Evaluation of augmented reality guidance for glenoid pin placement in total shoulder arthroplasty
International Journal of Computer Assisted Radiology and Surgery · 2025
DOI: 10.1007/s11548-025-03444-8 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, 2025. Verfasst von Taylor Frantz und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electronics and Informatics (ETRO), Vrije Universiteit Brussel (VUB), Pleinlaan 2, 1050, Brussels, Belgium. tfrantz@vub.be. Hauptthemen laut MEDLINE: Augmented Reality, Arthroplasty, Replacement, Shoulder, Surgery, Computer-Assisted, Bone Nails. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Agentschap Innoveren en Ondernemen.
Abstract (Originalsprache)
Purpose
Computer-aided navigation and patient-specific 3D printed guides have demonstrated superior outcomes in total shoulder arthroplasty (TSA). Nevertheless, few TSAs are inserted using these technologies. Head-worn augmented reality (AR) devices can provide intuitive 3D computer navigation to the surgeon. This study investigates AR navigation in conjunction with adaptive spatial drift correction toward TSA.
Methods
A phantom study was performed to assess the performance of AR navigated pin placement in TSA. Two medical experts performed a total of 12 pin placements into phantom scapula; six were placed using an end-to-end AR-navigated technique, and six using a common freehand technique. Inside-out infrared (IR) tracking was designed and integrated into the AR headset to correct for device drift and provide tool tracking. Additionally, the impact of IR tool tracking, registration, and superposed/juxtaposed visualization techniques was investigated.
Results
The AR-navigated pin placement resulted in a mean entry point error of 1.06 mm ± 0.64 mm and directional error of $${1.66^\circ \pm 0.65^\circ }$$ 1 . 66 ∘ ± 0 . 65 ∘ . Compared with the freehand technique, AR navigation resulted in improved directional outcomes ( $$p=0.03$$ p = 0.03 ), while entry point accuracy was not significantly different ( $$p=0.44$$ p = 0.44 ). IR tool tracking error was 1.47 mm ± 0.69 mm and $${0.92^\circ \pm 0.50^\circ }$$ 0 . 92 ∘ ± 0 . 50 ∘ , and registration error was 4.32 mm ± 1.75 mm and $${2.56^\circ \pm 0.82^\circ }$$ 2 . 56 ∘ ± 0 . 82 ∘ . No statistical difference between AR visualization techniques was found in entry point ( $$p=0.22$$ p = 0.22 ) or directional ( $$p=0.31$$ p = 0.31 ) errors.
Conclusion
AR navigation allowed for comparable pin placement outcomes with those reported in the literature for patient-specific 3D printed guides; moreover, it complements the patient-specific planning without the need for the guides themselves.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electronics and Informatics (ETRO), Vrije Universiteit Brussel (VUB), Pleinlaan 2, 1050, Brussels, Belgium. tfrantz@vub.be.
- imec, Kapeldreef 75, 3001, Leuven, Belgium. tfrantz@vub.be.
- Department of Neurosurgery, Universitair Ziekenhuis Brussel (UZ Brussel), Laarbeeklaan 101, 1090, Brussels, Belgium.
- Materialise NV, Technologielaan 15, 3001, Leuven, Belgium.
- Department of Orthopedic Surgery and Traumatology, Universitair Ziekenhuis Brussel (UZ Brussel), Laarbeeklaan 101, 1090, Brussels, Belgium.
- Department of Electronics and Informatics (ETRO), Vrije Universiteit Brussel (VUB), Pleinlaan 2, 1050, Brussels, Belgium.
Förderung
- Agentschap Innoveren en Ondernemen
Zitieren
APA
Frantz, T., van Gestel, F., Slagmolen, P., Duerinck, J., Scheerlinck, T., & Vandemeulebroucke, J. (2025). Evaluation of augmented reality guidance for glenoid pin placement in total shoulder arthroplasty. International Journal of Computer Assisted Radiology and Surgery. https://doi.org/10.1007/s11548-025-03444-8
Vancouver
Frantz T, van Gestel F, Slagmolen P, Duerinck J, Scheerlinck T, Vandemeulebroucke J. Evaluation of augmented reality guidance for glenoid pin placement in total shoulder arthroplasty. International Journal of Computer Assisted Radiology and Surgery. 2025. doi:10.1007/s11548-025-03444-8
BibTeX
@article{Frantz2025,
author = {Frantz, Taylor and van Gestel, Frederick and Slagmolen, Pieter and Duerinck, Johnny and Scheerlinck, Thierry and Vandemeulebroucke, Jef},
title = {Evaluation of augmented reality guidance for glenoid pin placement in total shoulder arthroplasty},
journal = {International Journal of Computer Assisted Radiology and Surgery},
year = {2025},
doi = {10.1007/s11548-025-03444-8}
}
EUVIMED