Determination of cortisol cut-off limits and steroid dynamics in the ACTH stimulation test: a comparative analysis using Roche Elecsys Cortisol II immunoassay and LC-MS/MS
Endocrine · 09.03.2024 · 85 (1) · 321-330
DOI: 10.1007/s12020-024-03752-0 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Endocrine, online erschienen am 09.03.2024. Verfasst von Sema Okutan und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Endocrinology and Metabolism, Copenhagen University Hospital, Rigshospitalet, Copenhagen, Denmark. Hauptthemen laut MEDLINE: Hydrocortisone, Adrenocorticotropic Hormone, Tandem Mass Spectrometry. Die Arbeit wurde bisher 27-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Novo Nordisk.
Abstract (Originalsprache)
Purpose
Measurement of cortisol concentrations is method dependent. The study aimed to establish assay-specific cut-off limits for cortisol after adrenocorticotropic hormone (ACTH) stimulation, comparing Roche Elecsys Cortisol II immunoassay to liquid chromatography-mass spectrometry (LC-MS/MS), and to assess the impact of patient characteristics, estrogen containing oral contraceptives as well as relation to other adrenocortical steroid hormone dynamics.
Methods
One hundred healthy participants underwent a 250 μg ACTH-test, with plasma samples analyzed using ElecsysCortI, ElecsysCortII, and LC-MS/MS. Cortisone, corticosterone, 17-OH-progesterone, dehydroepiandrosterone sulfate (DHEAS), androstenedione, and testosterone were additionally analyzed with LC-MS/MS. Cut-off limit for a normal cortisol response to the ACTH-test was defined as: 2.5th percentile–1.96 × SE.
Results
ElecsysCort II measured cortisol concentrations 21% (95% CI: 19–22%) lower than ElecsysCort I. Cut-off limits for cortisol 30 and 60 min after ACTH were 426 and 485 nmol/L (ElecsysCort II) and 411 and 470 nmol/L (LC-MS/MS). Cut-offs were unaffected by gender, or body-composition. The ACTH-test resulted in significantly increased adrenocortical steroid hormones, except for decreased cortisone concentrations (both sexes), and decreased testosterone in men (1.9 nmol/L, 95% CI: 1.3–2.5). Testosterone was increased in women (0.07 nmol/L, 95% CI: 0.02–0.13).
Conclusion
ElecsysCort II has high analytical performance and yields significantly lower cortisol concentrations than prior polyclonal immunoassays. This clinically relevant difference underscores the necessity for revised cut-off limits for improved diagnostic precision. Suggested 30-minute cortisol cutoff limits are 411 nmol/L (LC-MS/MS) and 426 nmol/L (ElecsysCort II). Adrenocortical steroids increased upon ACTH stimulation, except for cortisone in both sexes and testosterone in men, both of which decreased.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Endocrinology and Metabolism, Copenhagen University Hospital, Rigshospitalet, Copenhagen, Denmark.
- Department of Clinical Medicine, Faculty of Health and Medical Sciences, Copenhagen University, Copenhagen, Denmark.
- Department of Clinical Biochemistry, Næstved, Slagelse and Ringsted Hospitals, Slagelse, Denmark.
- Department of Clinical Biochemistry, Copenhagen University Hospital, Rigshospitalet, Copenhagen, Denmark.
- Department of Clinical Biochemistry, Copenhagen University Hospital, Bispebjerg Hospital, Copenhagen, Denmark.
- Department of Endocrinology and Metabolism, Copenhagen University Hospital, Rigshospitalet, Copenhagen, Denmark. marianne.christina.klose.01@regionh.dk.
ORCID-Profile: Sema Okutan, Stina Willemoes Borresen, Ulla Feldt-Rasmussen, Marianne Klose
Förderung
- Novo Nordisk
Zitieren
APA
Okutan, S., Jørgensen, N. T., Pedersen, L. E., Borresen, S. W., Hilsted, L., Hansen, L. F., Feldt-Rasmussen, U., & Klose, M. (2024). Determination of cortisol cut-off limits and steroid dynamics in the ACTH stimulation test: a comparative analysis using Roche Elecsys Cortisol II immunoassay and LC-MS/MS. Endocrine, 85(1), 321-330. https://doi.org/10.1007/s12020-024-03752-0
Vancouver
Okutan S, Jørgensen NT, Pedersen LE, Borresen SW, Hilsted L, Hansen LF, et al. Determination of cortisol cut-off limits and steroid dynamics in the ACTH stimulation test: a comparative analysis using Roche Elecsys Cortisol II immunoassay and LC-MS/MS. Endocrine. 2024;85(1):321-330. doi:10.1007/s12020-024-03752-0
BibTeX
@article{Okutan2024,
author = {Okutan, Sema and Jørgensen, Nanna Thurmann and Pedersen, Lars Engers and Borresen, Stina Willemoes and Hilsted, Linda and Hansen, Lennart Friis and Feldt-Rasmussen, Ulla and Klose, Marianne},
title = {Determination of cortisol cut-off limits and steroid dynamics in the ACTH stimulation test: a comparative analysis using Roche Elecsys Cortisol II immunoassay and LC-MS/MS},
journal = {Endocrine},
year = {2024},
volume = {85},
number = {1},
pages = {321--330},
doi = {10.1007/s12020-024-03752-0}
}
EUVIMED