Auf einen Blick

FDM- und ANF-Modelle zeigten bei 3 CI-Patient:innen: Ein apikaler Erdungskontakt senkte die Stimulationsschwellen apikal und erhöhte sie basaler im Vergleich zur extrakochleären Erdung.

Fragestellung

Welchen Einfluss hat die apicale Platzierung eines Erdungskontakts auf die Stimulation apikaler auditorischer Nervenfasern bei Cochlea-Implantaten?

Methode

FDM-Simulationen auf Basis hochauflösender Resistivitätskarten aus postoperativen CTs von 3 CI-Patient:innen; ANF-Modelle für 75 Bündel entlang der Cochlea; Vergleich apikaler vs. extrakochleärer Erdung.

Ergebnisse

Bei allen 3 Patient:innen waren die Stimulationsschwellen für apikale ANFs mit apikaler Erdung geringer als mit extrakochleärer Erdung. Für basale ANFs stiegen die Schwellen unter apikaler Erdung.

Schlussfolgerung der Autor:innen

Apikale Erdungskontakte können die Aktivierung meist unterstimulierter niederfrequenter ANFs erhöhen, ohne längere Kontaktarrays zu benötigen.

Kernaussagen

  • Apikale Erdung senkte die Schwellen apikaler ANFs in allen 3 Patient:innen
  • Basale ANFs zeigten unter apikaler Erdung höhere Schwellen
  • Mögliche Verbesserung der Wahrnehmung tiefer Frequenzen

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, online erschienen am 17.05.2026. Verfasst von Jared A Rybarczyk und 3 weiteren Autor:innen. Beteiligte Einrichtung: Vanderbilt University, Nashville, USA. jared.a.rybarczyk@vanderbilt.edu. Hauptthemen laut MEDLINE: Cochlear Implants, Cochlear Nerve, Computer Simulation, Cochlear Implantation. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIDCD NIH HHS, National Institute on Deafness and Other Communication Disorders.

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Cochlear implant (CI) electrode arrays typically do not extend along the entire cochlear duct, often resulting in poor stimulation of the most apical nerves. To better stimulate apical auditory nerve fibers (ANFs), previous studies have examined placing an isolated ground contact within the helicotrema, providing the ability to direct current apically. In this study, we utilize finite difference method (FDM) simulations and computational ANF models to study the effect of apical ground placement on apical ANF stimulation.

Methods

High-resolution resistivity maps based on postoperative CT scans for 3 CI patients were used in FDM-based generation of cochlear voltage maps. Models corresponding to the use of a typical extracochlear ground and a helicotrema ground were used to drive ANF models, allowing for generation of activation thresholds at each of 75 modeled ANF bundles throughout the cochlea. Thresholds between the two models were compared to assess the effects of an apical grounding contact on improving apical ANF stimulation.

Results

In all 3 patients, we observe a relative reduction in stimulation thresholds for apical ANFs when an apical ground is present as opposed to an extracochlear ground. Additionally, when an apical ground is used, we observe an increase in stimulation thresholds for more basal ANFs.

Conclusions

Apical ground contacts can increase activation of typically under-stimulated low frequency ANFs without the need for longer contact arrays, providing a solution for improving low frequency sound perception and potentially leading to better hearing quality.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cochlear ImplantsCochlear NerveComputer SimulationCochlear ImplantationHumansTomography, X-Ray ComputedElectric StimulationCochlea

Stichwörter der Autor:innen

Auditory nerve fiber modelingCochlear implantsImage-based modelingNeural stimulation

Autor:innen und Institutionen

  • Vanderbilt University, Nashville, USA. jared.a.rybarczyk@vanderbilt.edu.
  • Vanderbilt University, Nashville, USA.
  • Medical University of South Carolina, Charleston, USA.

Förderung

  • NIDCD NIH HHS
  • National Institute on Deafness and Other Communication Disorders

Zitieren

APA

Rybarczyk, J. A., Bratu, E. L., Labadie, R. F., & Noble, J. H. (2026). Simulation of apically grounded cochlear implant stimuli using neural stimulation models. International Journal of Computer Assisted Radiology and Surgery, 21(11), 2277-2285. https://doi.org/10.1007/s11548-026-03702-3

Vancouver

Rybarczyk JA, Bratu EL, Labadie RF, Noble JH. Simulation of apically grounded cochlear implant stimuli using neural stimulation models. International Journal of Computer Assisted Radiology and Surgery. 2026;21(11):2277-2285. doi:10.1007/s11548-026-03702-3

BibTeX
@article{Rybarczyk2026,
  author = {Rybarczyk, Jared A and Bratu, Erin L and Labadie, Robert F and Noble, Jack H},
  title = {Simulation of apically grounded cochlear implant stimuli using neural stimulation models},
  journal = {International Journal of Computer Assisted Radiology and Surgery},
  year = {2026},
  volume = {21},
  number = {11},
  pages = {2277--2285},
  doi = {10.1007/s11548-026-03702-3}
}
Kostenlos starten →