An IMU-Based Wearable System for Respiratory Rate Estimation in Static and Dynamic Conditions
Cardiovascular Engineering and Technology · 2023-06 · 14 (3) · 351-363
DOI: 10.1007/s13239-023-00657-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, online erschienen am 27.02.2023. Verfasst von Alessandra Angelucci und 1 weiteren Autor:innen. Beteiligte Einrichtung: Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Piazza Leonardo Da Vinci 32, 20133, Milan, Italy. alessandra.angelu…. Hauptthemen laut MEDLINE: Respiratory Rate, Wearable Electronic Devices. Die Arbeit wurde bisher 40-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: e-Novia S.p.A., Politecnico di Milano.
Abstract (Originalsprache)
Purpose
Breathing parameters change with activity and posture, but currently available solutions can perform measurements only during static conditions.
Methods
This article presents an innovative wearable sensor system constituted by three inertial measurement units to simultaneously estimate respiratory rate (RR) in static and dynamic conditions and perform human activity recognition (HAR) with the same sensing principle. Two units are aimed at detecting chest wall breathing-related movements (one on the thorax, one on the abdomen); the third is on the lower back. All units compute the quaternions describing the subject’s movement and send data continuously with the ANT transmission protocol to an app. The 20 healthy subjects involved in the research (9 men, 11 women) were between 23 and 54 years old, with mean age 26.8, mean height 172.5 cm and mean weight 66.9 kg. Data from these subjects during different postures or activities were collected and analyzed to extract RR.
Results
Statistically significant differences between dynamic activities (“walking slow”, “walking fast”, “running” and “cycling”) and static postures were detected ( p < 0.05), confirming the obtained measurements are in line with physiology even during dynamic activities. Data from the reference unit only and from all three units were used as inputs to artificial intelligence methods for HAR. When the data from the reference unit were used, the Gated Recurrent Unit was the best performing method (97% accuracy). With three units, a 1D Convolutional Neural Network was the best performing (99% accuracy).
Conclusion
Overall, the proposed solution shows it is possible to perform simultaneous HAR and RR measurements in static and dynamic conditions with the same sensor system.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Piazza Leonardo Da Vinci 32, 20133, Milan, Italy. alessandra.angelucci@polimi.it.
- Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Piazza Leonardo Da Vinci 32, 20133, Milan, Italy.
ORCID-Profile: Alessandra Angelucci
Förderung
- e-Novia S.p.A.
- Politecnico di Milano
Zitieren
APA
Angelucci, A., & Aliverti, A. (2023). An IMU-Based Wearable System for Respiratory Rate Estimation in Static and Dynamic Conditions. Cardiovascular Engineering and Technology, 14(3), 351-363. https://doi.org/10.1007/s13239-023-00657-3
Vancouver
Angelucci A, Aliverti A. An IMU-Based Wearable System for Respiratory Rate Estimation in Static and Dynamic Conditions. Cardiovascular Engineering and Technology. 2023;14(3):351-363. doi:10.1007/s13239-023-00657-3
BibTeX
@article{Angelucci2023,
author = {Angelucci, Alessandra and Aliverti, Andrea},
title = {An IMU-Based Wearable System for Respiratory Rate Estimation in Static and Dynamic Conditions},
journal = {Cardiovascular Engineering and Technology},
year = {2023},
volume = {14},
number = {3},
pages = {351--363},
doi = {10.1007/s13239-023-00657-3}
}
EUVIMED