Development of a human machine interface for robotically assisted surgery optimized for laparoscopic workflows
International Journal of Computer Assisted Radiology and Surgery · 2024
DOI: 10.1007/s11548-024-03239-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, 2024. Verfasst von Luca Wegener und 3 weiteren Autor:innen. Beteiligte Einrichtung: Klinikum rechts der Isar, Research Group MITI, TUM School of Medicine and Health, Technical University of Munich, Munich, Germany. l.wegener@tum.de. Hauptthemen laut MEDLINE: Laparoscopy, Robotic Surgical Procedures, Workflow, User-Computer Interface. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Technische Universität München.
Abstract (Originalsprache)
Introduction
In robotic-assisted surgery (RAS), the input device is the primary site for the flow of information between the user and the robot. Most RAS systems remove the surgeon’s console from the sterile surgical site. Beneficial for performing lengthy procedures with complex systems, this ultimately lacks the flexibility that comes with the surgeon being able to remain at the sterile site.
Methods
A prototype of an input device for RAS is constructed. The focus lies on intuitive control for surgeons and a seamless integration into the surgical workflow within the sterile environment. The kinematic design is translated from the kinematics of laparoscopic surgery. The input device uses three degrees of freedom from a flexible instrument as input. The prototype’s performance is compared to that of a commercially available device in an evaluation. Metrics are used to evaluate the surgeons’ performance with the respective input device in a virtual environment implemented for the evaluation.
Results
The evaluation of the two input devices shows statistically significant differences in the performance metrics. With the proposed prototype, the surgeons perform the tasks faster, more precisely, and with fewer errors.
Conclusion
The prototype is an efficient and intuitive input device for surgeons with laparoscopic experience. The placement in the sterile working area allows for seamless integration into the surgical workflow and can potentially enable new robotic approaches.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Klinikum rechts der Isar, Research Group MITI, TUM School of Medicine and Health, Technical University of Munich, Munich, Germany. l.wegener@tum.de.
- Klinikum rechts der Isar, Research Group MITI, TUM School of Medicine and Health, Technical University of Munich, Munich, Germany.
- Klinikum rechts der Isar, Department of Surgery, TUM School of Medicine and Health, Technical University of Munich, Munich, Germany.
- Munich Institute of Robotics and Machine Intelligence, Technical University of Munich, Munich, Germany.
Förderung
- Technische Universität München
Zitieren
APA
Wegener, L., Wilhelm, D., Berlet, M., & Fuchtmann, J. (2024). Development of a human machine interface for robotically assisted surgery optimized for laparoscopic workflows. International Journal of Computer Assisted Radiology and Surgery. https://doi.org/10.1007/s11548-024-03239-3
Vancouver
Wegener L, Wilhelm D, Berlet M, Fuchtmann J. Development of a human machine interface for robotically assisted surgery optimized for laparoscopic workflows. International Journal of Computer Assisted Radiology and Surgery. 2024. doi:10.1007/s11548-024-03239-3
BibTeX
@article{Wegener2024,
author = {Wegener, Luca and Wilhelm, Dirk and Berlet, Maximilian and Fuchtmann, Jonas},
title = {Development of a human machine interface for robotically assisted surgery optimized for laparoscopic workflows},
journal = {International Journal of Computer Assisted Radiology and Surgery},
year = {2024},
doi = {10.1007/s11548-024-03239-3}
}
EUVIMED