MR-safe robotic needle driver for real-time MRI-guided minimally invasive procedures: a feasibility study
International Journal of Computer Assisted Radiology and Surgery · 2025
DOI: 10.1007/s11548-025-03545-4 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, 2025. Verfasst von Atharva Paralikar und 5 weiteren Autor:innen. Beteiligte Einrichtung: Sheikh Zayed Institute for Pediatric Surgical Innovation, Children's National Hospital, Washington, D.C., USA. aparalikar@childrensnational.org. Hauptthemen laut MEDLINE: Needles, Robotic Surgical Procedures, Magnetic Resonance Imaging, Interventional, Minimally Invasive Surgical Procedures. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, Foundation for the National Institutes of Health.
Abstract (Originalsprache)
PURPOSE: This article reports on the development and feasibility testing of an MR-safe robotic needle driver. The needle driver is pneumatically actuated and designed for automatic insertion and extraction of needles along a straight trajectory within the MRI scanner. METHOD: All parts use plastic resins and composite materials to ensure MR-safe operation. A needle could be clamped in the needle carriage using a pneumatically operated clamp. The clamp is designed to be easily attached and detached from the needle driver. Clamps with different opening sizes could accommodate a range of needles from 18 to 22 gauge. To mimic the manual procedure of needle insertion, a pneumatically operated rack-and-pinion mechanism simultaneously translates and rotates the needle carriage along a helical slot. Signal-to-noise ratio (SNR) and 2-D geometric distortion were measured to evaluate the MRI compatibility. Targeting was measured with an electromagnetic tracker. We also evaluated the maximum force that could be generated at the tip of the needle with different clamping pressures using a force sensor. RESULTS: We recorded the maximum percentage change in SNR for multiple configurations of needle drivers as 6.6% and the maximum geometric distortion at 0.24%. The needle driver's mean positioning accuracy for 105 targets at 50 mm depth was 2.38 ± 1.00 mm in a composite tissue phantom. The angulation error for the straight trajectory was 0.51°, and the mean linear trajectory deviation was statistically negligible. The measured force at the needle tip was 1.17N, 1.6N, and 2.12N at 30, 40, and 50 psi, respectively. CONCLUSION: This preliminary study showed that the prototype of our robotic needle driver works as intended for the insertion and extraction of the needle. The driver is MR-safe and serves as a suitable platform for MRI-guided interventions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Sheikh Zayed Institute for Pediatric Surgical Innovation, Children's National Hospital, Washington, D.C., USA. aparalikar@childrensnational.org.
- Sheikh Zayed Institute for Pediatric Surgical Innovation, Children's National Hospital, Washington, D.C., USA.
- Department of Neurosurgery, Children's National Hospital, Washington, D.C., USA.
- Sheikh Zayed Institute for Pediatric Surgical Innovation, Children's National Hospital, Washington, D.C., USA. rmonfare@childrensnational.org.
Förderung
- NIBIB NIH HHS
- Foundation for the National Institutes of Health
Zitieren
APA
Paralikar, A., Li, G., Oluigbo, C., Yarmolenko, P., Cleary, K., & Monfaredi, R. (2025). MR-safe robotic needle driver for real-time MRI-guided minimally invasive procedures: a feasibility study. International Journal of Computer Assisted Radiology and Surgery. https://doi.org/10.1007/s11548-025-03545-4
Vancouver
Paralikar A, Li G, Oluigbo C, Yarmolenko P, Cleary K, Monfaredi R. MR-safe robotic needle driver for real-time MRI-guided minimally invasive procedures: a feasibility study. International Journal of Computer Assisted Radiology and Surgery. 2025. doi:10.1007/s11548-025-03545-4
BibTeX
@article{Paralikar2025,
author = {Paralikar, Atharva and Li, Gang and Oluigbo, Chima and Yarmolenko, Pavel and Cleary, Kevin and Monfaredi, Reza},
title = {MR-safe robotic needle driver for real-time MRI-guided minimally invasive procedures: a feasibility study},
journal = {International Journal of Computer Assisted Radiology and Surgery},
year = {2025},
doi = {10.1007/s11548-025-03545-4}
}
EUVIMED