Improving hand-eye-coordination in laparoscopy using a virtual endoscope monitor
International Journal of Computer Assisted Radiology and Surgery · 2026
DOI: 10.1007/s11548-026-03620-4 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, 2026. Verfasst von Bernhard Michael Weber und 8 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Robotics and Mechatronics, German Aerospace Center, Muenchener Str. 20, 82234, Wessling, Germany. Bernhard.Weber@dlr.de. Hauptthemen laut MEDLINE: Laparoscopy, User-Computer Interface, Psychomotor Performance, Virtual Reality. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bayerische Forschungsstiftung.
Abstract (Originalsprache)
Purpose
While laparoscopic procedures offer considerable advantages for the patient, they pose substantial challenges for the surgeon, whose hand-eye coordination is impaired. One reason for this is the surgeon’s lost visual access to the surgical field, who is instead provided with an endoscopic image on a stationary monitor. In the present study, an optical see-through head-mounted display (OST-HMD) was used to display a virtual endoscope monitor. Unlike stationary monitors, this allows the endoscope image to be projected at arbitrary positions in the room.
Methods
The benefits of this approach were evaluated in a user study. N = 10 surgeons carried out a threading task in a pelvic trainer integrated in an operating room setup. Four positions of the endoscope image were compared: frontal view ( $$0^{\circ }$$ 0 ∘ ), offset by $$60^{\circ }$$ 60 ∘ to the left and right, and projected directly onto the surgical field.
Results
The results show that for complex bimanual tasks, projecting the endoscope image onto the surgical field led to a halving of execution times in the study. In addition, the forces applied during threading were measured, but no significant effects of perspective were found. More experienced surgeons also reported significantly lower mental workload with the frontal view ( $$0^{\circ }$$ 0 ∘ ) and the projection onto the surgical field compared to the other conditions.
Conclusion
This study provides first evidence that an OST-HMD could be a suitable way for flexibly displaying endoscopic images in laparoscopy and could help improve the surgeon’s hand-eye coordination, while at the same time being easily integrated into the clinical environment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Robotics and Mechatronics, German Aerospace Center, Muenchener Str. 20, 82234, Wessling, Germany. Bernhard.Weber@dlr.de.
- Institute of Robotics and Mechatronics, German Aerospace Center, Muenchener Str. 20, 82234, Wessling, Germany.
- Research Group Minimally Invasive Interdisciplinary Therapeutical Interventions (MITI), Technical University of Munich, Trogerstr. 10, 81675, Munich, Germany.
- TUM School of Medicine and Health, Technical University of Munich, Klinikum Rechts der Isar, Department of Surgery, Ismaninger Str. 22, 81675, Munich, Germany.
Förderung
- Bayerische Forschungsstiftung
Zitieren
APA
Weber, B. M., Connan, M., Kirst, A., Schneider, D., Wegener, L., Berlet, M., Fuchtmann, J., Wilhelm, D., & Klodmann, J. (2026). Improving hand-eye-coordination in laparoscopy using a virtual endoscope monitor. International Journal of Computer Assisted Radiology and Surgery. https://doi.org/10.1007/s11548-026-03620-4
Vancouver
Weber BM, Connan M, Kirst A, Schneider D, Wegener L, Berlet M, et al. Improving hand-eye-coordination in laparoscopy using a virtual endoscope monitor. International Journal of Computer Assisted Radiology and Surgery. 2026. doi:10.1007/s11548-026-03620-4
BibTeX
@article{Weber2026,
author = {Weber, Bernhard Michael and Connan, Mathilde and Kirst, Alexander and Schneider, Dominik and Wegener, Luca and Berlet, Maximilian and Fuchtmann, Jonas and Wilhelm, Dirk and Klodmann, Julian},
title = {Improving hand-eye-coordination in laparoscopy using a virtual endoscope monitor},
journal = {International Journal of Computer Assisted Radiology and Surgery},
year = {2026},
doi = {10.1007/s11548-026-03620-4}
}