaped Inductor-Based Wireless Implantable Sensor for Surveillance of Abdominal Aortic Aneurysm Post-Endovascular Repair},
journal = {Cardiovascular Engineering and Technology},
year = {2025},
volume = {16},
number = {1},
pages = {1--19},
doi = {10.1007/s13239-024-00753-y}
}