EUVIMEDEUVIMED entdecken
Crossref

Effects of source-to-image distance and dose mode on image quality and radiation dose in high-definition (Hi-Def) angiography

Kunihito Tsuboi, Hiroshi Matsushima, Takamasa Kanbe, Yumiko Iguchi, Ken Tanikawa, Masanori Kaneko

Radiological Physics and Technology · 2026

DOI: 10.1007/s12194-026-01079-1 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiological Physics and Technology, 2026. Verfasst von Kunihito Tsuboi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Central Radiology, Gifu Prefectural Gero Hospital, 2211 Mori, Gero, Gifu, 509-2292, Japan. tsuboi-kunihito@gero-hp.jp. Hauptthemen laut MEDLINE: Radiation Dosage, Angiography, Image Processing, Computer-Assisted. Die Arbeit nennt 12 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study evaluated the effects of source-to-image distance (SID) and dose mode on image quality and radiation dose in a high-definition (Hi-Def) angiography compared with conventional flat-panel detector (FPD). A phantom study was performed with varying SID (90–120 cm) and dose mode (low, normal, high). Fluoroscopy images were assessed using standard deviation (SD), contrast-to-noise ratio (CNR), air kerma rate, and spatial resolution (visual assessment). SD increased with SID in both systems, but Hi-Def exhibited more stable noise performance compared with FPD. Increasing SID reduced spatial resolution and increased air kerma rate in Hi-Def. CNR improved with increasing dose mode, whereas spatial resolution showed no significant differences. These results demonstrate that Hi-Def imaging exhibits unique characteristics, including increased radiation dose and SID-sensitive spatial resolution. Optimizing SID and dose mode is essential to preserve spatial resolution, avoid unnecessary radiation exposure, and achieve a balance between image quality and dose in Hi-Def.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation DosageAngiographyImage Processing, Computer-AssistedPhantoms, ImagingSignal-To-Noise RatioFluoroscopy

Stichwörter der Autor:innen

AngiographyHi-DefRadiation doseResolutionSIDX-ray imaging

Autor:innen und Institutionen

  • Department of Central Radiology, Gifu Prefectural Gero Hospital, 2211 Mori, Gero, Gifu, 509-2292, Japan. tsuboi-kunihito@gero-hp.jp.
  • Department of Central Radiology, Gifu Prefectural Gero Hospital, 2211 Mori, Gero, Gifu, 509-2292, Japan.

Zitieren

APA

Tsuboi, K., Matsushima, H., Kanbe, T., Iguchi, Y., Tanikawa, K., & Kaneko, M. (2026). Effects of source-to-image distance and dose mode on image quality and radiation dose in high-definition (Hi-Def) angiography. Radiological Physics and Technology. https://doi.org/10.1007/s12194-026-01079-1

Vancouver

Tsuboi K, Matsushima H, Kanbe T, Iguchi Y, Tanikawa K, Kaneko M. Effects of source-to-image distance and dose mode on image quality and radiation dose in high-definition (Hi-Def) angiography. Radiological Physics and Technology. 2026. doi:10.1007/s12194-026-01079-1

BibTeX
@article{Tsuboi2026,
  author = {Tsuboi, Kunihito and Matsushima, Hiroshi and Kanbe, Takamasa and Iguchi, Yumiko and Tanikawa, Ken and Kaneko, Masanori},
  title = {Effects of source-to-image distance and dose mode on image quality and radiation dose in high-definition (Hi-Def) angiography},
  journal = {Radiological Physics and Technology},
  year = {2026},
  doi = {10.1007/s12194-026-01079-1}
}
Kostenlos starten →